ESSENZ Venous clampRecalls & reported problems

LivaNova Deutschland GmbH · Class 2

All models of ESSENZ (8)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This LivaNova ESSENZ venous clamp controls blood flow tubing on a heart-lung machine during cardiopulmonary bypass surgery. The 28 reports mostly describe the clamp being hard to open or close, becoming unstable, or showing error codes and warning messages on the control screen during procedures, sometimes forcing the care team to finish with manual clamps. There are 28 malfunctions, with no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
21
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 04033817903093), received between Oct 5, 2023 and Aug 7, 2026 — classified as 21 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (17 reports).

1 other recall cover devices cleared under the same FDA submission (K232291); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Open or Close17
  • Unstable4
  • Calibration Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions21

Recent adverse event reports

The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2026Difficult to Open or Close

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50169928 ↗
  • MalfunctionMar 25, 2026Difficult to Open or Close

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN ESSENZ HEART-LUNG MACHINE (HLM) VENOUS CLAMP (VC) DISPLAYED THE ERROR MESSAGE "VC DEFECTIVE" (ERROR CODE 2551). THE ISSUE OCCURRED DURING PROCEDURE. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688631 ↗
  • MalfunctionMar 23, 2026Unstable

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT ABOUT MULTIPLE ERROR MESSAGES "VENOUS CLAMP DEFECTIVE" RELATED TO THE AN ESSENZ ELECTRONIC VENOUS CLAMP (EVO), DISPLAYED ON THE RELATIVE ESSENZ HLM CONSOLE. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24664975 ↗
  • MalfunctionMar 9, 2026Unstable

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT ABOUT AN ELECTRONIC ESSENZ VENOUS CLAMP (EVO), WHOSE CONTROL UNIT SHOWED 39% OPENING VALUE WHILE IT WAS CLOSED, DESPITE HAVING BEEN CALIBRATED. ON ANOTHER OCCASION, THE CLAMP INDICATED 1% OPENING VALUE WHILE IT WAS FULLY OPEN. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24544923 ↗
  • MalfunctionFeb 6, 2026Difficult to Open or Close

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT ERROR MESSAGE 2551 'VENOUS CLAMP DEFECTIVE' CONCERNING AN ESSENZ VENOUS CLAMP APPEARED ON THE SCREEN DURING PRIMING. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24279389 ↗
  • MalfunctionDec 30, 2025Difficult to Open or Close; Difficult to Open or Close

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT THE ESSENZ VENOUS CLAMP WAS NOT WORKING PROPERLY AND ERROR CODE 2551 WAS SHOWED. THE ISSUE OCCURRED DURING PROCEDURE. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23920522 ↗
  • MalfunctionDec 29, 2025Difficult to Open or Close; Difficult to Open or Close

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT OF AN ERROR MESSAGE RELATED TO AN ESSENZ VENOUS CLAMP DISPLAYED ONTO THE UNIT ("VC DEFECTIVE"). SUBSEQUENTLY, THE CLAMP WAS CALIBRATED AND COULD WORK PROPERLY. THE EVENT OCCURRED OUT OF PROCEDURE.THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23904696 ↗
  • MalfunctionDec 5, 2025Difficult to Open or Close; Difficult to Open or Close

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23728166 ↗
  • MalfunctionNov 28, 2025Difficult to Open or Close; Difficult to Open or Close

    SEE INTIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23672106 ↗
  • MalfunctionNov 5, 2025Unstable; Unstable

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23479254 ↗
  • MalfunctionSep 3, 2025Difficult to Open or Close; Difficult to Open or Close

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22950219 ↗
  • MalfunctionJun 20, 2025Difficult to Open or Close; Difficult to Open or Close

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT, DURING A PROCEDURE, THE VENOUS CLAMP MALFUNCTIONED. THE ESSENZ COCKPIT DISPLAYED RED BOX WITH A GEAR AND AN UNLINKED ICON. CUSTOMER HAD TO REMOVE THE TUBING FROM THE CLAMP AND MANUALLY CLAMP. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22297992 ↗
  • MalfunctionMay 9, 2025Difficult to Open or Close; Difficult to Open or Close

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT, DURING A PROCEDURE, IT IS NOT POSSIBLE TO MODIFY THE STATUS OF THE VENOUS CLAMP: WHEN THEY TURN IT ON THEY FEEL THAT THE CLAMP DOES SOMETHING INSIDE MECHANICALLY, BUT THE PISTON DOES NOT MOVE AND REMAINING STILL EVEN AFTER SEVERAL ATTEMPTS TO TURN THE SYSTEM ON AND OFF, WITH ALSO CONSEQUENT CHANGE OF DOOR IN THE POWER-PACK, THE PROBLEM REMAINS. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21992761 ↗
  • MalfunctionMar 25, 2025Difficult to Open or Close; Difficult to Open or Close

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21685595 ↗
  • MalfunctionFeb 11, 2025Difficult to Open or Close; Difficult to Open or Close

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT RELATED TO A VENOUS CLAMP ISSUE DURING PROCEDURE. EVEN IF 20% OPENING WAS REQUESTED, THE CLAMP WAS 100% OPEN. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21353084 ↗
  • MalfunctionJan 10, 2025Difficult to Open or Close; Difficult to Open or Close

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21122423 ↗
  • MalfunctionOct 31, 2024Difficult to Open or Close; Difficult to Open or Close

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT, DURING A PROCEDURE, THE VENOUS CLAMP BEHAVED AS NO LONGER CALIBRATED AND WAS NOT OCCLUDING CORRECTLY: AT 0.9%, THE BLOOD FLOW METER MEASURED MORE THAN 2 LITERS OF VENOUS RETURN. ALARM 156 WAS DISPLAYED ACCORDING TO PERFUSIONIST, THE EVENT HAS ALREADY PREVIOUSLY OCCURRED WITHOUT CAUSING ALARM AS IN THIS CASE. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20578260 ↗
  • MalfunctionSep 3, 2024Unstable; Unstable

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20125732 ↗
  • MalfunctionAug 8, 2024Difficult to Open or Close; Difficult to Open or Close

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19938393 ↗
  • MalfunctionJun 20, 2024Calibration Problem; Difficult to Open or Close; Calibration Problem; Difficult to Open or Close

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN ESSSENZ ELECTRICAL VENOUS OCCLUDER (EVO) BEHAVES ERRATICALLY: IT DOES NOT PROVIDE CONSISTENT CLAMPING AND OCCASIONALLY OPENS OR CLOSES ON IT'S OWN. THE UNIT IS IN USE AT A PERFUSION SCHOOL LABORATORY AND NOT IN A CLINICAL SETTING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19575031 ↗
  • MalfunctionOct 5, 2023Difficult to Open or Close; Difficult to Open or Close

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN ESSENZ ELECTRICAL VENOUS OCCLUDER (EVO) GAVE AN ERROR MESSAGE AT SETUP RELATED TO ITS MOTOR DEFECTIVE. ONCE THE ALARM HAS BEEN CLEARED, THE UNIT CONTINUED TO WORK PROPERLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17876409 ↗

FDA clearance and classification

Cleared through Special 510(k) K232291 on Aug 24, 2023 (decision: Substantially Equivalent). Submitted by Livanova Deutschland, GmbH as “Essenz HLM, Essenz ILBM”. The FDA review took 23 days. FDA record ↗

The FDA regulates product code DTQ (console, heart-lung machine, cardiopulmonary bypass) under 21 CFR 870.4220 as a Class II device. 21 CFR 870.4220 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K232291. One submission often covers many models, so check the product and lot details in each record.

  • Z-0571-2025 · Oct 18, 2024 · Open, Classified
    Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software

Common questions about the ESSENZ Venous clamp

Has the ESSENZ Venous clamp been recalled?

No FDA recall record in our data names this exact device identifier (DI 04033817903093), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the ESSENZ Venous clamp?

The FDA's MAUDE database lists 21 adverse event reports for this device, received between Oct 5, 2023 and Aug 7, 2026. In those reports, the most common specific device problems are “Difficult to Open or Close” (17), “Unstable” (4) and “Calibration Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ESSENZ Venous clamp?

None of the reports are classified as deaths, 0 as injuries and 21 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ESSENZ Venous clamp FDA approved?

It was cleared through Special 510(k) K232291 on Aug 24, 2023. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the ESSENZ Venous clamp, and is it still sold?

The FDA's device database lists LivaNova Deutschland GmbH as the company responsible for it (brand ESSENZ, model Venous clamp). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices

All CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page