Corox OTW 85-BPRecalls & reported problems
BIOTRONIK SE & Co. KG · Class 3 · Implant
All models of Corox (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 04035479093549), received between Mar 17, 2019 and Jan 13, 2026 — classified as 3 injuries, 16 malfunctions.
The most frequently reported specific device problem is “High Capture Threshold” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- High Capture Threshold4
- Device Dislodged or Dislocated3
- Failure to Capture3
- High impedance3
- Intermittent Capture2
- Ambient Noise Problem1
- Break1
Patient problems
- No Known Impact Or Consequence To Patient8
- No Clinical Signs, Symptoms or Conditions6
- Unspecified Infection3
- Undesired Nerve Stimulation2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2026Failure to Capture; High impedance
THIS LEAD WAS CAPPED DUE TO HIGH IMPEDANCE AND NO CAPTURE THRESHOLD. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24058541 ↗InjuryOct 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS LEAD WAS PARTIALLY EXPLANTED ON (B)(6) 2023 DUE TO INFECTION, SURGEON WAS UNABLE TO REMOVE DISTAL END OUT OF FAR END OF CS BRANCH DURING OPEN HEART SURGERY FIVE DAYS AFTER ICD REMOVAL. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: Device Embedded In Tissue or Plaque; Unspecified Infection
MAUDE report 17979790 ↗MalfunctionOct 5, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS LEAD WAS CAPPED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17876830 ↗MalfunctionMar 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS LV LEAD WAS EXPLANTED DUE TO PHRENIC NERVE STIMULATION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.
Patient problems: Undesired Nerve Stimulation
MAUDE report 13637494 ↗MalfunctionJun 25, 2021Intermittent Capture; Intermittent Capture
LEAD CAPPED DUE TO INTERMITTENT CAPTURE AT MAX OUTPUT. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12065970 ↗MalfunctionMay 18, 2021Extracardiac stimulation; High Capture Threshold; Extracardiac stimulation; High Capture Threshold
THIS LV LEAD WAS EXPLANTED DUE TO HIGH THRESHOLDS AND POCKET STIMULATION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: Undesired Nerve Stimulation
MAUDE report 11842056 ↗MalfunctionApr 15, 2021Failure to Capture; Over-Sensing; Failure to Capture; Over-Sensing
LEAD REMAINS IMPLANTED. REPRESENTATIVE REPORTED OVERSENSING ON THE LV CHANNEL. LV IEGM IS NOT VISIBLE ON TRANSMITTED RECORDINGS, HOWEVER THE SECOND LVS EVENT APPEARS TO LINE UP WITH THE T-WAVE ON THE FAR-FIELD CHANNEL. LOSS OF CAPTURE WAS ALSO REPORTED. LEAD TO BE EVALUATED FURTHER. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11675803 ↗InjuryDec 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS SYSTEM WAS EXPLANTED DUE TO SUSPECTED INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: Unspecified Infection
MAUDE report 11091954 ↗MalfunctionOct 16, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
AT ERI DEVICE CHANGE OUT IT WAS NOTED THAT THE LV LEAD HAD DISLODGED INTO THE RA. THIS WAS NOTED ON X-RAYS DATED 2017 (UNAVAILABLE). PHYSICIAN CUT LEAD AND TIED OFF END. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10690441 ↗MalfunctionOct 1, 2020Ambient Noise Problem; Ambient Noise Problem
THIS LEAD WAS CAPPED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10612308 ↗MalfunctionJun 2, 2020Pacing Problem; Material Integrity Problem; Pacing Problem; Material Integrity Problem
IT WAS REPORTED THAT INTRINSIC LV EVENTS FALLING INTO A REFRACTORY PERIOD FOLLOWED BY PACING ON THE LV CHANNEL WERE OBSERVED. PATIENT LATER NEEDED RV LEAD REVISION; IN THE COURSE OF LASER EXTRACTION OF RV LEAD, THIS LEAD WAS DAMAGED AND ULTIMATELY EXPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10112845 ↗MalfunctionApr 13, 2020High Capture Threshold; High Capture Threshold
THIS LEAD WAS CAPPED AND REPLACED DUE TO HIGH THRESHOLDS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9955036 ↗MalfunctionJan 22, 2020Failure to Capture; Device Sensing Problem; Failure to Capture; Device Sensing Problem
THIS LEAD WAS CAPPED AND REPLACED DUE TO SENSING ISSUES AND LOSS OF CAPTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9617916 ↗InjuryJan 16, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. NEW SYSTEM WAS IMPLANTED (B)(6) 2020. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: Unspecified Infection
MAUDE report 9599280 ↗MalfunctionJan 6, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS LEAD WAS EXPLANTED AND REPLACED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9555763 ↗MalfunctionAug 14, 2019High impedance; High impedance
THIS LEAD WAS CAPPED DUE TO HIGH IMPEDANCE OVER 3000OHMS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8895470 ↗MalfunctionJul 11, 2019Intermittent Capture; High Capture Threshold; Intermittent Capture; High Capture Threshold
THIS LEAD WAS EXPLANTED AND REPLACED DUE TO HIGH THRESHOLDS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8781186 ↗MalfunctionApr 16, 2019Break; High Capture Threshold; Break; High Capture Threshold
DURING A GEN CHANGE LV LEAD INSULATION WAS FOUND TO HAVE A BREACH NEAR THE PIN. LEAD WAS CAPPED AND REMAINS IMPLANTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8520861 ↗MalfunctionMar 17, 2019High impedance; High impedance
THIS LEAD WAS CAPPED AND REPLACED DUE TO IMPEDANCES OF GREATER THAN 3000 OHMS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8427300 ↗
Other Drug eluting permanent left ventricular (lv) pacemaker electrode devices
- Attain Stability™ Quad MRI SureScan™ 479888Medtronic, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode28 deaths2,213 reports
- Attain Performa™ Straight MRI SureScan™ 439888Medtronic, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode26 deaths1,595 reports
- Sentus ProMRI OTW QP L-85/49 408719BIOTRONIK SE & Co. KG · Drug eluting permanent left ventricular (lv) pacemaker electrode7 deaths1,385 reports
- Quartet™ 1457Q-86St. Jude Medical, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode21 deaths1,061 reports
- Quickflex™ 1258TSt. Jude Medical, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode26 deaths996 reports
- ATTAIN® BIPOLAR OTW 419488Medtronic, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode15 deaths730 reports
- Attain Ability® Plus 429688Medtronic, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode11 deaths529 reports
- Sentus ProMRI OTW QP S-85/49 406082BIOTRONIK SE & Co. KG · Drug eluting permanent left ventricular (lv) pacemaker electrode1 death511 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.