ESSENZ RP85Recalls & reported problems
LivaNova Deutschland GmbH · Class 2
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04033817903086), received between Apr 5, 2023 and Sep 18, 2026 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Pumping Stopped” (7 reports).
1 other recall cover devices cleared under the same FDA submission (K232291); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Pumping Stopped7
- Pumping Problem4
Patient problems
- No Clinical Signs, Symptoms or Conditions11
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2026Pumping Stopped
LivaNova Deutschland received a report that Essenz HLM CPL-C pump displayed an error code indicating that the pump was not working properly (2335). Reportedly, the error could not be cleared despite powering down and rebooting the HLM system.The affected roller pump was then replaced with a loan unit. The event occurred during priming.There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50814442 ↗MalfunctionJul 24, 2025Pumping Problem; Pumping Problem
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN ESSENZ ROLLER PUMP 85, USED AS BLOOD CARDIOPLEGIA (CPL-B) ONE, WAS DEFECTIVE DURING PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22588365 ↗MalfunctionJul 21, 2025Pumping Problem; Pumping Problem
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN ESSENZ ROLLER PUMP 85, USED AS SUCTION PUMP, GAVE THE FOLLOWING ERROR MESSAGE: "SUCTION 3 PUMP ISSUE (REPLACE SUCTION 3 PUMP AFTER CASE)" DURING PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22561607 ↗MalfunctionJul 10, 2025Pumping Problem; Pumping Problem
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22466337 ↗MalfunctionMay 28, 2025Pumping Stopped; Pumping Stopped
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN ESSENZ ROLLER PUMP 85, USED FOR CRYSTALLOID CARDIOPLEGIA DELIVERY WITH 4:1 RATIO, DID NOT TURN AND AN ALARM WENT OFF. THE ISSUE OCCURRED DURING PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22096181 ↗MalfunctionMay 26, 2025Pumping Stopped; Pumping Stopped
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22084449 ↗MalfunctionFeb 19, 2025Pumping Stopped; Pumping Stopped
LIVANOVA RECEIVED REPORT THAT, DURING A PROCEDURE, THE CARDIOPLEGIA ROLLER PUMP 85 NO LONGER WORKED. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21411424 ↗MalfunctionOct 31, 2024Pumping Problem; Pumping Problem
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT ROLLER PUMP 85 DISPLAYED AN ALARM PUMP DEFECTIVE DURING A PROCEDURE. MEDICAL TEAM REPLACED THE CARDIOPLEGIA CRYSTAL PUMP. THERE IS NO REPORT OF ANY PATIENT INJURY. CUSTOMER: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20578102 ↗MalfunctionAug 7, 2024Pumping Stopped; Pumping Stopped
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN ESSENZ DOUBLE ROLLER PUMP GAVE AN ERROR MESSAGE ('PUMP DEFECTIVE') AND WOULD NOT TURN WHEN STARTED DURING PROCEDURE. NOT RESPONSE EVEN WHEN OPERATED AFTER SEVERAL ATTEMPTS. THEREFORE, PUMP WAS SWAPPED OUT WITH A LOANER ONE BY THE CUSTOMER. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19928099 ↗MalfunctionMar 20, 2024Pumping Stopped; Pumping Stopped
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18939212 ↗MalfunctionApr 5, 2023Pumping Stopped; Pumping Stopped
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN ESSENZ DOUBLE HEAD PUMP, USED AS SUCTION PUMP, STOPPED DURING A PEDIATRIC CASE AND GAVE A RUNAWAY ERROR MESSAGE (THE PUMP RAN TOO FAST - SET VALUE DIFFERENT TO ACTUAL VALUE). THE CUSTOMER UNPLUGGED AND RE-PLUGGED THE CONNECTION CABLE OF THE INVOLVED PUMP, THEN THE PUMP RAN FOR TWENTY (20) SECONDS BEFORE STOPPING AGAIN AND GIVING THE SAME ERROR MESSAGE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16681605 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K232291. One submission often covers many models, so check the product and lot details in each record.
- Z-0571-2025 · Oct 18, 2024 · Open, ClassifiedEssenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software
Other CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices
- Terumo® Advanced Perfusion System 1 801188Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass3 recalls447 reports
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
- Terumo® Advanced Perfusion System 1 816300Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass300 reports
- Terumo® Advanced Perfusion System 1 816571Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass230 reports
- Terumo® Advanced Perfusion System 1 801763Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 death230 reports
- ESSENZ HLMLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death110 reports
- Terumo® Advanced Perfusion System 1 5773Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall78 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.