Corox OTW-L 85-BPRecalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

All models of Corox (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Corox OTW-L85-BP is an implanted lead that delivers drug-eluting therapy to the heart's left ventricle. The 73 reports mostly describe the lead not working as expected — the wire moving out of position, failing to pace the heart, needing unusually high energy to capture a beat, and one case where the lead fractured and a piece was left in the patient during removal. Reports also mention infections and unwanted nerve or muscle stimulation. There were 27 injuries and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
73
Deaths reported
0
Injuries reported
27
Recalls naming it
0

Summary

The FDA's MAUDE database contains 73 adverse event reports that list this exact device (DI 04035479112172), received between Jul 2, 2018 and Aug 19, 2025 — classified as 27 injuries, 46 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (16 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem24
  • High Capture Threshold16
  • Device Dislodged or Dislocated11
  • Failure to Capture10
  • High impedance6
  • Extracardiac stimulation5
  • Fracture3
  • Ambient Noise Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • No Known Impact Or Consequence To Patient21
  • Unspecified Infection19
  • Undesired Nerve Stimulation7
  • Muscle Stimulation4
  • Cardiac Perforation1
  • Device Embedded In Tissue or Plaque1
  • Pocket Erosion1

Recent adverse event reports

The 25 most recent of 73 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22844225 ↗
  • MalfunctionAug 7, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22742549 ↗
  • MalfunctionJul 23, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS CAPPED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22578001 ↗
  • InjuryMar 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 18977721 ↗
  • MalfunctionMar 18, 2024High Capture Threshold; High Capture Threshold

    THIS LEAD SHOWS HIGH THRESHOLDS. PATIENT WAS BROUGHT IN TO ATTEMPT PLACING A LBB LEAD, BUT DUE TO SUBCLAVIAN STENOSIS, CASE WAS ABORTED. CONSIDERING PATIENT AGE AND POOR STATUS, PHYSICIAN DECIDED TO CHANGE OUT THE DEVICE ONLY. LEAD REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18920420 ↗
  • MalfunctionSep 27, 2023Failure to Capture; High impedance; Failure to Capture; High impedance

    THIS LV LEAD WAS CAPPED DUE TO NO CAPTURE AT MAXIMUM OUTPUT AND HIGH IMPEDANCE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17828897 ↗
  • MalfunctionSep 27, 2023Failure to Capture; Device Sensing Problem; Failure to Capture; Device Sensing Problem

    THIS LV LEAD WAS EXPLANTED DUE TO LOSS OF CAPTURE AND SENSING. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17825430 ↗
  • InjuryFeb 6, 2023Fracture; Fracture

    THIS LV LEAD WAS IN THE PROCESS OF BEING EXPLANTED FOR A QUADRIPOLAR LEAD IN ORDER TO ELIMINATE DIAPHRAGMATIC STIMULATION WHEN IT FRACTURED. DURING THIS, THE PATIENT STARTED TO BLEED AND PART OF THE LEAD REMAINED IN THE PATIENT. THE PATIENT WAS TAKEN TO CARDIAC SURGERY FOR THE BLEED AND THERE MAY HAVE BEEN A PERFORATION. THE SURGERY WAS SUCCESSFUL. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Cardiac Perforation; Device Embedded In Tissue or Plaque; Undesired Nerve Stimulation

    MAUDE report 16308258 ↗
  • MalfunctionDec 28, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    LEAD EXPLANTED DUE TO IT HAD CHRONICALLY DISLODGED INTO THE SVC. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16065003 ↗
  • MalfunctionNov 29, 2022Device Sensing Problem; Device Dislodged or Dislocated; High Capture Threshold; Device Sensing Problem; Device Dislodged or Dislocated; High Capture Threshold

    LEAD WAS CAPPED DUE TO INCREASED THRESHOLDS WITH DECREASING SENSING CAUSED BY DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15877089 ↗
  • MalfunctionNov 3, 2022Extracardiac stimulation; Extracardiac stimulation

    THIS LEAD WAS EXPLANTED DUE TO PHRENIC STIMULATION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 15723069 ↗
  • MalfunctionNov 2, 2022Failure to Capture; Failure to Capture

    LEAD WAS CAPPED DUE TO NO CAPTURE AT MAX OUTPUT. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15713794 ↗
  • InjurySep 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 15417988 ↗
  • MalfunctionSep 8, 2022High Capture Threshold; High Capture Threshold

    THIS LEAD WAS EXPLANTED DUE TO HIGH PACING THRESHOLDS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15385189 ↗
  • InjuryJun 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. DATE UNKNOWN. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 14706936 ↗
  • InjuryJun 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 14703836 ↗
  • MalfunctionApr 29, 2022Failure to Capture; Failure to Capture

    LEAD WAS CAPPED DUE TO LOSS OF CAPTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14254937 ↗
  • InjuryApr 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 14172246 ↗
  • MalfunctionMar 10, 2022Failure to Capture; Failure to Capture

    THIS LEAD WAS EXPLANTED DUE TO LOSS OF CAPTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13726781 ↗
  • MalfunctionMar 8, 2022High Capture Threshold; High Capture Threshold

    LEAD EXPLANTED DUE TO HIGH THRESHOLDS. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13692936 ↗
  • MalfunctionMar 1, 2022Malposition of Device; Malposition of Device

    THIS LV LEAD WAS CAPPED AND REPLACED BECAUSE THE PHYSICIAN FELT LIKE IT WAS IN A BAD POSITION. THE PHYSICIAN IMPLANTED A NEW IS4 LV LEAD. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13637490 ↗
  • MalfunctionJan 19, 2022Failure to Capture; High impedance; Failure to Capture; High impedance

    LEAD WAS SHOWING HIGH IMPEDANCES OF GREATER THAN 3000 OHMS VIA HOME MONITORING IN (B)(6) 2018. SINCE THEN IT HAS EXHIBITED INTERMITTENT NON-CAPTURE. THE LEAD WAS EXPLANTED AND REPLACED ON (B)(6) 2022. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13288748 ↗
  • MalfunctionJan 12, 2022Failure to Capture; Extracardiac stimulation; Failure to Capture; Extracardiac stimulation

    LOSS OF CAPTURE SEEN IN PERIODIC IEGM. PATIENT WILL BE BROUGHT IN TO EVALUATE THRESHOLDS. DIAPHRAGMATIC STIMULATION WAS SEEN DURING PREVIOUS FOLLOW UPS BUT TROUBLESHOOTING WILL BE DONE IN OFFICE. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 13239999 ↗
  • InjuryJan 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    EXPLANTED DUE TO PATIENT SYSTEMIC INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 13226313 ↗
  • MalfunctionJan 4, 2022Malposition of Device; High Capture Threshold; Malposition of Device; High Capture Threshold

    THIS LEAD WAS CAPPED AND REPLACED DUE TO HIGH THRESHOLDS AND A BAD POSITION. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13154742 ↗

Other Drug eluting permanent left ventricular (lv) pacemaker electrode devices

All Drug eluting permanent left ventricular (lv) pacemaker electrode devices →