Skip to content

activL® SW990KRecalls & reported problems

Aesculap AG · ACTIV L INF.PLATE SIZE L 0°/SPIKES · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04038653425884), received between Jun 22, 2022 and Jan 23, 2023 — classified as 4 injuries.

The most frequently reported specific device problem is “Migration” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Migration2
  • Patient Device Interaction Problem2
  • Migration or Expulsion of Device1

Patient problems

  • Failure of Implant2
  • Peritonitis2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 23, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    UPDATE: PER THE SURGEON, THE PATIENT IS DOING VERY WELL POSTOPERATIVELY. CERTAIN CULTURES TAKEN WERE POSITIVE FOR STREPTOCOCCUS PYOGENES. SEVERAL SITES WERE TESTED INCLUDING THE DEVICE ONCE IT WAS REMOVED. IT WAS NOTED THAT THE PATIENT MAY BE HAVING SURGERY AGAIN TO REINFORCE THE ANTERIOR CAGES AT BOTH LEVELS. ASSOCIATED MEDWATCH-REPORTS: 9610612-2022-00394 (INTERNAL AESCULAP AG REF. NO. (B)(4) - SW990K). 9610612-2022-00395 (INTERNAL AESCULAP AG REF. NO. (B)(4) - SW991K). 9610612-2022-00396 (INTERNAL AESCULAP AG REF. NO. (B)(4) - SW966).

    Patient problems: Peritonitis

    MAUDE report 16230498 ↗
  • InjuryJan 23, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED TO AESCULAP INC. THAT AN ACTIV L INF.PLATE SIZE L 0°/SPIKES (PART# SW990K) WAS IMPLANTED DURING A HYBRID AT L4/5 AND FUSION DEVICE L5/S1 PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE PATIENT UNDERWENT A REOPERATION AND REMOVAL ON DECEMBER 27 2022 AS PATIENT HAS A RETROPERITONEAL ABSCESS AT L4/5-L5/S1. THE SURGEON STATED THAT THE PATIENT WOULD BE HEADING BACK TO SURGERY VERY SOON FOR AN ADDITIONAL POSTERIOR CONSTRUCT TO REINFORCE THE ANTERIOR CAGES AT L4/5 AND L5/S1.AS IT WAS REPORTED THE PATIENT HAD A RETROPERITONEAL ABSCESS AT L4/5-L5/S1. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. A REVISION SURGERY WAS NECESSARY. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE XC 100032168; CC 400582800. ASSOCIATED MEDWATCH REPORTS: 400582801_2916714-2023-00022_MDR, 400582802_2916714-2023-00023_MDR.

    Patient problems: Peritonitis

    MAUDE report 16230491 ↗
  • InjuryJun 22, 2022Migration or Expulsion of Device; Migration; Migration or Expulsion of Device; Migration

    IT WAS REPORTED TO AESCULAP AAG THAT AN ACTIV L IMPLANT SYSTEM (PART # UNKNOWN) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE DEVICE MIGRATED ONE (1) MONTH POST SURGERY. REPORTEDLY, THE PATIENT HAD A MULTI-LEVEL LUMBAR ARTHROPLASTY FROM TWO (2) PRIOR SURGERIES. THE PATIENT HAD A COMPETITIVE DISC REPLACEMENT AT L5/S1 PERFORMED ON AN UNKNOWN DATE. DURING A SECONDARY SURGERY, PERFORMED IN (B)(6) 2022, THE ACTIV L DEVICES WERE IMPLANTED AT LEVELS L4/L5 AND AT L3/L4. THE TOP LEVEL, L3/L4, THE UPPERMOST OF THE THREE (3) IMPLANTS, MIGRATED ANTERIORLY, WHICH REQUIRED EXPLANTATION. THE PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2022. THE COMPLAINT DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. A REVISION SURGERY WAS NECESSARY. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION SURGERY. THE ACTIV L DEVICE WAS REMOVED WITHOUT INCIDENT DURING THE LATERAL REVISION PROCEDURE. THAT THERE WAS AN ISSUE WITH AE-QAS-SP42 - COLLECT.NO.QAS SPINE ANTERIOR STABILIS. THE ACTUAL DEVICE IS UNKNOWN. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4). ASSOCIATED MEDWATCH-REPORTS: 9610612-2022-00133 (400556960 - AE-QAS-

    Patient problems: Failure of Implant

    MAUDE report 14777144 ↗
  • InjuryJun 22, 2022Migration; Migration

    IT WAS REPORTED TO AESCULAP INC. THAT AN ACTIV L IMPLANT SYSTEM (PART # UNKNOWN) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE DEVICE MIGRATED ONE (1) MONTH POST SURGERY. REPORTEDLY, THE PATIENT HAD A MULTI-LEVEL LUMBAR ARTHROPLASTY FROM TWO (2) PRIOR SURGERIES. THE PATIENT HAD A COMPETITIVE DISC REPLACEMENT AT L5/S1 PERFORMED ON AN UNKNOWN DATE. DURING A SECONDARY SURGERY, PERFORMED IN (B)(6) 2022, THE ACTIV L DEVICES WERE IMPLANTED AT LEVELS L4/L5 AND AT L3/L4. THE TOP LEVEL, L3/L4, THE UPPERMOST OF THE THREE (3) IMPLANTS, MIGRATED ANTERIORLY, WHICH REQUIRED EXPLANTATION. THE PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2022. THE COMPLAINT DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. A REVISION SURGERY WAS NECESSARY. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION SURGERY. THE ACTIV L DEVICE WAS REMOVED WITHOUT INCIDENT DURING THE LATERAL REVISION PROCEDURE. THE ADVERSE EVENT IS FILED UNDER AIC REFERENCE 100030868. ASSOCIATED MEDWATCH-REPORTS: 2916714-2022-00065 (400556959 - AE-QAS-SP42) 2916714-2022-00066 (400556960 - AE-QAS-SP42) 2916714-2022-00067 (400556961 - AE-QAS-SP42)

    Patient problems: Failure of Implant

    MAUDE report 14777141 ↗

Other PROSTHESIS, INTERVERTEBRAL DISC devices

All PROSTHESIS, INTERVERTEBRAL DISC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page