Sentus ProMRI OTW QP L-75/49 408718Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Sentus ProMRI is a pacemaker lead placed in the left side of the heart. Many of the 134 reports describe the lead moving out of place, infections, and the lead failing to pace the heart, and some describe the lead breaking or being removed. There are no deaths and 55 injuries, and reports rose from 7 in the previous 12 months to 17 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
135
Deaths reported
0
Injuries reported
56
Recalls naming it
0

Summary

The FDA's MAUDE database contains 135 adverse event reports that list this exact device (DI 04035479148942), received between Jun 18, 2018 and Sep 3, 2026 — classified as 56 injuries, 79 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (55 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 4 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 4 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated55
  • Adverse Event Without Identified Device or Use Problem40
  • Failure to Capture17
  • High impedance12
  • High Capture Threshold9
  • Fracture6
  • Ambient Noise Problem3
  • Break3

Patient problems

  • No Clinical Signs, Symptoms or Conditions64
  • Unspecified Infection33
  • No Known Impact Or Consequence To Patient21
  • Pocket Erosion3
  • Twiddlers Syndrome3
  • Cardiac Perforation2
  • Fall2
  • Muscle Stimulation2

Recent adverse event reports

The 25 most recent of 135 reports, in the reporter's own words as published by the FDA.

  • InjurySep 3, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to Vegetation noted in Echocardiogram of RV lead. Should additional information be received, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50590296 ↗
  • InjuryAug 26, 2026Break

    Lead was capped due to it broke. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456529 ↗
  • InjuryAug 25, 2026Device Dislodged or Dislocated

    This lead was explanted due to dislodgement. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50438156 ↗
  • InjuryAug 13, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infection. Should additional information be received, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50234036 ↗
  • InjuryAug 13, 2026Device Dislodged or Dislocated

    This LV lead was explanted due to dislodgement. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50230300 ↗
  • InjuryJul 6, 2026High impedance

    THIS LEAD WAS EXPLANTED DUE TO HIGH IMPEDANCE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25719533 ↗
  • InjuryMay 22, 2026Failure to Capture; Device Dislodged or Dislocated

    THIS LEAD WAS CAPPED ON (B)(6) 2026 DUE TO DISLODGEMENT RESULTING IN NON-CAPTURE OF ALL LV VECTORS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25255054 ↗
  • InjuryApr 28, 2026Fracture

    THIS LEAD WAS CAPPED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023720 ↗
  • InjuryMar 30, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24719402 ↗
  • InjuryFeb 26, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24464069 ↗
  • InjuryFeb 26, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24462502 ↗
  • InjuryFeb 24, 2026Failure to Capture; Device Sensing Problem

    THIS LV LEAD WAS CAPPED DUE TO LOSS OF SENSING/CAPTURE (STEADY DECLINE OVER TIME). NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24429075 ↗
  • InjuryFeb 9, 2026Adverse Event Without Identified Device or Use Problem

    DURING IMPLANT ATTEMPT THIS LEAD DISSECTED THE CS. IMPLANT CONTINUED WITH A NEW LEAD IN A DIFFERENT LOCATION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Vascular Dissection

    MAUDE report 24299888 ↗
  • InjuryJan 26, 2026Device Dislodged or Dislocated

    THIS LV LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24181336 ↗
  • InjuryJan 19, 2026Failure to Capture

    THIS LEAD WAS EXPLANTED DUE TO NO CAPTURE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24108755 ↗
  • MalfunctionDec 31, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23935960 ↗
  • MalfunctionDec 22, 2025Failure to Capture; Failure to Capture

    THIS LV LEAD WAS EXPLANTED DUE TO FAILURE TO CAPTURE IN ANY VECTOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23873534 ↗
  • InjuryNov 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 23658246 ↗
  • InjuryJun 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22364605 ↗
  • MalfunctionJun 2, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22128549 ↗
  • MalfunctionMay 28, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22095551 ↗
  • MalfunctionMay 16, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    LEAD WAS EXPLANTED BECAUSE IT DISLODGED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035875 ↗
  • MalfunctionMar 14, 2025Break; Break

    DURING DEVICE CHANGE OUT, THIS LEAD WAS CAPPED DUE TO IT BROKE APART AT THE CONNECTOR PIN. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21605605 ↗
  • InjuryFeb 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 21327351 ↗
  • MalfunctionJan 21, 2025High impedance; High impedance

    THIS LEAD WAS EXPLANTED DUE TO HIGH IMPEDANCE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21203739 ↗

Other Drug eluting permanent left ventricular (lv) pacemaker electrode devices