Skip to content

Perfusor® 8713080URecalls & reported problems

B. Braun Medical Inc. · Perfusor® PCA Syringe Pump · Class 2

All models of Perfusor (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 04046963915308), received between Mar 18, 2021 and Jun 23, 2026 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Excess Flow or Over-Infusion” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 14, 2024Class IIOpen, Classified

    Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

    Z-2435-2024 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Excess Flow or Over-Infusion5
  • Insufficient Flow or Under Infusion5
  • Infusion or Flow Problem2
  • Data Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Irritability1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2026Excess Flow or Over-Infusion

    AS REPORTED BY THE USER FACILITY: FENTANYL CONTINUOUS INFUSION SET UP AND CHECKED. PER BEDSIDE NURSE SHE HAD TO PAUSE THE INFUSION PUMP TO CHECK PRESSURES-PUMP PLACED IN STANDBY. NOTHING WAS MODIFIED UPON RESTARTING AND WITH 20 MINUTES THE SYRINGE HAD INFUSED. THE ORIGINAL INFUSION WAS AT .05CC AND WAS LATER FOUND INFUSING AT 25ML/HR.. IT WAS ORIGINALLY STARTED AS A CONTINUOUS SEDATION. GTB AND PLASTICS NOTIFIED. TREATMENT ORDERS GIVEN. PCA STOPPED. PUMP AND TUBING REMOVED.. PUMP TURNED IN TO MAINTENANCE. AFTER SPEAKING TO GTB THIS SIMILAR ISSUE WAS NOTED WIN @D WITH A PCA SYRINGE BEING INFUSED AS A BOLUS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25596938 ↗
  • MalfunctionMay 4, 2026Excess Flow or Over-Infusion

    AS REPORTED BY THE USER FACILITY: POTENTIAL ISSUE WITH THE IV INFUSION PUMP/PROGRAMMING LEADING TO LIKELY MEDICATION ERROR RECOGNIZED AS FENTANYL SYRINGE SCANNED AND STARTED AT 1854, IMMEDIATELY PRIOR TO CHANGE OF SHIFT ALARMED WITH VTBI WARNING AND APPEARS EMPTY AT APPROX. 1945. GTB APP, CHARGE NURSE, AND PHARMACY NOTIFIED WITHOUT FURTHER INTERVENTIONS NEEDED AS PATIENT IS ALREADY INTUBATED AND MECHANICALLY VENTILATED. NO SIGNS OF COMPLICATIONS NOTED ASIDE FROM DECREASED RAAS. PROPOFOL DOSE CUT SIGNIFICANTLY TO ASSESS NEURO STATUS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25072227 ↗
  • MalfunctionJan 15, 2026Insufficient Flow or Under Infusion

    AS REPORTED BY THE USER FACILITY: PATIENT (CMO) GOT A NEW SYRINGE (15ML) FENTANYL INITIATED AT SHIFT CHANGE (1914). MED CONCENTRATION WAS 50MCG/1ML. RATE 75MCG PER HOUR. SYRINGE SHOULD HAVE BEEN EMPTY AT 0514 (10 HOURS). WHEN RN WENT IN AT 5AM TO CHANGE SYRINGE, THE MACHINE SAID VTBI WAS 0.76, WHICH WOULD MATCH UP WITH RATE/TIMING. HOWEVER, WHEN SYRINGE WAS CHANGED, THERE WAS STILL >5ML IN SYRINGE TO WASTE. WHEN SYRINGE WAS CHANGED AT SHIFT CHANGE, THERE WAS NO ISSUES.

    Patient problems: Irritability

    MAUDE report 24084147 ↗
  • MalfunctionJun 13, 2025Infusion or Flow Problem; Infusion or Flow Problem

    AS REPORTED BY THE USER FACILITY: WHEN PUMP DISCONTINUED THERE WAS 4 ML/4 MG MORPHINE LEFT IN SYRINGE AND PUMP SHOWED 22.8 ML/MG.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22206093 ↗
  • MalfunctionMay 16, 2025Excess Flow or Over-Infusion; Data Problem; Excess Flow or Over-Infusion; Data Problem

    AS REPORTED BY THE USER FACILITY: PATIENT INTUBATED AND SEDATED WITH PROPOFOL AND 50MCG/HR OF FENTANYL. WHEN FENTANYL SYRINGE WAS DUE TO BE CHANGED THE PUMP SWITCHED TO A DIFFERENT MODE OF THE THERAPY AND INSTEAD WAS PROGRAMMED TO RUN AT 50 ML/HR. THE ENTIRETY OF THE 15 ML FENTANYL SYRINGE WAS INFUSED INTO THE PT. AT 50 ML/HR INSTEAD OF 50 MCG/HR. PUMP WAS REMOVED FROM ROOM AND TAGGED FOR SERVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22039920 ↗
  • MalfunctionJun 12, 2023Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    AS REPORTED BY THE USER FACILITY: RN IN PACU SET PCA PUMP TO INFUSE 50ML PER MD'S ORDER. PATIENT THEN MOVED TO A NURSING FLOOR. WHEN RN WAS GIVING REPORT, SHE NOTICED THE 50MLS FIRST PROGRAMMED WAS NOW SET TO 0. PATIENT WAS STILL ABLE TO GET INTERMITTENT INFUSIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17106197 ↗
  • MalfunctionMay 8, 2023Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    AS REPORTED BY THE USER FACILITY: THE NURSE CLAIMED THE PCA PUMP INACCURACIES. SAID THAT PUMP ALERTED VIAL EMPTY BUT THERE WAS STILL 9ML'S LEFT IN SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16885603 ↗
  • MalfunctionMay 8, 2023Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    AS REPORTED BY THE USER FACILITY: THE NURSE CLAIMED THE PCA PUMP INNACURACIES. SAID THAT PUMP ALARTED VIAL EMPTY BUT THERE WAS STILL 9ML'S LEFT IN SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16885581 ↗
  • MalfunctionMay 8, 2023Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    AS REPORTED BY THE USER FACILITY: THE NURSE CLAIMED THE PCA PUMP INACCURACIES. SAID THAT PUMP ALERTED VIAL EMPTY BUT THERE WAS STILL 9ML'S LEFT IN SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16885144 ↗
  • MalfunctionMay 19, 2021Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    AS REPORTED BY USER FACILITY: WHEN THE PATIENT WAS DISCHARGED, THE REMAINING VOLUME IN THE SYRINGE WAS 19ML, BUT THE PUMP INDICATED THAT THERE SHOULD BE 29.98ML REMAINING TO BE INFUSED. THE INFUSION WAS STARTED WITH A 50ML BD SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11847558 ↗
  • MalfunctionMay 19, 2021Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    AS REPORTED BY USER FACILITY: THE INFUSION WAS STARTED WITH A 50ML BD SYRINGE, INFUSING MORPHINE. THE PUMP ALARMED THAT THE SYRINGE WAS EMPTY, BUT IT WAS READING THAT 34.64MLS WERE DELIVERED, WITH 15.36MLS REMAINING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11847556 ↗
  • MalfunctionMar 18, 2021Infusion or Flow Problem; Infusion or Flow Problem

    AS REPORTED BY USER FACILITY: A PATIENT'S MEDICATION (MORPHINE) DOSING ON PCA PUMP SHOULD HAVE BEEN STEADY BUT NURSE NOTICED IT WAS CHANGED WITHOUT AUTHORIZATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11515414 ↗

FDA clearance and classification

The FDA regulates product code FRN (pump, infusion) under 21 CFR 880.5725 as a Class II device. 21 CFR 880.5725 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the Perfusor® 8713080U

Has the Perfusor® 8713080U been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2435-2024 (Class II), began on Jun 14, 2024: Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Perfusor® 8713080U?

The FDA's MAUDE database lists 12 adverse event reports for this device, received between Mar 18, 2021 and Jun 23, 2026. In those reports, the most common specific device problems are “Excess Flow or Over-Infusion” (5), “Insufficient Flow or Under Infusion” (5) and “Infusion or Flow Problem” (2); the most common patient problem is “Irritability” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Perfusor® 8713080U?

None of the reports are classified as deaths, 0 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Perfusor® 8713080U MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Perfusor® 8713080U, and is it still sold?

The FDA's device database lists B. Braun Medical Inc. as the company responsible for it (brand Perfusor®, model 8713080U). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Pump, infusion devices

All Pump, infusion devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page