AESCULAP PL579TRecalls & reported problems
Aesculap AG · CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR. · Class 2 · Implant
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04046963398835), received between Mar 26, 2021 and Dec 27, 2023 — classified as 5 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Unintended Ejection” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Ejection2
- Break1
- Detachment of Device or Device Component1
- Ejection Problem1
- Failure to Eject1
- Material Fragmentation1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information3
- Liver Laceration(s)1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryDec 27, 2023Unintended Ejection; Unintended Ejection
IT WAS REPORTED TO AESCULAP INC. THAT A CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR. (PART# PL579T) WAS USED DURING A BARIATRIC GASTRIC BYPASS ON DECEMBER 06 2023. ACCORDING TO THE COMPLAINANT THE CLIP APPLIER SUCCESSFULLY FIRED AND CLIPPED TWICE. FURTHER USE OF THE CLIP APPLIER CAUSED THE CLIPS TO BACK UP IN CARTRIDGE. REPORTEDLY THE STAFF TRIED AGAIN TO ENGAGE THE CLIP APPLIER AND 3 TITANIUM CLIPS DROPPED INTO THE PATIENT. THE PIECES WERE RETRIEVED DURING SURGERY AND REMOVED FROM INSIDE OF PATIENT. A DELAY OF TWENTY MINUTES WAS REPORTED AS A RESULT OF THIS INCIDENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE ADVERSE EVENT IS FILED UNDER AIC REFERENCE (B)(4). ASSOCIATED MEDWATCH REPORTS: (B)(4) (2916714-2023-00125) PL520R, (B)(4) (2916714-2023-00127) PL536R.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18405707 ↗InjuryFeb 17, 2022Unintended Ejection; Unintended Ejection
IT WAS REPORTED TO AESCULAP AG THAT A CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR (PART # PL579T) WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE DEVICE IS USED IN LAPAROSCOPIC CHOLECYST RESECTION PROCEDURES.THE PROBLEM ENCOUNTERED WAS THE DISLOCATION OF THE CLIP BLISTER (PL579T) FROM THE INSTRUMENT, WHEN USING THE DEVICE. WHEN APPLYING THE SECOND / THIRD CLIP, THE HANDLING MECHANISM JAMS AND THIS BLOCKAGE CAUSES THE BLISTER TO DISLODGE AND FALL IN THE PATIENT'S ABDOMEN. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. A SURGICAL DELAY OF FIFTEEN (15) MINUTES WAS REPORTED. AN ADDITIONAL MEDICAL INTERVENTION WAS NECESSARY. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4). INVOLVED COMPONENTS: PL606R - MULTIF.CLIP APPL.TI-P 10/370MM F/ML-CLIP - LOT UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13550337 ↗MalfunctionFeb 3, 2022Failure to Eject; Failure to Eject
IT WAS REPORTED TO AESCULAP AG THAT A CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR. (PART # PL579T) WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE CLIPS PUSHED AMONG THEMSELVES DURING SURGERY. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4). INVOLVED COMPONENTS: PL536R - SHAFT COMPL.D:10MM L:370MM - LOT: 62463532; PL520R - CHALLENGER TI-P HANDLE - LOT: UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13444660 ↗MalfunctionNov 7, 2021Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH PL579T - CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR.. ACCORDING TO THE COMPLAINT DESCRIPTION, THE CARTRIDGE TIP BROKE DURING SURGERY. THE PATIENT HARM IS UNKNOWN. ADDITIONAL INFORMATION WAS NOT PROVIDED NOR AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12762902 ↗InjuryOct 29, 2021Ejection Problem; Ejection Problem
IT WAS REPORTED TO AESCULAP AG THAT A CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR (PART # PL579T) WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE CLIP WAS DAMAGED DURING SURGERY. REPORTEDLY, THE CLIP RAMP OPENED DURING THE SECOND APPLICATION AND BECAME DETACHED FROM THE REUSABLE GRIPPER. RISK OF BLEEDING FROM VESICULAR INJURY IN THE HEPATHIC PEDICLE. THE USER EXPERIENCED DIFFICULTY STORING THE CLIP RAMP AND PULLING IT OUT THROUGH THE LAPAROSCOPY TROCARS. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. PATIENT HARM IS UNKNOWN. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER (B)(4) REFERENCE (B)(4).
Patient problems: Insufficient Information; Liver Laceration(s)
MAUDE report 12719874 ↗MalfunctionJul 13, 2021Material Rupture; Material Rupture
IT WAS REPORTED TO AESCULAP AG THAT A CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR (PART # PL579T) WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE CLIP PIN BROKE AND DETACHED INTO THE PATIENT. THE PART WAS NOT ABLE TO BE RECOVERED. THE FRAGMENT REMAINED IN SITU. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED AS OF THIS REPORT. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: Insufficient Information
MAUDE report 12159154 ↗InjuryMay 18, 2021Break; Break
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH PL579T-CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR. ACCORDING TO THE COMPLAINT DESCRIPTION, THE TIP OF THE MAGAZINE BREAKS OFF INTRAOPERATIVELY IN SITU. COULD BE RECOVERED. AN ADDITIONAL MEDICAL INTERVENTION WAS NECESSARY. ADDITIONAL INFORMATION WAS NOT PROVIDED NOR AVAILABLE / WAS NOT AVAILABLE. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: Insufficient Information
MAUDE report 11841563 ↗InjuryMar 26, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH PL579T - CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR.. ACCORDING TO THE COMPLAINT DESCRIPTION, THE CLIP DARTRIDGE FELL INTO PATIENTS´S BODY. WHEN THE MAIN BODY SHAFT WAS PULLED OUT FROM THE TROCAR, THE CARTRIDGE FELL INTO THE PATIENT'S BODY. AN ADDITIONAL MEDICAL INTERVENTION WAS NECESSARY. ADDITIONAL INFORMATION WAS NOT PROVIDED. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11574636 ↗
Other CLIP, IMPLANTABLE devices
- LIGAMAX EL5MLEthicon Endo-Surgery, LLC · CLIP, IMPLANTABLE3 deaths3,808 reports
- Endo Clip III 176630Covidien LP · CLIP, IMPLANTABLE1 death1,959 reports
- Epix® Universal Clip Applier CA500Applied Medical Resources Corporation · Clip, Implantable1 recall1 death1,183 reports
- Endo Clip II 176657Covidien LP · CLIP, IMPLANTABLE1,183 reports
- LIGACLIP ER320Ethicon Endo-Surgery, LLC · CLIP, IMPLANTABLE1,115 reports
- Direct Drive® Clip Applier CA090Applied Medical Resources Corporation · Clip, Implantable1 death806 reports
- LAPRA-TY XC200Ethicon Inc. · CLIP, IMPLANTABLE791 reports
- Endo Clip 176625Covidien LP · CLIP, IMPLANTABLE1 death768 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.