WECK IPN008818Recalls & reported problems

Teleflex Incorporated · VISISTAT 35R 6/BOX · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Teleflex Weck Visistat skin stapler is a staple used on the skin. Its 711 reports, none involving deaths or injuries, mostly describe staple jams, staples failing to fire or misfiring, and staples not forming properly. Recent reports say the skin was not captured properly and that jams forced a new stapler to be used. No recalls name this device, and reports in the last 12 months number 16, down from 231 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
711
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 711 adverse event reports that list this exact device (DI 04026704418190), received between May 13, 2021 and Mar 24, 2026 — classified as 711 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (500 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 17 reportsFeb 2024: 4 reportsMar 2024: 20 reportsApr 2024: 33 reportsMay 2024: 98 reportsJun 2024: 35 reportsJul 2024: 44 reportsAug 2024: 46 reportsSep 2024: 37 reportsOct 2024: 81 reportsNov 2024: 58 reportsDec 2024: 33 reportsJan 2025: 34 reportsFeb 2025: 27 reportsMar 2025: 12 reportsApr 2025: 3 reportsMay 2025: 20 reportsJun 2025: 12 reportsJul 2025: 12 reportsAug 2025: 10 reportsSep 2025: 6 reportsOct 2025: 4 reportsNov 2025: 4 reportsDec 2025: 5 reportsJan 2026: 4 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202698
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam500
  • Failure to Fire32
  • Misfire32
  • Failure to Form Staple29
  • Physical Resistance/Sticking28
  • Firing Problem23
  • Mechanical Problem16
  • Failure to Align11

Patient problems

  • No Clinical Signs, Symptoms or Conditions553
  • Insufficient Information155
  • Skin Inflammation/ Irritation3
  • Abrasion2
  • Skin Tears2

Recent adverse event reports

The 25 most recent of 711 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2026Failure to Form Staple

    IT WAS REPORTED THAT "THE SKIN WAS NOT CAPTURED PROPERLY WHEN APPLYING THE STAPLE". NO REPORT OF PATIENT HARM OR INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24681543 ↗
  • MalfunctionMar 24, 2026Failure to Form Staple

    IT WAS REPORTED THAT "THE SKIN WAS NOT CAPTURED PROPERLY WHEN APPLYING THE STAPLE". NO REPORT OF PATIENT HARM OR INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24681479 ↗
  • MalfunctionFeb 3, 2026Mechanical Jam

    IT WAS REPORTED THAT THE "STAPLES DID NOT FIRE FROM THE STAPLER". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24247892 ↗
  • MalfunctionJan 30, 2026Mechanical Jam

    IT WAS REPORTED THAT "STAPLE JAMMING". A NEW STAPLER WAS USED TO PROCEED WITH THE PROCEDURE. NO PATIENT HARM OR INJURY REPORTED. THE PATIENT'S STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24228493 ↗
  • MalfunctionJan 19, 2026Mechanical Jam

    IT WAS REPORTED THAT "STAPLE JAMMING." NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24111493 ↗
  • MalfunctionJan 16, 2026Mechanical Jam

    IT WAS REPORTED THAT "STAPLE JAMMING". A NEW STAPLER WAS USED TO PROCEED WITH THE PROCEDURE. NO PATIENT HARM OR INJURY REPORTED. THE PATIENT'S STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24099989 ↗
  • MalfunctionJan 9, 2026Mechanical Jam

    IT WAS REPORTED THAT "STAPLE JAMMING." NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24023590 ↗
  • MalfunctionDec 29, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT "STAPLE DEFORMED". A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO REPORT OF PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23907138 ↗
  • MalfunctionDec 23, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT "STAPLE DEFORMED". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23877856 ↗
  • MalfunctionDec 18, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT "TWO STAPLES WERE RELEASED AT ONCE." A NEW STAPLER WAS USED TO PROCEED WITH THE PROCEDURE. NO PATIENT HARM OR INJURY REPORTED. THE PATIENT'S STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23844015 ↗
  • MalfunctionDec 12, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT "STAPLE JAMMING". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23790167 ↗
  • MalfunctionDec 8, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT "IT WAS FOUND THAT THE STAPLER IS NOT FIRING DURING USE ON A PATIENT. THE PATIENT'S CONDITION IS FIND AND NO MEDICAL INTERVENTION REQUIRED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23742684 ↗
  • MalfunctionNov 27, 2025Failure to Align; Failure to Align

    IT WAS REPORTED THAT "THE STAPLES WERE COMING OUT ASKEW." MULTIPLE ATTEMPTS FOR ADDITIONAL INFORMATION HAVE BEEN REQUESTED FROM THE COMPLAINANT HAVE BEEN UNSUCCESSFUL AT THE TIME OF THIS REPORT.

    Patient problems: Insufficient Information

    MAUDE report 23666261 ↗
  • MalfunctionNov 14, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT "STAPLE JAMMING." A NEW SET WAS USED TO PROCEED WITH THE PROCEDURE. NO PATIENT HARM OR INJURY REPORTED. THE PATIENT'S STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23555106 ↗
  • MalfunctionNov 13, 2025Failure to Align; Failure to Align

    IT WAS REPORTED THAT "STAPLE MISALIGNED." A NEW SET WAS USED TO PROCEED WITH THE PROCEDURE. NO PATIENT HARM OR INJURY REPORTED. THE PATIENT'S STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23544287 ↗
  • MalfunctionNov 4, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT "THE STAPLER MALFUNCTIONED, IT JAMMED AND DELIVERED SEVERAL STAPLES AT ONCE. THE INCIDENT CONCERNED TWO DEVICES FROM THE SAME BATCH." NO REPORT OF PATIENT HARM OR INJURY. ASSOCIATED MDRS: 3003898360-2025-00807 AND 3003898360-2025-00811.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23470774 ↗
  • MalfunctionOct 30, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT "THE STAPLER MALFUNCTIONED, IT JAMMED AND DELIVERED SEVERAL STAPLES AT ONCE". NO REPORT OF PATIENT HARM OR INJURY.

    Patient problems: Insufficient Information

    MAUDE report 23426827 ↗
  • MalfunctionOct 15, 2025Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED THAT PRIOR TO USE IT WAS FOUND THAT THE STERILE PACKAGES WERE NOT SEALED PROPERLY. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23292303 ↗
  • MalfunctionOct 15, 2025Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED THAT PRIOR TO USE IT WAS FOUND THAT THE STERILE PACKAGES WERE NOT SEALED PROPERLY. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23292150 ↗
  • MalfunctionOct 15, 2025Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED THAT PRIOR TO USE IT WAS FOUND THAT THE STERILE PACKAGES WERE NOT SEALED PROPERLY. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23291627 ↗
  • MalfunctionSep 19, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT "THE USER RELEASED THE HANDLE AFTER FIRING A STAPLE, BUT IT DID NOT RETURN TO THE ORIGINAL POSITION DURING USE. HE/SHE REPLACED THE STAPLER WITH A NEW UNIT, BUT THE SAME ISSUE OCCURRED. THEREFORE, THE THIRD UNIT WAS USED TO COMPLETE THE PROCEDURE. NO INJURY TO THE PATIENT OCCURRED". ASSOCIATED MDR # 3003898360-2025-00722.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23097931 ↗
  • MalfunctionSep 19, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT "THE USER RELEASED THE HANDLE AFTER FIRING A STAPLE, BUT IT DID NOT RETURN TO THE ORIGINAL POSITION DURING USE. HE/SHE REPLACED THE STAPLER WITH A NEW UNIT, BUT THE SAME ISSUE OCCURRED. THEREFORE, THE THIRD UNIT WAS USED TO COMPLETE THE PROCEDURE. NO INJURY TO THE PATIENT OCCURRED". ASSOCIATED MDR # 3003898360-2025-00721.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23097840 ↗
  • MalfunctionSep 19, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT PRIOR TO USE, "STAPLE JAMMING". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23097817 ↗
  • MalfunctionSep 17, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT "STAPLE JAMMING". A NEW STAPLER WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23075655 ↗
  • MalfunctionSep 17, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT "STAPLE JAMMING". A NEW STAPLER WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23074732 ↗

Other STAPLE, REMOVABLE (SKIN) devices