Rivacor 5 HF-T QP DF4 IS4 ProMRI 429561Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
1
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 04035479156879), received between Jan 10, 2020 and Jun 8, 2026 — classified as 1 death, 11 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Inappropriate/Inadequate Shock/Stimulation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem9
  • Inappropriate/Inadequate Shock/Stimulation2
  • Loose or Intermittent Connection2
  • Ambient Noise Problem1
  • Failure to Convert Rhythm1
  • Failure to Interrogate1
  • High impedance1

Patient problems

  • Unspecified Infection5
  • No Clinical Signs, Symptoms or Conditions2
  • Hematoma1
  • Pneumothorax1
  • Shock1
  • Shock from Patient Lead(s)1
  • Thrombosis/Thrombus1
  • Ventricular Fibrillation1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 8, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO SUSPECTED INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25427908 ↗
  • InjuryJun 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED POSTOPERATIVELY THAT A WOUND DEHISCENCE WAS OBSERVED. THE PATIENT WAS HOSPITALIZED. WOUND TREATMENT AND DISINFECTION WERE DONE. THE ICD LEAD WAS RELOCATED BEHIND THE DEVICE. DEVICE REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Wound Dehiscence

    MAUDE report 14881107 ↗
  • InjuryJun 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    TWO WEEKS AFTER IMPLANTATION THE PATIENT SUFFERED FROM AN INFECTION AND WAS HOSPITALIZED. THE PATIENT WAS GIVEN I.V. ANTIBIOTIC TREATMENT AND THE CRT-D WAS EXTRACTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 14784141 ↗
  • InjuryJun 7, 2022Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    AFTER AN IMPLANTATION PERIOD OF APPROX. 12 MONTHS, IT WAS OBSERVED THAT THE PATIENT RECEIVED AN INAPPROPRIATE SHOCK FROM THE CRT. DEVICE REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Shock

    MAUDE report 14620995 ↗
  • MalfunctionMay 3, 2022High impedance; Loose or Intermittent Connection; Ambient Noise Problem; High impedance; Loose or Intermittent Connection; Ambient Noise Problem

    IMPEDANCE HIGHER THAN 3000 OHM AND NOISE DETECTION WERE SEEN. AN RV LEAD ISSUE HAS BEEN RULED OUT. A FAILURE OF THE DEVICE HEADER WAS REPORTED. DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14271209 ↗
  • InjuryMar 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT SUFFERED FROM SYSTEMIC FEVER AND INFECTION IN THE AREA OF PACEMAKER. (B)(6)WAS FOUND. CRT-D AND ELECTRODES WERE REMOVED.

    Patient problems: Unspecified Infection

    MAUDE report 13953838 ↗
  • InjuryFeb 16, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT SUFFERED FROM POCKET INFECTION. THE PATIENT WAS HOSPITALIZED. SYSTEM WAS EXPLANTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 13541937 ↗
  • InjurySep 14, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AFTER AN IMPLANTATION PERIOD OF APPROXIMATELY 2 MONTHS, IT WAS REPORTED THAT THE PATIENT WAS DIAGNOSED WITH MULTI-STAGE THROMBOSIS OF THE SUBCLAVIAN VEIN. ORAL ANTICOAGULATION THERAPY WAS RECOMMENDED FOR AT LEAST 3-6 MONTHS.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 12467688 ↗
  • DeathAug 9, 2021Failure to Convert Rhythm; Failure to Convert Rhythm

    27 DAYS AFTER IMPLANTATION IT WAS REPORTED THAT THE PATIENT DIED AFTER AN ELECTRICAL STORM AND UNSUCCESSFUL REANIMATION.

    Patient problems: Ventricular Fibrillation

    MAUDE report 12294048 ↗
  • InjuryMar 24, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT SUFFERED FROM A LARGE HEMATOMA AFTER CRT IMPLANTATION. THE INVESTIGATOR ASSESSED THIS EVENT AS POSSIBLE RELATED TO THE PATIENTS MEDICAL HISTORY AND CONCOMITANT MEDICATION WITH APIXABAN 2.5MG. THE PATIENT WAS HOSPITALIZED. ANALGESIC AS NEEDED AND NOVALGIN PILLS WERE GIVEN.

    Patient problems: Hematoma

    MAUDE report 11565767 ↗
  • MalfunctionMar 22, 2021Failure to Interrogate; Failure to Interrogate

    IT WAS REPORTED THAT THIS DEVICE COULD NOT BE INTERROGATED. THE DEVICE WAS EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11547449 ↗
  • InjuryOct 16, 2020Loose or Intermittent Connection; Inappropriate/Inadequate Shock/Stimulation; Loose or Intermittent Connection; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT APPROX. 15 DAYS AFTER THE IMPLANTATION THERE MAY BE AN INAPPROPRIATE SHOCK DUE TO A CRT-D FAILURE. FURTHERMORE, PATIENT WAS DIAGNOSED WITH HYPERKALEMIA. THE PATIENT WAS HOSPITALIZED AND PULSE GENERATOR REPROGRAMMING WAS DONE. FOR DIAGNOSTIC REASONS CHEST X-RAY AND BLOOD TESTS WERE DONE. PATIENT WAS TREATED WITH ANTIBIOTICS AND TREATMENT WITH ALDACTONE WAS STOPPED. PATIENT GOT A SURGICAL INTERVENTION (SECURING SCREW CONNECTION OF RV LEAD TO CRTD).

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 10693184 ↗
  • InjuryJul 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT SUFFERED FROM A POST-INTERVENTIONAL PNEUMOTHORAX. THE PATIENT WAS HOSPITALIZED AND A SUCTION DRAIN WAS IMPLANTED IN THE MONALDI POSITION LEFT. FOR DIAGNOSTICS AN X-RAY WAS DONE.

    Patient problems: Pneumothorax

    MAUDE report 10227365 ↗
  • InjuryJan 10, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A REVISION OF THE CRT SYSTEM WAS PERFORMED. HOWEVER, THE TISSUE WAS UNUSUALLY SOFT AND ALTERED, GIVING THE CLINICAL IMPRESSION OF A POSSIBLE CHRONIC WOUND INFECTION. THE MICROBIOLOGICAL TESTING OF THE SMEAR DID NOT SHOW ANY BACTERIA OR FUNGUS. THE REVISION WAS COMPLICATED BY EXCESSIVE BLEEDING. THE INVESTIGATOR ASSESSED THIS AS A POSSIBLE CHRONIC PACEMAKER WOUND INFECTION. WOUND DEBRIDEMENT WAS PERFORMED AND AN INTRAOPERATIVE SMEAR WAS TAKEN. COAGULATION WAS PERFORMED TO STOP BLEEDING. PATIENT WAS PRESCRIBED A PROLONGED COURSE OF ANTIBIOTICS AFTER CRT REVISION.

    Patient problems: Unspecified Infection

    MAUDE report 9574559 ↗

Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices

All Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices →