Proximate PMW35Recalls & reported problems
Ethicon Endo-Surgery, LLC · Proximate Plus MD Multi-Directional Release Skin Staplers contains 35 Stainless Steel Staples approximate Closed Dimensions: 6.9mm x 3.9mm Wide · Class 1
All models of PROXIMATE (41)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Proximate skin stapler has 131 adverse event reports on record, with no deaths and 40 injuries. The reports mostly describe staples that did not close, bent, or fell out of the device, and several say the problem appeared with different users. Some mention wounds reopening, slow healing, bleeding, or infection. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 131
- Deaths reported
- 0
- Injuries reported
- 40
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 131 adverse event reports that list this exact device (DI 10705036002642), received between Apr 1, 2013 and Mar 13, 2026 — classified as 40 injuries, 91 malfunctions.
The most frequently reported specific device problem is “Mechanics Altered” (107 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanics Altered107
- Firing Problem18
- Adverse Event Without Identified Device or Use Problem13
- Patient Device Interaction Problem11
- Delivered as Unsterile Product6
- Defective Device1
- Device Slipped1
- Failure to Eject1
Patient problems
- No Clinical Signs, Symptoms or Conditions81
- Wound Dehiscence33
- Impaired Healing17
- Hemorrhage/Blood Loss/Bleeding15
- Pain15
- Unspecified Infection10
- No Consequences Or Impact To Patient5
- Insufficient Information2
Recent adverse event reports
The 25 most recent of 131 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 13, 2026Mechanics Altered; Firing Problem
THE CUSTOMER REPORTED THAT THE STAPLES ON THE STAPLER WERE NOT CLOSING, WERE BENDING, OR WERE SIMPLY FALLING OUT OF THE DEVICE. THE PROBLEM WAS INITIALLY MONITORED TO RULE OUT USER ERROR, BUT THE ISSUE OCCURRED WITH DIFFERENT USERS. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24590895 ↗MalfunctionMar 12, 2026Mechanics Altered; Firing Problem
THE CUSTOMER REPORTED THAT THE STAPLES ON THE STAPLER WERE NOT CLOSING, WERE BENDING, OR WERE SIMPLY FALLING OUT OF THE DEVICE. THE PROBLEM WAS INITIALLY MONITORED TO RULE OUT USER ERROR, BUT THE ISSUE OCCURRED WITH DIFFERENT USERS. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24582898 ↗MalfunctionMar 12, 2026Mechanics Altered; Firing Problem
THE CUSTOMER REPORTED THAT THE STAPLES ON THE STAPLER WERE NOT CLOSING, WERE BENDING, OR WERE SIMPLY FALLING OUT OF THE DEVICE. THE PROBLEM WAS INITIALLY MONITORED TO RULE OUT USER ERROR, BUT THE ISSUE OCCURRED WITH DIFFERENT USERS. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24582730 ↗MalfunctionMar 12, 2026Mechanics Altered; Firing Problem
THE CUSTOMER REPORTED THAT THE STAPLES ON THE STAPLER WERE NOT CLOSING, WERE BENDING, OR WERE SIMPLY FALLING OUT OF THE DEVICE. THE PROBLEM WAS INITIALLY MONITORED TO RULE OUT USER ERROR, BUT THE ISSUE OCCURRED WITH DIFFERENT USERS. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24582167 ↗MalfunctionMar 12, 2026Mechanics Altered; Firing Problem
THE CUSTOMER REPORTED THAT THE STAPLES ON THE STAPLER WERE NOT CLOSING, WERE BENDING, OR WERE SIMPLY FALLING OUT OF THE DEVICE. THE PROBLEM WAS INITIALLY MONITORED TO RULE OUT USER ERROR, BUT THE ISSUE OCCURRED WITH DIFFERENT USERS. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24582149 ↗MalfunctionMar 12, 2026Mechanics Altered; Firing Problem
THE CUSTOMER REPORTED THAT THE STAPLES ON THE STAPLER WERE NOT CLOSING, WERE BENDING, OR WERE SIMPLY FALLING OUT OF THE DEVICE. THE PROBLEM WAS INITIALLY MONITORED TO RULE OUT USER ERROR, BUT THE ISSUE OCCURRED WITH DIFFERENT USERS. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24582126 ↗MalfunctionMar 12, 2026Mechanics Altered; Firing Problem
THE CUSTOMER REPORTED THAT THE STAPLES ON THE STAPLER WERE NOT CLOSING, WERE BENDING, OR WERE SIMPLY FALLING OUT OF THE DEVICE. THE PROBLEM WAS INITIALLY MONITORED TO RULE OUT USER ERROR, BUT THE ISSUE OCCURRED WITH DIFFERENT USERS. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24582098 ↗MalfunctionMar 12, 2026Mechanics Altered; Firing Problem
THE CUSTOMER REPORTED THAT THE STAPLES ON THE STAPLER WERE NOT CLOSING, WERE BENDING, OR WERE SIMPLY FALLING OUT OF THE DEVICE. THE PROBLEM WAS INITIALLY MONITORED TO RULE OUT USER ERROR, BUT THE ISSUE OCCURRED WITH DIFFERENT USERS. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24581973 ↗MalfunctionMar 12, 2026Mechanics Altered; Firing Problem
THE CUSTOMER REPORTED THAT THE STAPLES ON THE STAPLER WERE NOT CLOSING, WERE BENDING, OR WERE SIMPLY FALLING OUT OF THE DEVICE. THE PROBLEM WAS INITIALLY MONITORED TO RULE OUT USER ERROR, BUT THE ISSUE OCCURRED WITH DIFFERENT USERS. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24581874 ↗MalfunctionMar 12, 2026Mechanics Altered; Firing Problem
THE CUSTOMER REPORTED THAT THE STAPLES ON THE STAPLER WERE NOT CLOSING, WERE BENDING, OR WERE SIMPLY FALLING OUT OF THE DEVICE. THE PROBLEM WAS INITIALLY MONITORED TO RULE OUT USER ERROR, BUT THE ISSUE OCCURRED WITH DIFFERENT USERS. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24581787 ↗MalfunctionMar 2, 2026Mechanics Altered
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, USING A SKIN STAPLER RESULTED IN POOR STAPLE FORMATION. CHANGED THE NEW ONE TO COMPLETE SURGERY. NO ADDITIONAL INFORMATION CAN BE PROVIDED. NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24490337 ↗MalfunctionMar 2, 2026Mechanics Altered
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, USING A SKIN STAPLER RESULTED IN POOR STAPLE FORMATION. CHANGED THE NEW ONE TO COMPLETE SURGERY. NO ADDITIONAL INFORMATION CAN BE PROVIDED. NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24490301 ↗MalfunctionFeb 27, 2026Mechanics Altered
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24474827 ↗MalfunctionFeb 27, 2026Mechanics Altered
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24474265 ↗MalfunctionFeb 24, 2026Mechanics Altered
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24431400 ↗MalfunctionJan 14, 2026Mechanics Altered
IT WAS REPORTED THAT DURING THE HEART SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24066601 ↗MalfunctionJan 14, 2026Mechanics Altered
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24066485 ↗MalfunctionJan 14, 2026Mechanics Altered
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24066475 ↗MalfunctionJan 12, 2026Mechanics Altered
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24037441 ↗MalfunctionJan 12, 2026Mechanics Altered
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24037116 ↗MalfunctionJan 8, 2026Mechanics Altered
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24010773 ↗MalfunctionJan 8, 2026Mechanics Altered
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24010730 ↗MalfunctionJan 5, 2026Mechanics Altered
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23956085 ↗MalfunctionJan 5, 2026Mechanics Altered
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE STAPLER CANNOT HOLD THE STAPLES. CHANGED ANOTHER ONE TO COMPLETE SURGERY. NO ADDITIONAL INFORMATION CAN BE PROVIDED. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23955959 ↗MalfunctionDec 31, 2025Mechanics Altered; Mechanics Altered
IT WAS REPORTED THAT DURING THE NEPHRECTOMY SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY, THE DEVICE WAS DAMAGED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23934386 ↗
FDA clearance and classification
The FDA regulates product code GDT (staple, removable (skin)) under 21 CFR 878.4760 as a Class I device. 21 CFR 878.4760 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Proximate PMW35
Has the Proximate PMW35 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036002642), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Proximate PMW35?
The FDA's MAUDE database lists 131 adverse event reports for this device, received between Apr 1, 2013 and Mar 13, 2026. In those reports, the most common specific device problems are “Mechanics Altered” (107), “Firing Problem” (18) and “Patient Device Interaction Problem” (11); the most common patient problems are “Wound Dehiscence” (33), “Impaired Healing” (17) and “Hemorrhage/Blood Loss/Bleeding” (15). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Proximate PMW35?
None of the reports are classified as deaths, 40 as injuries and 91 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Proximate PMW35 FDA approved?
Devices of this type (product code GDT) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Ethicon Endo-Surgery, LLC.
Is the Proximate PMW35 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Proximate PMW35, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand Proximate, model PMW35). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STAPLE, REMOVABLE (SKIN) devices
- WECK IPN028492Teleflex Incorporated · STAPLE, REMOVABLE (SKIN)1,377 reports
- WECK IPN008818Teleflex Incorporated · STAPLE, REMOVABLE (SKIN)711 reports
- WECK IPN028493Teleflex Incorporated · STAPLE, REMOVABLE (SKIN)562 reports
- Appose ULC 8886803712Covidien LP · STAPLE, REMOVABLE (SKIN)68 reports
- Appose ULC 8886803512Covidien LP · STAPLE, REMOVABLE (SKIN)65 reports
- Proximate PXW35Ethicon Endo-Surgery, LLC · STAPLE, REMOVABLE (SKIN)37 reports
- Proximate PRW35Ethicon Endo-Surgery, LLC · STAPLE, REMOVABLE (SKIN)20 reports
- WECK IPN008820Teleflex Incorporated · STAPLE, REMOVABLE (SKIN)14 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page