Acticor 7 HF-T DF4 IS-1 ProMRI 429523Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BIOTRONIK Acticor 7 HF-T is an implanted defibrillator with heart resynchronization. The 70 reports mostly describe the unit being removed (explanted) because of infection at the implant site, along with pocket erosion and a hematoma that wouldn't heal; there are 53 injuries and no deaths. Some reports also note the device reaching end-of-service status sooner than expected or failing to be read during a check. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
73
Deaths reported
0
Injuries reported
56
Recalls naming it
0

Summary

The FDA's MAUDE database contains 73 adverse event reports that list this exact device (DI 04035479156688), received between Jan 10, 2020 and Sep 16, 2026 — classified as 56 injuries, 17 malfunctions.

The most frequently reported specific device problem is “Premature Elective Replacement Indicator” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 4 reportsApr 2024: 2 reportsMay 2024: 3 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 5 reportsAug 2025: 3 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem53
  • Premature Elective Replacement Indicator9
  • Premature End-of-Life Indicator3
  • Ambient Noise Problem2
  • Failure to Interrogate2
  • Inappropriate/Inadequate Shock/Stimulation2
  • Device Sensing Problem1
  • Failure to Capture1

Patient problems

  • Unspecified Infection41
  • No Clinical Signs, Symptoms or Conditions18
  • Hematoma4
  • Pocket Erosion4
  • Bradycardia1
  • Discomfort1
  • Drug Resistant Bacterial Infection1
  • Fever1

Recent adverse event reports

The 25 most recent of 73 reports, in the reporter's own words as published by the FDA.

  • InjurySep 16, 2026Premature End-of-Life Indicator

    This device was explanted due to EOS after approximately 65 appropriate shocks since (b)(6) 2026 (patient was out of the country). Should additional information become available, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50782711 ↗
  • InjurySep 2, 2026Adverse Event Without Identified Device or Use Problem

    After successful implantation of this device, a Medtronic Spere 9 was used to ablate the AV node via RF. The first burn was successful. The second burn resulted in inducing the patient into VF. The patient was restored to sinus rhythm after external defibrillation. Device remains implanted.

    Patient problems: Ventricular Fibrillation

    MAUDE report 50557235 ↗
  • InjuryAug 17, 2026Adverse Event Without Identified Device or Use Problem

    System explanted due to pocket erosion. Should additional information be received, this file will be updated.

    Patient problems: Pocket Erosion

    MAUDE report 50278178 ↗
  • InjuryAug 11, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to non-healing hematoma. Should additional information become available, this file will be updated.

    Patient problems: Hematoma

    MAUDE report 50213809 ↗
  • InjuryAug 4, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to lead vegetation. Should additional information become available, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50134482 ↗
  • InjuryJun 10, 2026Adverse Event Without Identified Device or Use Problem

    THIS DEVICE WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25457457 ↗
  • InjuryApr 27, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25000675 ↗
  • MalfunctionApr 13, 2026Premature End-of-Life Indicator

    DEVICE WAS PUT INTO AUTO MRI MODE FOR A SCAN AND THE SESSION WAS ENDED WITHOUT ISSUE. TWO DAYS LATER, A FACILITY CLINICIAN INTERROGATED THE DEVICE AND REPORTED THE DEVICE IS EOS. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24860578 ↗
  • InjuryMar 11, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24573469 ↗
  • InjuryJan 22, 2026Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24151014 ↗
  • InjuryJan 21, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24133678 ↗
  • InjuryJan 12, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24042713 ↗
  • InjuryDec 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    POCKET REVISION/EVACUATION DUE TO HEMATOMA. DEVICE REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Hematoma

    MAUDE report 23910895 ↗
  • InjuryDec 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 23694332 ↗
  • MalfunctionNov 18, 2025Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE WAS EXPLANTED DUE TO ERI WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23586483 ↗
  • InjuryNov 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO POCKET EROSION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Pocket Erosion

    MAUDE report 23568860 ↗
  • InjurySep 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22965753 ↗
  • InjuryAug 25, 2025Migration or Expulsion of Device; Migration or Expulsion of Device

    DEVICE CHANGED AT TIME OF POCKET REVISION DUE TO MIGRATION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22882988 ↗
  • InjuryAug 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS DEVICE WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22788065 ↗
  • MalfunctionAug 11, 2025Ambient Noise Problem; Ambient Noise Problem

    DEVICE EXHIBITED NOISE ON RV LEAD AFTER NEW RV LEAD WAS IMPLANTED. PHYSICIAN WAS IN A HURRY AND ADEQUATE TIME FOR TROUBLE-SHOOTING WAS NOT PRACTICAL. PHYSICIAN ASKED FOR NEW DEVICE, SO THIS ONE WAS EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22771097 ↗
  • InjuryJul 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22670123 ↗
  • MalfunctionJul 30, 2025Premature End-of-Life Indicator; Inappropriate/Inadequate Shock/Stimulation; Premature End-of-Life Indicator; Inappropriate/Inadequate Shock/Stimulation

    THIS DEVICE WAS EXPLANTED DUE TO EOS, POSSIBLY DUE TO 40 PLUS INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22669288 ↗
  • InjuryJul 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22666427 ↗
  • MalfunctionJul 22, 2025Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE SHOWS ERI WITH UNEXPECTED BATTERY BEHAVIOR. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22568654 ↗
  • MalfunctionJul 17, 2025Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE WAS EXPLANTED DUE TO ERI WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22538524 ↗

Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices

All Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices →