Skip to content

Infusomat® 480258Recalls & reported problems

B. Braun Medical Inc. · 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04046955324835), received between Jan 5, 2022 and Nov 13, 2022 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Device Misassembled During Manufacturing /Shipping” (3 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Oct 29, 2025Open, Classified

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

    Z-0640-2026 · Root cause: Process control · FDA record ↗

  2. Nov 1, 2023Open, Classified

    In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.

    Z-0355-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Device Misassembled During Manufacturing /Shipping3
  • Leak/Splash3

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 13, 2022Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: GREEN CLAMP OF PUMP SEGMENT IS ON THE WRONG SIDE OF THE PUMP SEGMENT. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15786260 ↗
  • MalfunctionNov 13, 2022Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: GREEN CLAMP OF PUMP SEGMENT IS ON THE WRONG SIDE OF THE PUMP SEGMENT. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15786259 ↗
  • MalfunctionFeb 23, 2022Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: GREEN SLIDE CLAMP ON WRONG SIDE OF PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION: WHILE PRIMING, THE NURSE NOTICED THAT THE GREEN CLAMP WAS ON THE WRONG SIDE OF THE PUMP SEGMENT. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13583869 ↗
  • MalfunctionJan 5, 2022Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: EVENT 3: DETAILED INQUIRY DESCRIPTION THE DISTAL FILTERED EXTENSION SET HAS BECOME DISCONNECTED FROM THE PRIMARY PUMP TUBING DURING INFUSION RESULTING IN FLUID SPILLING ON THE PATIENT, FLOOR, NURSES AND IN SEVERAL INSTANCES BLOOD BACK FLOWED FROM PORT RESULTING IN ENVIRONMENTAL EXPOSURE. THE NURSES MENTIONED TIGHTENING THE CONNECTION PRIOR TO PRIMING. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13172997 ↗
  • MalfunctionJan 5, 2022Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: EVENT 1: DETAILED INQUIRY DESCRIPTION: THE DISTAL FILTERED EXTENSION SET HAS BECOME DISCONNECTED FROM THE PRIMARY PUMP TUBING DURING INFUSION RESULTING IN FLUID SPILLING ON THE PATIENT, FLOOR, NURSES AND IN SEVERAL INSTANCES BLOOD BACK FLOWED FROM PORT RESULTING IN ENVIRONMENTAL EXPOSURE. THE NURSES MENTIONED TIGHTENING THE CONNECTION PRIOR TO PRIMING. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13172995 ↗
  • MalfunctionJan 5, 2022Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: EVENT 2: DETAILED INQUIRY DESCRIPTION: THE DISTAL FILTERED EXTENSION SET HAS BECOME DISCONNECTED FROM THE PRIMARY PUMP TUBING DURING INFUSION RESULTING IN FLUID SPILLING ON THE PATIENT, FLOOR, NURSES AND IN SEVERAL INSTANCES BLOOD BACK FLOWED FROM PORT RESULTING IN ENVIRONMENTAL EXPOSURE. THE NURSES MENTIONED TIGHTENING THE CONNECTION PRIOR TO PRIMING. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13172993 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →