Infusomat recallZ-0640-2026 · B Braun Medical Inc · initiated Oct 29, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from B Braun Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA root cause category: Process control
What the company is doing
B Braun notified consignees on about 10/29/2025 via letter. Consignees were instructed to follow "Interim Measures for Users" as outlined in the letter for secondary infusions via gravity sets and if using secondary infusions with infusion pumps. Consignees were also instructed to inform all users and post the applicable attachment(s) in care areas where secondary infusions may be administered and where IV administration sets and pump administration sets may be stored until further notice. B. Braun will provide additional notification when users may resume their normal secondary administration process without the need for the modified workflow. Consider priming all IV administration sets and pump administration sets with normal saline or other standard solutions prior to priming with expensive medications to avoid loss of expensive medications or occupational exposure to hazardous medications in the event the set is unable to be primed. If difficulty priming is encountered, obtain a replacement device and file a customer complaint with B. Braun. Additionally, consignees were requested to ensure all pertinent employees and customers are informed, post the notification where affected products are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form."
Recall details
- Recall number
- Z-0640-2026
- Status
- Open, Classified
- Initiated by the company
- Oct 29, 2025
- Posted by the FDA
- Nov 26, 2025
- Recalling firm
- B Braun Medical Inc
- Quantity
- 49,330,370
- Product code
- FPA
- FDA submissions
- K142036
- Distribution
- Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Product description
Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. UNIV.15DROP PUMP SET W/3 SAFELINE LL; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 362032. 2. UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; Catalog Number: 362033. 3. 60DROP METRISET PUMP SET, 3 SAFELINE; Priming volume: 19mL, length: 120in, SafeLine; Catalog Number: 362034. 4. Y TYPE BLOOD PUMP SET, SAFELINE INJ; Priming volume: 39mL, length: 120in, SafeLine; Catalog Number: 362043. 5. UNIV. 15 DROP PUMP SET W/ 0.2 FILTER; 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; Catalog Number: 362050. 6. SPACE PUMP SET 15 DROP W/2 SAFEDAY; SPACE PUMP SET 15 DROP W/2 SAFEDAY; Catalog Number: 362420. 7. UNIV. 15 DROP PUMP SET W/ 3 ULTRASITE LL; 15 drops/mL, priming volume: 21mL, length: 120in, ULTRASITE; Catalog Number: 362431. 8. SPACE PUMP IV SET, 120 IN.; 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; Catalog Number: 362432. 9. 60 DROP METRISET PUMP SET,3 SFLINE ASV; Priming volume: 21mL, length: 121in, SafeLine; Catalog Number: 363010. 10. Y-TYPE BLOOD PUMP SET, SAFELINE, ASV; Priming volume: 40mL, length: 123in, SafeLine; Catalog Number: 363019. 11. UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 363030. 12. UNIV. 15 DROP PUMP SET, 3 ULTRASITE, ASV; 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; Catalog Number: 363230. 13. Y-TYPE BLOOD PUMP SET, CARESITE, ASV; Priming volume: 41mL, length: 120in, CARESITE; Catalog Number: 363419. 14. SPACE PUMP IV SET, 2 CARESITES, ASV; 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; Catalog Number: 363420. 15. UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 363421. 16. 60DROP METRISET PUMP, 3 CARESITES, ASV; Priming volume: 21mL, length: 130in, CARESITE; Catalog Number: 363422. 17. UNIV. 60 DROP PUMP SET, 2 CARESITES, ASV; 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; Catalog Number: 363423. 18. UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; Catalog Number: 363424. 19. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; Catalog Number: 363430. 20. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; Catalog Number: 363433. 21. SPACE PUMP IV SET 15D, 110 IN.; "SPACE PUMP IV SET 15D, 110 IN. CLOSE"; Catalog Number: 480254. 22. SPACE PUMP IV SET, 127 IN.; SPACE PUMP IV SET, 127 IN.; Catalog Number: 480255. 23. SPACE PUMP SET 15D, CARESITE LAD 122 IN. ; 15 drops/mL, priming volume: 23mL, length: 122in, CARESITE; Catalog Number: 480256. 24. 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; Catalog Number: 480258. 25. SPACE PUMP IV SET 15D, 90 IN.; SPACE PUMP IV SET 15D, 90 IN.; Catalog Number: 480263. 26. UNIV. 15 DROP PUMP SET W 3 CARESITE L.L.; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 490100. 27. 60 DROP METRISET PUMP SET, 3 CARESITES; Priming volume: 19mL, length: 120in, CARESITE; Catalog Number: 490101.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
1. Catalog Number: 362032; Primary UDI-DI: 04046964182556; Unit of Dose UDI-DI: 04046964182549. Shelf Life: Earliest Exp of Comp or 60mths. 2. Catalog Number: 362033; Primary UDI-DI: 04046964837821; Unit of Dose UDI-DI: 04046964837814. Shelf Life: Earliest Exp of Comp or 36mths. 3. Catalog Number: 362034; Primary UDI-DI: 04046964182570; Unit of Dose UDI-DI: 04046964182563. Shelf Life: Earliest Exp of Comp or 48mths. 4. Catalog Number: 362043; Primary UDI-DI: 04046964182631; Unit of Dose UDI-DI: 04046964182624. Shelf Life: Earliest Exp of Comp or 60mths. 5. Catalog Number: 362050; Primary UDI-DI: 04046964182655; Unit of Dose UDI-DI: 04046964182648. Shelf Life: Earliest Exp of Comp or 60mths. 6. Catalog Number: 362420; Primary UDI-DI: 04046955592715; Unit of Dose UDI-DI: 04046955592708. Shelf Life: 36mths from Start Date. 7. Catalog Number: 362431; Primary UDI-DI: 04046964294013; Unit of Dose UDI-DI: 04046964294006. Shelf Life: Earliest Exp of Comp or 60mths. 8. Catalog Number: 362432; Primary UDI-DI: 04046964182679; Unit of Dose UDI-DI: 04046964182662. Shelf Life: Earliest Exp of Comp or 60mths. 9. Catalog Number: 363010; Primary UDI-DI: 04046964294174; Unit of Dose UDI-DI: 04046964294167. Shelf Life: Earliest Exp of Comp or 48mths. 10. Catalog Number: 363019; Primary UDI-DI: 04046964294198; Unit of Dose UDI-DI: 04046964294181. Shelf Life: Earliest Exp of Comp or 60mths. 11. Catalog Number: 363030; Primary UDI-DI: 04046964294259; Unit of Dose UDI-DI: 04046964294242. Shelf Life: Earliest Exp of Comp or 60mths. 12. Catalog Number: 363230; Primary UDI-DI: 04046964294495; Unit of Dose UDI-DI: 04046964294488. Shelf Life: Earliest Exp of Comp or 60mths. 13. Catalog Number: 363419; Primary UDI-DI: 04046964294570; Unit of Dose UDI-DI: 04046964294563. Shelf Life: Earliest Exp of Comp or 36mths. 14. Catalog Number: 363420; Primary UDI-DI: 04046964294594; Unit of Dose UDI-DI: 04046964294587. Shelf Life: Earliest Exp of Comp or 36mths. 15. Catalog Number: 363421; Primary UDI-DI: 04046964294617; Unit of Dose UDI-DI: 04046964294600. Shelf Life: Earliest Exp of Comp or 36mths. 16. Catalog Number: 363422; Primary UDI-DI: 04046964294631; Unit of Dose UDI-DI: 04046964294624. Shelf Life: Earliest Exp of Comp or 36mths. 17. Catalog Number: 363423; Primary UDI-DI: 04046964294655; Unit of Dose UDI-DI: 04046964294648. Shelf Life: Earliest Exp of Comp or 36mths. 18. Catalog Number: 363424; Primary UDI-DI: 04046964837906; Unit of Dose UDI-DI: 04046964837890. Shelf Life: Earliest Exp of Comp or 36mths. 19. Catalog Number: 363430; Primary UDI-DI: 04046964294679; Unit of Dose UDI-DI: 04046964294662. Shelf Life: 36mths from Start Date. 20. Catalog Number: 363433; Primary UDI-DI: 04046964837920; Unit of Dose UDI-DI: 04046964837913. Shelf Life: Earliest Exp of Comp or 36mths. 21. Catalog Number: 480254; Primary UDI-DI: 04046955324798; Unit of Dose UDI-DI: 04046955343904. Shelf Life: Earliest Exp of Comp or 36mths. 22. Catalog Number: 480255; Primary UDI-DI: 04046955324804; Unit of Dose UDI-DI: 04046955585403. Shelf Life: Earliest Exp of Comp or 36mths. 23. Catalog Number: 480256; Primary UDI-DI: 04046955324811; Unit of Dose UDI-DI: 04046955585861. Shelf Life: Earliest Exp of Comp or 36mths. 24. Catalog Number: 480258; Primary UDI-DI: 04046955324835; Unit of Dose UDI-DI: 04046955585885. Shelf Life: Earliest Exp of Comp or 36mths. 25. Catalog Number: 480263; Primary UDI-DI: 04046955693931; Unit of Dose UDI-DI: 04046955693924. Shelf Life: Earliest Exp of Comp or 36mths. 26. Catalog Number: 490100; Primary UDI-DI: 04046964186080; Unit of Dose UDI-DI: 04046964186073. Shelf Life: 36mths from Start Date. 27. Catalog Number: 490101; Primary UDI-DI: 04046964186103; Unit of Dose UDI-DI: 04046964186097. Shelf Life: Earliest Exp of Comp or 36mths. 28. Catalog Number: 490102; Primary UDI-DI: 04046964186127; Unit of Dose UDI-DI: 04046964186110. Shelf Life: Earliest Exp of Comp or 36mths. 29. Catalog Number: 490103; Primary UDI-DI: 04046964302107; Unit of Dose UDI-DI: 04046964302091. Shelf Life: Earliest Exp of Comp or 36mths. 30. Catalog Number: 490104; Primary UDI-DI: 04046964186141; Unit of Dose UDI-DI: 04046964186134. Shelf Life: Earliest Exp of Comp or 36mths. 31. Catalog Number: 490105; Primary UDI-DI: 04046964186165; Unit of Dose UDI-DI: 04046964186158. Shelf Life: Earliest Exp of Comp or 36mths. 32. Catalog Number: 490655; Primary UDI-DI: 04046955836383; Unit of Dose UDI-DI: 04046955836376. Shelf Life: Earliest Exp of Comp or 36mths.
Devices named in this recall
32 devices in the FDA device registry whose identifier appears in the recall's code information.
- Infusomat® 490100B. Braun Medical Inc. · Set, administration, intravascular3 recalls2 deaths408 reports
- Infusomat® 490102B. Braun Medical Inc. · Set, administration, intravascular3 recalls3 deaths349 reports
- Infusomat® 363430B. Braun Medical Inc. · Set, administration, intravascular3 recalls245 reports
- Infusomat® 490105B. Braun Medical Inc. · Set, administration, intravascular2 recalls174 reports
- Infusomat® 490103B. Braun Medical Inc. · Set, administration, intravascular3 recalls108 reports
- Infusomat® 363421B. Braun Medical Inc. · Set, administration, intravascular3 recalls86 reports
- Infusomat® 363419B. Braun Medical Inc. · Set, administration, intravascular2 recalls109 reports
- Infusomat® 362420B. Braun Medical Inc. · Set, administration, intravascular3 recalls29 reports
- Infusomat® 363420B. Braun Medical Inc. · Set, administration, intravascular3 recalls29 reports
- Infusomat® 490101B. Braun Medical Inc. · Set, administration, intravascular3 recalls13 reports
- Infusomat® 363230B. Braun Medical Inc. · Set, administration, intravascular3 recalls7 reports
- Infusomat® 363423B. Braun Medical Inc. · Set, administration, intravascular3 recalls5 reports
- Infusomat® 363424B. Braun Medical Inc. · Set, administration, intravascular3 recalls4 reports
- Infusomat® 480254B. Braun Medical Inc. · Set, administration, intravascular3 recalls3 reports
- Infusomat® 490104B. Braun Medical Inc. · Set, administration, intravascular3 recalls3 reports
- Infusomat® 362033B. Braun Medical Inc. · Set, administration, intravascular3 recalls3 reports
- Infusomat® 362050B. Braun Medical Inc. · Set, administration, intravascular3 recalls1 report
- Infusomat® 363010B. Braun Medical Inc. · Set, administration, intravascular3 recalls1 report
- Infusomat® 480255B. Braun Medical Inc. · Set, administration, intravascular3 recalls0 reports
- Infusomat® 480263B. Braun Medical Inc. · Set, administration, intravascular3 recalls0 reports
- Infusomat® 362032B. Braun Medical Inc. · Set, administration, intravascular3 recalls0 reports
- Infusomat® 362034B. Braun Medical Inc. · Set, administration, intravascular3 recalls0 reports
- Infusomat® 362432B. Braun Medical Inc. · Set, administration, intravascular3 recalls0 reports
- Infusomat® 363030B. Braun Medical Inc. · Set, administration, intravascular3 recalls0 reports
- Infusomat® 363422B. Braun Medical Inc. · Set, administration, intravascular3 recalls0 reports
- Infusomat® 363433B. Braun Medical Inc. · Set, administration, intravascular3 recalls0 reports
- Infusomat® 480258B. Braun Medical Inc. · Set, administration, intravascular2 recalls6 reports
- Infusomat® 363019B. Braun Medical Inc. · Set, administration, intravascular2 recalls2 reports
- Infusomat® 362431B. Braun Medical Inc. · Set, administration, intravascular2 recalls1 report
- Infusomat® 362043B. Braun Medical Inc. · Set, administration, intravascular2 recalls0 reports
- Infusomat® 490655B. Braun Medical Inc. · Set, administration, intravascular1 recall1 report
- Infusomat® 480256B. Braun Medical Inc. · Set, administration, intravascular1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97816).
- Z-0603-2026 · Open, ClassifiedADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled…
- Z-0604-2026 · Open, ClassifiedADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled…
- Z-0605-2026 · Open, ClassifiedADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled…
- Z-0606-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0607-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0608-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0609-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0610-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0611-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0612-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0613-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0614-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0615-2026 · Open, ClassifiedB.BRAUN utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled…
- Z-0616-2026 · Open, ClassifiedBlood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Bas…
- Z-0617-2026 · Open, ClassifiedBlood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Bas…
- Z-0618-2026 · Open, ClassifiedBlood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Bas…
- Z-0619-2026 · Open, ClassifiedBlood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Bas…
- Z-0620-2026 · Open, ClassifiedBlood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Bas…
- Z-0621-2026 · Open, ClassifiedBurette Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Lab…
- Z-0622-2026 · Open, ClassifiedBurette Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Lab…
- Z-0623-2026 · Open, ClassifiedCARESAFE utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labele…
- Z-0624-2026 · Open, ClassifiedCaresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labele…
- Z-0625-2026 · Open, ClassifiedCaresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labele…
- Z-0626-2026 · Open, ClassifiedCaresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labele…
- Z-0627-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
- Z-0628-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
- Z-0629-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
- Z-0630-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
- Z-0631-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
- Z-0632-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page