Skip to content

Proximate PXW35Recalls & reported problems

Ethicon Endo-Surgery, LLC · Proximate skin Staplers (35 Wide) Contains 35 Stainless Steel Staples (Fixed Head) · Class 1

Quick summary

This Ethicon Proximate PXW35 skin stapler is a disposable stapler loaded with 35 wide stainless steel staples for closing skin. The 37 reports mostly describe the stapler misfiring or placing staples crossed, upside down or not forming them properly, and packages arriving with holes in them. Twenty reports involved injuries, such as wounds reopening and staples possibly left inside a patient; no deaths were reported. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 10705036002673), received between Apr 24, 2018 and Apr 29, 2026 — classified as 20 injuries, 17 malfunctions.

The most frequently reported specific device problem is “Mechanics Altered” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanics Altered15
  • Device Slipped5
  • Failure to Form Staple5
  • Firing Problem5
  • Patient Device Interaction Problem4
  • Break3
  • Delivered as Unsterile Product3
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13
  • Wound Dehiscence13
  • No Code Available10
  • No Consequences Or Impact To Patient4
  • Foreign Body In Patient3
  • Insufficient Information2
  • Hemorrhage/Blood Loss/Bleeding1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Mechanics Altered

    IT WAS REPORTED THAT DURING UNKNOWN PROCEDURE THE STAPLERS WERE DEPLOYING STAPLES IN A CROSSED OR MISALIGNED MANNER. AS A RESULT, THE SURGEON WAS UNABLE TO PROPERLY USE THE STAPLERS AS INTENDED. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25036030 ↗
  • MalfunctionApr 29, 2026Mechanics Altered

    IT WAS REPORTED THAT DURING UNKNOWN PROCEDURE THE STAPLERS WERE DEPLOYING STAPLES IN A CROSSED OR MISALIGNED MANNER. AS A RESULT, THE SURGEON WAS UNABLE TO PROPERLY USE THE STAPLERS AS INTENDED. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25036013 ↗
  • MalfunctionMar 25, 2026Delivered as Unsterile Product

    IT WAS REPORTED THAT BEFORE AN UNK PROCEDURE THAT THE STAPLER PACKAGE CAME WITH HOLE IN IT - LOSS OF STERILITY COULDN'T BE USED AS INTENDED. THERE WAS NO PATIENT CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24691203 ↗
  • MalfunctionJan 5, 2026Mechanics Altered

    IT WAS REPORTED THAT DURING OPEN PANCREAS SURGERY WHILE CLOSING THE WOUND WITH SKIN STAPLER, THE DEVICE DIDN'T WORK PROPERLY. STAPLERS WERE PLACED KIND OF "UPSIDE DOWN". NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23956048 ↗
  • InjuryDec 29, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE A SKIN STAPLE MIGHT HAVE BEEN LEFT INSIDE A PATIENT INTERNALLY. SHE WAS NOT SURE OR DID NOT HAVE THE INFORMATION CURRENTLY IF IT WAS LEFT IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 23907512 ↗
  • InjuryDec 29, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE A SKIN STAPLE MIGHT HAVE BEEN LEFT INSIDE A PATIENT INTERNALLY. SHE WAS NOT SURE OR DID NOT HAVE THE INFORMATION CURRENTLY IF IT WAS LEFT IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 23907487 ↗
  • MalfunctionOct 15, 2025Mechanics Altered; Firing Problem; Mechanics Altered; Firing Problem

    IT WAS REPORTED THAT DURING A VASCULAR PROCEDURE, THE DEVICE NOT FIRING ALL THE STAPLERS AND THE STAPLERS ARE NOT FORMING CORRECTLY. NO FURTHER INFORMATION PROVIDED. THERE WERE NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23278650 ↗
  • MalfunctionOct 15, 2025Mechanics Altered; Firing Problem; Mechanics Altered; Firing Problem

    IT WAS REPORTED THAT DURING A VASCULAR PROCEDURE, THE DEVICE NOT FIRING ALL THE STAPLERS AND THE STAPLERS ARE NOT FORMING CORRECTLY. NO FURTHER INFORMATION PROVIDED. THERE WERE NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23278005 ↗
  • MalfunctionJun 16, 2025Mechanics Altered; Mechanics Altered

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE WAS NOT STAPLING CORRECTLY. NOT FORMING B FORMATION STAPES. ADVERSE IMPACT PATIENT/USER: NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22210364 ↗
  • MalfunctionMay 1, 2025Mechanics Altered; Firing Problem; Mechanics Altered; Firing Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN NEUROSURGERY, THE DEVICE COULD NOT BE DEPLOYED AND THE STAPLE WAS UNFORMED WHEN TRIED TO DEPLOY THE STAPLES. ANOTHER COMPETITIVE DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21941572 ↗
  • InjuryApr 24, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT POST OP TO A LACERATION CLOSURE THE SURGEON THAT THE PATIENT SHOWED UP IN ED (B)(6) WITH A SCALP LACERATION. LACERATION WAS REPAIRED WITH 3 STAPLES FROM PXW35. PATIENT RETURNED YESTERDAY (B)(6) TO HAVE THE STAPLES AND THE STAPLES WERE NO LONGER VISIBLE. X-RAY CONFIRMED THEY WERE STILL PRESENT AND HAD FULLY LOOPED AS THE SHOULD. PATIENT WENT TO THE OPERATING ROOM FOR REMOVAL AND AFTER THE WOUND WAS RE-OPENED IT WAS DISCOVERED THAT ALL 3 STAPLES WERE EMBEDDED IN THE SKULL.

    Patient problems: Foreign Body In Patient

    MAUDE report 21900528 ↗
  • MalfunctionMar 18, 2025Mechanics Altered; Mechanics Altered

    IT WAS REPORTED BY A SALES REP THAT, DURING A LAPAROSCOPIC SIGMOIDECTOMY, MALFORMATION WAS FOUND THAT IT WAS DEPLOYED AND SLID SIDEWAYS WHEN USED THE PRODUCT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21633502 ↗
  • InjuryOct 3, 2024Mechanics Altered; Mechanics Altered

    IT WAS REPORTED THAT POST OP TO AN UNKNOWN PROCEDURE THE PATIENT EXPERIENCED THAT THE WOUND APPROXIMATION IS NOT SO TIGHTLY CLOSED, AND THAT THE WOUND BURSTS MORE THAN USUAL WHEN THE PATIENT RETURNS FOR CHECKUP. THEY ALSO COMPLAINED THAT THEY HAVE TO CHANGE STAPLES AND THAT DON'T ADHERE WELL ENOUGH.

    Patient problems: Wound Dehiscence

    MAUDE report 20360740 ↗
  • InjuryOct 3, 2024Mechanics Altered; Mechanics Altered

    IT WAS REPORTED THAT POST OP TO AN UNKNOWN PROCEDURE THE PATIENT EXPERIENCED THAT THE WOUND APPROXIMATION IS NOT SO TIGHTLY CLOSED, AND THAT THE WOUND BURSTS MORE THAN USUAL WHEN THE PATIENT RETURNS FOR CHECKUP. THEY ALSO COMPLAINED THAT THEY HAVE TO CHANGE STAPLES AND THAT DON'T ADHERE WELL ENOUGH.

    Patient problems: Wound Dehiscence

    MAUDE report 20360704 ↗
  • InjuryOct 3, 2024Mechanics Altered; Mechanics Altered

    IT WAS REPORTED THAT POST OP TO AN UNKNOWN PROCEDURE THE PATIENT EXPERIENCED THAT THE WOUND APPROXIMATION IS NOT SO TIGHTLY CLOSED, AND THAT THE WOUND BURSTS MORE THAN USUAL WHEN THE PATIENT RETURNS FOR CHECKUP. THEY ALSO COMPLAINED THAT THEY HAVE TO CHANGE STAPLES AND THAT DON'T ADHERE WELL ENOUGH.

    Patient problems: Wound Dehiscence

    MAUDE report 20360547 ↗
  • InjuryJul 27, 2023Insufficient Device Problem Information; Firing Problem; Insufficient Device Problem Information; Firing Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE PXW35 WAS NOT STAPLING CORRECTLY, JAMMED STAPLES BEING RELEASE AT A TIME WHICH ENDED UP TEARING PATIENT'S SKIN. THERE WAS PATIENT CONSEQUENCE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17404674 ↗
  • InjuryNov 1, 2022Mechanics Altered; Mechanics Altered

    IT WAS REPORTED THAT THE STAPLES WERE NOT FIXED TO THE SKIN. POSE OF STERI-STRIPS. PATIENT HOSPITALIZED FOR CAESAREAN SECTION. THE DAY AFTER THE OPERATION, THE STAPLES DID NOT ATTACH TO THE SKIN.

    Patient problems: Insufficient Information

    MAUDE report 15708304 ↗
  • MalfunctionOct 24, 2022Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT PRE-OP TO THE UNKNOWN PROCEDURE, THE PRODUCT PACKAGING WAS UNSEALED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15657616 ↗
  • InjuryJun 20, 2022Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT 2-WEEKS POST-OP FOLLOW-UP THE PATIENT¿S INCISION IS LOOKING INVERTED (SKIN EDGES ARE FOLDING OVER ON EACH OTHER). POOR ALIGNMENT OF STAPLE LINE SKIN CLOSURE. PROCEDURE: SKIN CLOSURE.

    Patient problems: Insufficient Information

    MAUDE report 14743035 ↗
  • MalfunctionJun 6, 2022Mechanics Altered; Mechanics Altered

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE STAPLES TOOK OFF. THE STAPLES DID NOT HOLD. PASSING THE HAND ON THEM MADE THEM FALL. SURGICAL DELAY. A SUTURE BY THREAD HAD BEEN PERFORMED INSTEAD. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14605452 ↗
  • InjuryMay 23, 2022Mechanics Altered; Adverse Event Without Identified Device or Use Problem; Failure to Eject; Firing Problem; Mechanics Altered; Adverse Event Without Identified Device or Use Problem; Failure to Eject; Firing Problem

    IT WAS REPORTED THAT DURING THE CATH LAB PROCEDURE, STAPLER DOES NOT GRAB THE SKIN VERY WELL, SO THE SKIN IS NOT APPROXIMATED. ONE THIRD OF THE TIME THE STAPLE JAMS. THE SURGEON STATED WHEN THEY DO GET IT TO WORK AND GRAB THE SKIN, THEN HALF OF TIME, IT DOESN'T RELEASE EASILY. THEY WOULD OPEN MULTIPLE DEVICES UNTIL THEY GOT ENOUGH STAPLES TO CLOSE THE SKIN. THERE WAS EXTRA BLOOD LOSS. THE PROCEDURE WAS COMPLETED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 14470249 ↗
  • MalfunctionApr 27, 2022Break; Break

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE A PIECE OF METAL CAME OUT OF THE STAPLER. A NEW DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14217979 ↗
  • MalfunctionJun 16, 2021Mechanics Altered; Mechanics Altered

    IT WAS REPORTED THAT THE PHYSICIAN USED SKIN STAPLER PXW35 ON A PATIENT TO CLOSE THE ABDOMINAL INCISION. WHEN THE DOCTOR HAD PATIENT COME BACK FOR CHECK UP HE PULLED DRESSING OFF AND THE STAPLES WERE OFF THE PATIENT AND THE INCISION WAS OPENED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12007856 ↗
  • MalfunctionJan 31, 2020Break

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE WAS BROKEN OFF AT THE FIRST FIRING. NO PIECES WERE LEFT INSIDE THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9652740 ↗
  • MalfunctionMay 16, 2019Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE ONLY THING WRONG WITH THIS ITEM WAS THAT THE PACKAGING HAD A HOLE IN IT. AS IT COULD NO LONGER BE CONSIDERED STERILE, THE OR COULD NOT USE IT. PATIENT CONSEQUENCE WAS NOT REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8615389 ↗

Other STAPLE, REMOVABLE (SKIN) devices

All STAPLE, REMOVABLE (SKIN) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.