Skip to content

AESCULAP PL569TRecalls & reported problems

Aesculap AG · LIGATING CLIPS M/L 12/BOX · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04038653181353), received between Apr 23, 2021 and Mar 8, 2023 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Failure to Eject” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Failure to Eject2
  • Connection Problem1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 8, 2023Connection Problem; Connection Problem

    IT WAS REPORTED TO AESCULAP AG THAT A LIGATING CLIPS M/L 12/BOX (PART # PL569T) WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE PATIENT VERBALLY CONTACTED TORONTO HUMANE SOCIETY (THS) WITH REGARDS TO PRODUCT. IT WAS POTENTIALLY USED DURING A PROCEDURE IN (B)(6) 2015. PATIENT REPORTED FEELING UNWELL SINCE THIS TIME. PATIENT HAS REQUESTED FURTHER INFORMATION WITH REGARDS TO THE MANUFACTURER/COMPOSITION OF PRODUCT IN ORDER TO PROCEED WITH FURTHER TESTING THROUGH HIS/HER PHYSICIAN. THS HAS NOT RECEIVED A COMPLAINT FROM THE HOSPITAL WITH REGARDS THE PRODUCT. PATIENT HARM WAS UNKNOWN. ADDITIONAL INFORMATION WAS NOT PROVIDED NOR AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 16507189 ↗
  • InjuryFeb 11, 2022Failure to Eject; Failure to Eject

    IT WAS REPORTED TO AESCULAP AG THAT A LIGATING CLIPS M/L 12/BOX (PART # PL569T) WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE FIRST TWO (2) TO THREE (3) CLIPS WORKED FINE, BUT THEN THE MAGAZINE WAS PUSHED AWAY FROM THE CLIP BAR (PART # PL538R) AT THE NEXT ACTIVATION. THIS HAPPENED INSIDE THE PATIENT'S ABDOMEN. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE PATIENT WAS NOT INJURED. HOWEVER, AN ADDITIONAL MEDICAL INTERVENTION WAS NECESSARY. ADDITIONAL INFORMATION WAS NOT PROVIDED NOR AVAILABLE / WAS NOT AVAILABLE. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4). INVOLVED COMPONENTS: PL538R - SHAFT COMPL.D:10MM L:260MM - LOT UNKNOWN. PL510R - HANDLE F/PL506R & PL508R - LOT UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13517269 ↗
  • MalfunctionNov 19, 2021Material Separation; Material Separation

    IT WAS REPORTED TO AESCULAP AG THAT A LIGATING CLIPS M/L 12/BOX (PART # PL569T) WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE DEVICE WAS COMPRESSED, THE TOP BROKE OFF AND ENDED UP IN THE PATIENT. THE FRAGMENT WAS RETRIEVED AND THEN THE PROCEDURE WAS CONTINUED AND SUCCESSFULLY COMPLETED. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12844614 ↗
  • InjuryApr 23, 2021Failure to Eject; Failure to Eject

    IT WAS REPORTED TO AESCULAP AG THAT A LIGATING CLIPS M/L 12/BOX (PART # PL569T) WAS USED DURING AN UNKNOWN PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CLIPS FELL IN SITU. REPORTEDLYM, AFTER THE THIRD OR FOURTH CLIP HAD BEEN PLACED, THE COMPLETE MAGAZINE SUDDENLY DETACHED INTO THE ABDOMINAL CAVITY AND HAD TO BE RETRIEVED. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4). INVOLVED COMPONENTS: PL510R - HANDLE F/PL506R & PL508R - LOT UNKNOWN. PL536R - SHAFT COMPL.D:10MM L:370MM - LOT 62424289.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11720789 ↗

FDA clearance and classification

The FDA regulates product code FZP (clip, implantable) under 21 CFR 878.4300 as a Class II device. 21 CFR 878.4300 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the AESCULAP PL569T

Has the AESCULAP PL569T been recalled?

No FDA recall record in our data names this exact device identifier (DI 04038653181353), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the AESCULAP PL569T?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Apr 23, 2021 and Mar 8, 2023. In those reports, the most common specific device problems are “Failure to Eject” (2), “Connection Problem” (1) and “Material Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AESCULAP PL569T?

None of the reports are classified as deaths, 2 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AESCULAP PL569T MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the AESCULAP PL569T, and is it still sold?

The FDA's device database lists Aesculap AG as the company responsible for it (brand AESCULAP, model PL569T). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CLIP, IMPLANTABLE devices

All CLIP, IMPLANTABLE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page