Skip to content

Appose ULC 8886803712Recalls & reported problems

Covidien LP · Slim Body Skin Stapler · Class 1

All models of Appose ULC (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Covidien Appose Slim Body skin stapler is a disposable device that closes surgical skin incisions with staples. The 69 reports mostly describe the stapler jamming or misfiring, staples breaking inside the device or failing to close the skin, and parts detaching, sometimes forcing the wound to be closed with stitches instead. Eight injuries were reported and no deaths;61 reports describe device problems with no breakdown recorded. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
68
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 68 adverse event reports that list this exact device (DI 10884521104372), received between Jul 8, 2016 and Aug 6, 2026 — classified as 8 injuries, 60 malfunctions.

The most frequently reported specific device problem is “Entrapment of Device” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 6 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 3 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem26
  • Entrapment of Device14
  • Failure to Form Staple13
  • Break9
  • Detachment of Device or Device Component8
  • Failure to Fire7
  • Failure to Align4
  • Mechanical Jam3

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • No Known Impact Or Consequence To Patient13
  • Unspecified Tissue Injury13
  • Insufficient Information12
  • No Consequences Or Impact To Patient5
  • Post Operative Wound Infection3
  • Unintended Radiation Exposure3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2

Recent adverse event reports

The 25 most recent of 68 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2026Failure to Form Staple; Adverse Event Without Identified Device or Use Problem

    Patient problems: Laceration(s)

    MAUDE report 50161193 ↗
  • MalfunctionMay 19, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING THE PROCEDURE, THE STAPLES DID NOT CLOSE PROPERLY FOR THE THREE SKIN STAPLERS. ABSORBABLE SUTURE WAS USED IN ORDER TO CLOSE THE WOUND AND COMPLETE THE CASE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 25224597 ↗
  • MalfunctionMay 19, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING THE PROCEDURE, THE STAPLES DID NOT CLOSE PROPERLY FOR THE THREE SKIN STAPLERS. ABSORBABLE SUTURE WAS USED IN ORDER TO CLOSE THE WOUND AND COMPLETE THE CASE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 25224563 ↗
  • MalfunctionApr 28, 2026Failure to Align; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER INTRA-OPERATIVELY, THE STAPLER MISFIRED AND WAS NOT ALIGNED PROPERLY DURING FIRING WHICH CAUSED TISSUE DAMAGE TO THE PATIENT. THE DEVICE USE WAS DISCONTINUED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Tissue Injury

    MAUDE report 25011819 ↗
  • MalfunctionMar 25, 2026Material Separation

    IT WAS REPORTED THAT DURING THE PROCEDURE, IMMEDIATELY AFTER THE SQUEEZE, THE STAPLES BROKE INSIDE THE DEVICE, WHICH SUBSEQUENTLY CAUSED THE JAMMING OF THE MECHANISM. ANOTHER SAME TYPE OF DEVICE FROM THE SAME COMPANY WAS USED TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24692042 ↗
  • MalfunctionJan 15, 2026Unintended Ejection; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, THE PATIENTS DEVELOPED WOUND DEHISCENCE BECAUSE THE SKIN STAPLES WERE RELEASED TOO EARLY.

    Patient problems: Wound Dehiscence

    MAUDE report 24079942 ↗
  • MalfunctionJan 15, 2026Unintended Ejection; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, THE PATIENTS DEVELOPED WOUND DEHISCENCE BECAUSE THE SKIN STAPLES WERE RELEASED TOO EARLY.

    Patient problems: Wound Dehiscence

    MAUDE report 24079919 ↗
  • MalfunctionDec 12, 2025Failure to Form Staple; Adverse Event Without Identified Device or Use Problem; Failure to Form Staple; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A PROCEDURE USING THE SKIN STAPLER, THE DEVICE WAS ACTIVATED AND THE STAPLES WERE DEPLOYED; HOWEVER, THE STAPLES FAILED TO PROPERLY ENGAGE OR PENETRATE THE SKIN, RESULTING IN THE WOUND REMAINING OPEN. THE DEVICE MALFUNCTION REQUIRED ADDITIONAL MANIPULATION OF THE WOUND, AND EXTENDED THE SURGERY TIME. ANOTHER SAME TYPE OF DEVICE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 23786527 ↗
  • MalfunctionOct 22, 2025Material Separation; Failure to Form Staple; Material Separation; Failure to Form Staple

    IT WAS REPORTED THAT DURING A PROCEDURE, AT SKIN CLOSURE, THE STAPLE BROKE, AND A STAPLE WENT INSIDE WITHOUT BEING CAUGHT. ANOTHER DEVICE OF THE SAME TYPE WAS USED TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23358983 ↗
  • MalfunctionSep 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, POST-OPERATIVELY, THE PATIENT HAD SUBCUTANEOUS INFECTION AT THE SITE WHERE THE STAPLES WERE USED FOR SKIN CLOSURE. AN ORAL DOXYCYCLINE COURSE WAS DONE TO RESOLVE THE ISSUE.

    Patient problems: Post Operative Wound Infection

    MAUDE report 23009514 ↗
  • MalfunctionAug 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, POST-OPERATIVELY, THE PATIENT HAD SUBCUTANEOUS INFECTION AT THE SITE WHERE THE STAPLES WERE USED FOR SKIN CLOSURE. AN ORAL DOXYCYCLINE COURSE WAS DONE TO RESOLVE THE ISSUE.

    Patient problems: Post Operative Wound Infection

    MAUDE report 22920620 ↗
  • MalfunctionAug 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER POST-OPERATIVELY, THE PATIENT HAD SUBCUTANEOUS INFECTION AT THE SITE WHERE THE STAPLES WERE USED FOR SKIN CLOSURE. AN ORAL DOXYCYCLINE COURSE WAS DONE TO RESOLVE THE ISSUE.

    Patient problems: Post Operative Wound Infection

    MAUDE report 22920597 ↗
  • MalfunctionAug 27, 2025Break; Mechanics Altered; Break; Mechanics Altered

    ACCORDING TO THE REPORTER PREOPERATIVELY UPON OPENING THE PACKAGING, THE STAPLER BROKE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22903521 ↗
  • InjuryJul 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING A WOUND CLOSURE, THE SKIN STAPLER HAD 38 STAPLES IN IT. AN X-RAY WAS TAKEN.

    Patient problems: Unintended Radiation Exposure

    MAUDE report 22607898 ↗
  • MalfunctionJun 24, 2025Entrapment of Device; Entrapment of Device

    IT WAS REPORTED THAT DURING AN OPEN PROCEDURE TO TREAT A HEAD LACERATION, THE SKIN STAPLER MALFUNCTIONED AND BECAME STUCK IN THE PAT IENT'S HEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 22319397 ↗
  • MalfunctionMar 5, 2025Entrapment of Device; Failure to Align; Entrapment of Device; Failure to Align

    ACCORDING TO THE REPORTER, DURING A SKIN CLOSURE, AT THE TIME OF STAPLING, THE STAPLE REMAINED STUCK ON THE PATIENT'S SCALP. THERE WAS DIFFICULTY FOR THE SURGEON TO GET IT OFF, AND THE STAPLES APPEARED TO BE POORLY POSITIONED IN THE SKIN STAPLER. ANOTHER SKIN STAPLER FROM A DIFFERENT BATCH PRESENTED THE SAME ISSUE. THE SURGEON REMOVED THE STAPLE FROM THE WOUND AND SUTURED IT IN PLACE. SURGICAL TIME WAS EXTENDED FOR LESS THAN 30 MINUTES.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 21527168 ↗
  • MalfunctionMar 5, 2025Entrapment of Device; Failure to Align; Entrapment of Device; Failure to Align

    ACCORDING TO THE REPORTER, DURING A SKIN CLOSURE, AT THE TIME OF STAPLING, THE STAPLE REMAINED STUCK ON THE PATIENT'S SCALP. THERE WAS DIFFICULTY FOR THE SURGEON TO GET IT OFF, AND THE STAPLES APPEARED TO BE POORLY POSITIONED IN THE SKIN STAPLER. ANOTHER SKIN STAPLER FROM A DIFFERENT BATCH PRESENTED THE SAME ISSUE. THE SURGEON REMOVED THE STAPLE FROM THE WOUND AND SUTURED IT IN PLACE. SURGICAL TIME WAS EXTENDED FOR LESS THAN 30 MINUTES.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 21527125 ↗
  • MalfunctionFeb 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, INTRAOPERATIVELY, THE SKIN STAPLES WERE NOT HOLDING AND FELL OUT EASILY PRIOR TO FIRING.

    Patient problems: Insufficient Information

    MAUDE report 21305873 ↗
  • MalfunctionFeb 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, INTRAOPERATIVELY, THE SKIN STAPLES WERE NOT HOLDING AND FELL OUT EASILY PRIOR TO FIRING.

    Patient problems: Insufficient Information

    MAUDE report 21305869 ↗
  • MalfunctionFeb 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, INTRAOPERATIVELY, THE SKIN STAPLES WERE NOT HOLDING AND FELL OUT EASILY PRIOR TO FIRING.

    Patient problems: Insufficient Information

    MAUDE report 21305848 ↗
  • MalfunctionFeb 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, INTRAOPERATIVELY, THE SKIN STAPLES WERE NOT HOLDING AND FELL OUT EASILY PRIOR TO FIRING.

    Patient problems: Insufficient Information

    MAUDE report 21305325 ↗
  • MalfunctionFeb 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT INTRAOPERATIVELY, THE SKIN STAPLES WERE NOT HOLDING AND FELL OUT EASILY PRIOR TO FIRING.

    Patient problems: Insufficient Information

    MAUDE report 21304284 ↗
  • MalfunctionFeb 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING THE TOTAL KNEE ARTHROPLASTY, THE SKIN STAPLES WERE NOT HOLDING AND FELL OUT EASILY PRIOR TO FIRING. A NEW STAPLER WITH A DIFFERENT LOT NUMBER WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21293526 ↗
  • MalfunctionJan 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING AN OPEN BREAST PROCEDURE, IN SKIN CLOSURE, THE SKIN TISSUE TORE UPON THE APPLICATION OF THE STAPLER.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 21161397 ↗
  • MalfunctionMar 27, 2024Break; Entrapment of Device; Failure to Form Staple; Failure to Fire; Detachment of Device or Device Component; Break; Entrapment of Device; Failure to Form Staple; Failure to Fire; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A SKIN CLOSURE PROCEDURE, ALL SIX DEVICES DISINTEGRATE INTO THEIR INDIVIDUAL PARTS DURING USE, THE INDIVIDUAL STAPLES DO NOT CLOSE COMPLETELY SO THAT THERE WAS NO HOLD IN THE SKIN OR THE WOUND SUTURE DID NOT CLOSE. IT WAS ALSO REPORTED THAT THE STAPLING WAS NOT POSSIBLE OR THE INDIVIDUAL STAPLES GET CAUGHT IN THE DEVICE. THERE WAS NO PATIENT INJURY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18988344 ↗

FDA clearance and classification

The FDA regulates product code GDT (staple, removable (skin)) under 21 CFR 878.4760 as a Class I device. 21 CFR 878.4760 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Appose ULC 8886803712

Has the Appose ULC 8886803712 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10884521104372), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Appose ULC 8886803712?

The FDA's MAUDE database lists 68 adverse event reports for this device, received between Jul 8, 2016 and Aug 6, 2026. In those reports, the most common specific device problems are “Entrapment of Device” (14), “Failure to Form Staple” (13) and “Break” (9); the most common patient problems are “Unspecified Tissue Injury” (13), “Post Operative Wound Infection” (3) and “Unintended Radiation Exposure” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Appose ULC 8886803712?

None of the reports are classified as deaths, 8 as injuries and 60 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Appose ULC 8886803712 FDA approved?

Devices of this type (product code GDT) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Covidien LP.

Is the Appose ULC 8886803712 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Appose ULC 8886803712, and is it still sold?

The FDA's device database lists Covidien LP as the company responsible for it (brand Appose ULC, model 8886803712). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other STAPLE, REMOVABLE (SKIN) devices

All STAPLE, REMOVABLE (SKIN) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page