Skip to content

Olympus WA50011ARecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus laparoscope accessory is used with video endoscopy systems. The 23 reports mostly describe physical damage such as dents, cracks and breaks and poor or dim images, often noticed during inspection or servicing, with no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
25
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 04042761061728), received between Mar 9, 2023 and Sep 22, 2026 — classified as 25 malfunctions.

The most frequently reported specific device problem is “Poor Quality Image” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Poor Quality Image8
  • Break4
  • Dent in Material4
  • Crack2
  • Detachment of Device or Device Component2
  • Leak/Splash2
  • Material Deformation2
  • Communication or Transmission Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions25

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026Break; Poor Quality Image; Dent in Material; Scratched Material

    The customer reported a black shadow in the upper right corner and side of the viewing area involving an Olympus Laparoscopes, Video. Additionally, during the initial evaluation of the device, fiber breakage, dents, and scratches on the distal frame, and dents and scratches on the outer tube were observed. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50857098 ↗
  • MalfunctionSep 14, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50763285 ↗
  • MalfunctionAug 4, 2026Break; Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50131974 ↗
  • MalfunctionJun 5, 2026Crack

    THE CUSTOMER REPORTED A CRACKED LIGHT SOURCE DURING INSPECTION FOR USE INVOLVING AN OLYMPUS VIDEO LAPAROSCOPE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25414171 ↗
  • MalfunctionApr 23, 2026Detachment of Device or Device Component; Material Split, Cut or Torn

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE LAPAROSCOPES, VIDEO PRESENTED MISSING ADAPTOR AND CUT IN THE LIGHT GUIDE CABLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24966411 ↗
  • MalfunctionFeb 25, 2026Material Deformation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE LAPAROSCOPES VIDEO SCOPES HAD DENTS ON THE DISTAL EDGE AND OUTER TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442913 ↗
  • MalfunctionFeb 11, 2026Poor Quality Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24315589 ↗
  • MalfunctionJan 2, 2026Crack; Material Deformation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO TELESCOPE EXHIBITED DENTS AND BENT ON OUTER TUBE, CRACK ON LEFT SIDE OF VIDEO CONNECTOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23949769 ↗
  • MalfunctionJul 16, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE EXPERIENCE DISTURBANCE OF VIDEO IMAGE. THIS WAS DISCOVERED DURING SERVICE / MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22524622 ↗
  • MalfunctionJun 25, 2025Optical Problem; Optical Problem

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A DIM LIGHT. THE ISSUE OCCURRED DURING SEDATION (DEVICE OUTSIDE PATIENT) OF AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22329471 ↗
  • MalfunctionJun 4, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD POOR IMAGE QUALITY. THE ISSUE WAS FOUND DURING SET UP / INSPECTION FOR USE (DURING SET UP IN ROOM / BEFORE PATIENT IN ROOM). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22147158 ↗
  • MalfunctionApr 29, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD DISTURBANCE OF VIDEO IMAGE. THE ISSUE WAS FOUND DURING MAINTENANCE OF DEVICE. THERE ARE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21925246 ↗
  • MalfunctionApr 28, 2025Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT DURING SETUP/INSPECTION FOR USE, THE RIGID VIDEO SCOPE DISPLAYED INCOMPATIBLE SCOPE ERROR E315. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21918198 ↗
  • MalfunctionMar 6, 2025Dent in Material; Dent in Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED DENTED RIGID TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21536088 ↗
  • MalfunctionFeb 7, 2025Dent in Material; Dent in Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO SCOPE HAD A DENT IN INSERTION TUBE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21334860 ↗
  • MalfunctionOct 28, 2024Dent in Material; Dent in Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT RIGID VIDEO SCOPE EXHIBITED MINOR DENTS ON THE DISTAL EDGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20544563 ↗
  • MalfunctionSep 4, 2024Electrical Shorting; Electrical Shorting

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A SHORT IN THE WIRE. THERE WERE NO REPORTS OF PATIENT HARM. NO HARM WAS REPORTED DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20141076 ↗
  • MalfunctionJul 26, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED, THAT THE SUBJECT DEVICE HAD AIR LEAKAGE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19839850 ↗
  • MalfunctionJul 25, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED DURING REPROCESSING, THE VIDEO SCOPE HAD A BUBBLE COMING OUT WHEN THEY WERE LOOKING AT IT, IT HAD A LEAK. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19828369 ↗
  • MalfunctionMay 22, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THAT DURING PREPARATION FOR A LAPAROSCOPY PROCEDURE, THE SUBJECT DEVICE IMAGE WAS IN AND OUT INTERMITTENTLY. THE PROCEDURE WAS COMPLETED UTILIZING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19371768 ↗
  • MalfunctionMay 10, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED, THAT THE IMAGE IS DARK ON THE RIGID VIDEO SCOPE. THE ISSUE WAS FOUND, DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19287858 ↗
  • MalfunctionApr 18, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THE RIGID VIDEO SCOPE DISPLAYED A BLURRY IMAGE. THERE WAS NO PATIENT INVOLVEMENT. EVENT DESCRIPTION-SCOPE IS BLURRY CUTOVER Y/N-Y WHEN DURING PROCEDURE DID EVENT OCCUR?-UNKNOWN ^09.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19133527 ↗
  • MalfunctionAug 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED TO OLYMPUS, THE OPERATION BUTTON DETACHED FROM THE VIDEO TELESCOPE. THE ISSUE WAS FOUND DURING REPROCESSING. THE INTENDED PROCEDURE WAS AN UNKNOWN THERAPEUTIC PROCEDURE. THE PROCEDURE WAS COMPLETED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17645749 ↗
  • MalfunctionJun 23, 2023Break; Break

    THE CUSTOMER RETURNED THE VIDEO TELESCOPE "ENDOEYE", TO OLYMPUS REPAIR CENTER FOR EVALUATION OF A REPORTED DAMAGED CABLE. DURING THE EVALUATION, DAMAGED COVER GLASS, AND A BROKEN LIGHT GUIDE BUNDLE WERE DETECTED. THIS REPORT IS BEING SUBMITTED TO CAPTURE THE DAMAGED COVER GLASS, AND A BROKEN LIGHT GUIDE BUNDLE DEFECTS FOUND DURING THE REPAIR EVALUATION. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17193889 ↗
  • MalfunctionMar 9, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, THE AUTOCLAVABLE VIDEO TELESCOPE HAD A BROKEN LIGHT. NO PATIENT WAS INVOLVED IN THIS EVENT AND THE PLANNED PROCEDURE WAS THERAPEUTIC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16518749 ↗

Other Endoscope, Accessories, Narrow Band Spectrum devices

All Endoscope, Accessories, Narrow Band Spectrum devices →