ESSENZRecalls & reported problems across all models
LivaNova Deutschland GmbH · 8 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one ESSENZ model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 185
- Deaths reported
- 1
- Injuries reported
- 4
- Recalls
- 1
We found 8 models of ESSENZ by LivaNova Deutschland GmbH in the FDA's device registry (GUDID). Together they are named in 185 adverse event reports in the FDA's MAUDE database, received between Apr 5, 2023 and Sep 25, 2026 — classified as 1 death, 4 injuries, 180 malfunctions.
The most frequently reported specific device problem is “Application Program Problem” (40 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
ESSENZ recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Oct 18, 2024Open, Classified
Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.
Z-0571-2025 · Root cause: Software change control · Recall details · FDA record ↗
Most reported ESSENZ problems
Device problems
- Application Program Problem40
- Difficult to Open or Close40
- Pumping Stopped31
- Computer Software Problem18
- Pumping Problem14
- No Apparent Adverse Event7
- Unstable7
- Device Sensing Problem5
Patient problems
- No Clinical Signs, Symptoms or Conditions183
- Hemolysis1
- Insufficient Information1
ESSENZ models
8 models, most reported first. Each model has its own page with its reports, recalls and device details.
- ESSENZ HLMCONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death110 reports
- ESSENZ Venous clampCONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS22 reports
- ESSENZ CockpitCONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS18 reports
- ESSENZ RP85CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS11 reports
- ESSENZ Arterial Clamp- AC arm long (620 mm)CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS10 reports
- ESSENZ Arterial Clamp - AC arm short (500mm)CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS6 reports
- ESSENZ RP150CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 reports
- ESSENZ CPDCONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 reports
Recent ESSENZ adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026Difficult to Open or Close
LivaNova Deutschland received a report that the Essenz Venous Clamp (VC) displayed an error code indicating that the clamp was defective (E2551). The issue occurred during the procedure. There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50897289 ↗MalfunctionSep 24, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50884955 ↗MalfunctionSep 24, 2026Application Program Problem
LivaNova Deutschland received report that Essenz HLM cockpit rebooted by itself during procedure. Reportedly, the pump remained controllable using the backup control unit. There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50884930 ↗MalfunctionSep 23, 2026Difficult to Open or Close
LivaNova Deutschland received a report that Essenz venous clamp (VC) was not working properly, even if it was calibrated. As troubleshooting, the tube was removed and then placed in again but the problem was not solved.There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870563 ↗MalfunctionSep 21, 2026Application Program Problem
LivaNova Deutschland received report that, before initiating procedure, Essenz HLM cockpit rebooted by itself.Reportedly, perfusionist started a new case and the procedure was completed without further problems.There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50836341 ↗MalfunctionSep 18, 2026Pumping Stopped
LivaNova Deutschland received a report that Essenz HLM CPL-C pump displayed an error code indicating that the pump was not working properly (2335). Reportedly, the error could not be cleared despite powering down and rebooting the HLM system.The affected roller pump was then replaced with a loan unit. The event occurred during priming.There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50814442 ↗MalfunctionSep 16, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50785633 ↗MalfunctionAug 4, 2026Unstable
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50125986 ↗MalfunctionMar 30, 2026Pumping Stopped
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A SUCKER PUMP OF AN ESSENZ HEART-LUNG MACHINE (HLM) WAS NOT WORKING. THE ISSUE OCCURRED DURING PROCEDURE. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24712222 ↗MalfunctionMar 30, 2026Pumping Stopped
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT SUCKER AND VENT PUMPS OF AN ESSENZ HEART-LUNG MACHINE (HLM) WERE NOT WORKING. THE ISSUE OCCURRED DURING PROCEDURE. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24712216 ↗MalfunctionMar 25, 2026Difficult to Open or Close
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN ESSENZ HEART-LUNG MACHINE (HLM) VENOUS CLAMP (VC) DISPLAYED THE ERROR MESSAGE "VC DEFECTIVE" (ERROR CODE 2551). THE ISSUE OCCURRED DURING PROCEDURE. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24688631 ↗MalfunctionMar 23, 2026Unstable
LIVANOVA DEUTSCHLAND RECEIVED A REPORT ABOUT MULTIPLE ERROR MESSAGES "VENOUS CLAMP DEFECTIVE" RELATED TO THE AN ESSENZ ELECTRONIC VENOUS CLAMP (EVO), DISPLAYED ON THE RELATIVE ESSENZ HLM CONSOLE. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24664975 ↗MalfunctionMar 19, 2026Low Readings
LIVANOVA DEUTSCHLAND RECEIVED A REPORT OF A ESSENZ BUBBLE SENSOR WHICH COULD NOT PROPERLY FUNCTIONING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24637224 ↗MalfunctionMar 13, 2026Application Program Problem
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT ESSENZ HEART-LUNG MACHINE (HLM) COCKPIT HAS UNEXPECTEDLY REBOOTED DURING THE PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24591224 ↗MalfunctionMar 12, 2026Application Program Problem
LIVANOVA DEUTSCHLAND RECEIVED A REPORT OF AN ESSENZ HEART-LUNG MACHINE COCKPIT REBOOT DURING THE PROCEDURE. DURING THE OCCURRENCE, THE COCKPIT SCREEN TURNED BLACK, BUT IMMEDIATELY IT RETURNED TO THE HOME SCREEN. IT WAS ALSO REPORTED THAT THE CENTRIFUGAL PUMP AND ROLLER PUMPS CONTINUED TO OPERATE WHILE THE SYSTEM REBOOTED ITSELF, AND AFTER CLICKING BACK INTO THE CASE, THE PROCEDURE WAS COMPLETED WITH NO FURTHER ISSUES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24583437 ↗
Other LivaNova Deutschland GmbH products
- Heater-Cooler System2 models · 311 reports · 1 recall
- CP54 models · 96 reports · 1 recall
- S5 Console7 models · 141 reports
- S5 Roller Pump 1502 models · 135 reports
- Electronic Gas Blender3 models · 106 reports
- Electrical Remote-Controlled Tubing Clamp3 models · 85 reports
- S5 Double Roller Pump 852 models · 59 reports
- S5 Mast Roller Pump4 models · 47 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page