Skip to content

Infusomat® 480254Recalls & reported problems

B. Braun Medical Inc. · (15 drops/mL) Universal spike, pinch clamp, pressure limited check valve, four CARESITE® injection sites, 92 in., 61 in., 41 in.,6 in. above distal end, free flow protection clamp, backcheck valve, roller clamp, slide clamp, SPIN-LOCK® connector. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04046955324798), received between Apr 6, 2022 and May 23, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Misassembled During Manufacturing /Shipping” (2 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Oct 29, 2025Open, Classified

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

    Z-0640-2026 · Root cause: Process control · FDA record ↗

  2. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2394-2024 · Root cause: Process control · FDA record ↗

  3. Nov 1, 2023Open, Classified

    In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.

    Z-0355-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Device Misassembled During Manufacturing /Shipping2
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 23, 2023Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: LUER END/SPIN LOCK CRACKED, LEADING TO DISLODGEMENT AND CHEMO SPILL NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16982552 ↗
  • MalfunctionNov 4, 2022Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: THE NORMAL IV TUBING WAS MADE OR PUT TOGETHER BACKWARDS. THE WHOTE CASSETTE FILTER AND GREEN CLAMP THAT GO INTO THE PUMP WERE INVERTED. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15733370 ↗
  • MalfunctionApr 6, 2022Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    AS REPORTED BY THE USER FACILITY: IT WAS REPORTED THAT THE GREEN ANTI-FREE FLOW PROTECTION CLAMP WAS INCORRECTLY PLACED ON THE SET. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14030653 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page