Skip to content

Infusomat® 490101Recalls & reported problems

B. Braun Medical Inc. · Priming volume: 21mL, length: 130in, CARESITE® · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
13
Deaths reported
0
Injuries reported
1
Recalls naming it
3

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 04046964186103), received between Nov 16, 2018 and Apr 29, 2026 — classified as 1 injury, 12 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (7 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 29, 2025Open, Classified

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

    Z-0640-2026 · Root cause: Process control · FDA record ↗

  2. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2398-2024 · Root cause: Process control · FDA record ↗

  3. Nov 1, 2023Open, Classified

    In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.

    Z-0355-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Contamination /Decontamination Problem7
  • Air/Gas in Device6

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • No Consequences Or Impact To Patient2
  • Unspecified Infection1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: A PIECE OF PLASTIC WAS OBSERVED IN THE LINE. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25031651 ↗
  • MalfunctionMay 20, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY USER FACILITY: BRIEF INQUIRY DESCRIPTION AIR IN LINE FROM Y-CONNECTOR AT PROXIMAL PORT TO PATIENT DETAILED INQUIRY DESCRIPTION AIR IN LINE FROM Y-CONNECTOR AT PROXIMAL PORT TO PATIENT DESCRIPTION OF THE PATIENT EVENT PUMP INFUSING, NURSE NOTICED AIR IN LINE AND BUBBLING AT Y-CONNECTOR NO INJURY REPORTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19355625 ↗
  • MalfunctionSep 13, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 3: DESCRIPTION OF THE PATIENT EVENT - PER NURSING: BURETROLS APPEAR TO HAVE SOMETHING IN THE TUBING THAT PRECIPITATES W/NS CREATING A HARD PLUG. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15414358 ↗
  • MalfunctionSep 13, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 1: DESCRIPTION OF THE PATIENT EVENT - PER NURSING: BURETROLS APPEAR TO HAVE SOMETHING IN THE TUBING THAT PRECIPITATES W/NS CREATING A HARD PLUG. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15414354 ↗
  • MalfunctionSep 13, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 4: DESCRIPTION OF THE PATIENT EVENT - PER NURSING: BURETROLS APPEAR TO HAVE SOMETHING IN THE TUBING THAT PRECIPITATES W/NS CREATING A HARD PLUG. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15414353 ↗
  • MalfunctionSep 13, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 2: DESCRIPTION OF THE PATIENT EVENT - PER NURSING: BURETROLS APPEAR TO HAVE SOMETHING IN THE TUBING THAT PRECIPITATES W/NS CREATING A HARD PLUG. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15414352 ↗
  • MalfunctionJun 13, 2022Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: EVENT 3 : FIVE (5) PATIENTS HAVE HAD IV LINES RUPTURES. ASSESSMENT: NICU PATIENTS REQUIRING A LARGE VOLUME CONTINUOUS INFUSION (IVF, TPN ETC.), A SPACE PUMP BURETTE SET IS USED FOR THE IV ASSEMBLY. NICU STAFF HAVE OBSERVED AND RECOGNIZED FREQUENT PUMP ALARMS INDICATING AIR IN THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14674344 ↗
  • MalfunctionJun 13, 2022Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: EVENT 5 : FIVE (5) PATIENTS HAVE HAD IV LINES RUPTURES. ASSESSMENT: NICU PATIENTS REQUIRING A LARGE VOLUME CONTINUOUS INFUSION (IVF, TPN ETC.), A SPACE PUMP BURETTE SET IS USED FOR THE IV ASSEMBLY. NICU STAFF HAVE OBSERVED AND RECOGNIZED FREQUENT PUMP ALARMS INDICATING AIR IN THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14674343 ↗
  • InjuryJun 13, 2022Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: EVENT 1 : FIVE (5) PATIENTS HAVE HAD IV LINES RUPTURES. ASSESSMENT: NICU PATIENTS REQUIRING A LARGE VOLUME CONTINUOUS INFUSION (IVF, TPN ETC.), A SPACE PUMP BURETTE SET IS USED FOR THE IV ASSEMBLY. NICU STAFF HAVE OBSERVED AND RECOGNIZED FREQUENT PUMP ALARMS INDICATING AIR IN THE TUBING. ADDITIONAL INFORMATION ONE NICU PATIENT WITH A SPACE PUMP BURETTE SET WAS RECENTLY FOUND WITH A POSITIVE BC - S. AUREUS ON (B)(6) 2022. THE PATIENT HAD A PICC LINE IN PLACE FROM (B)(6), CST DETERMINED THE CASE TO BE A SECONDARY BSI DUE TO PNEUMONIA. PATIENT BORN ON (B)(6) 2022 WITH RDS, INTUBATED (B)(6) 2022, (B)(6) 2022 IMAGING WITH BILATERAL CONSOLIDATION, RESIDUAL HAZY OPACITIES, WHICH CONTINUED THROUGH ADMISSION. BC COLLECTED ON (B)(6) 2022 DUE TO RECURRENT APNEAS, DESATS, BRADYS. ID CONSULT FROM DR. (B)(6) ON (B)(6) 2022 INDICATED - ABSENCE OF NEUTROPHILIA AND RAPID RESOLUTION OF SYMPTOMATOLOGIC DOES SUGGEST POSSIBLE CONTAMINANT. IF THE REPEAT BC OBTAINED FOLLOWING ONLY SINGLE DOSE OF VANCOMYCIN REMAINS WITHOUT GROWTH, THIS MAY FURTHER SUGGEST POSSIBLE CONTAMINATED SPECIMEN (REPEAT BC ON (B)(6) 2022 NEGATIVE). LOW RISK FOR CNS INFECTION SUGGESTED GIVEN CLINICA

    Patient problems: Unspecified Infection

    MAUDE report 14674342 ↗
  • MalfunctionJun 13, 2022Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: EVENT 2 : FIVE (5) PATIENTS HAVE HAD IV LINES RUPTURES. ASSESSMENT: NICU PATIENTS REQUIRING A LARGE VOLUME CONTINUOUS INFUSION (IVF, TPN ETC.), A SPACE PUMP BURETTE SET IS USED FOR THE IV ASSEMBLY. NICU STAFF HAVE OBSERVED AND RECOGNIZED FREQUENT PUMP ALARMS INDICATING AIR IN THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14674341 ↗
  • MalfunctionJun 13, 2022Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: EVENT 4 : FIVE (5) PATIENTS HAVE HAD IV LINES RUPTURES. ASSESSMENT: NICU PATIENTS REQUIRING A LARGE VOLUME CONTINUOUS INFUSION (IVF, TPN ETC.), A SPACE PUMP BURETTE SET IS USED FOR THE IV ASSEMBLY. NICU STAFF HAVE OBSERVED AND RECOGNIZED FREQUENT PUMP ALARMS INDICATING AIR IN THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14674340 ↗
  • MalfunctionNov 16, 2018Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 2: CUSTOMER STATED WHEN THE NURSES WERE SETTING UP/PRIMING THE IV SET, THERE WAS DIFFICULTY GETTING FLUID TO FLOW INTO THE SET. ANOTHER IV SET WAS RETRIEVED AND THE SAME INCIDENT OCCURRED. THE NURSES THEN NOTICES A PIECE OF CLEAR PLASTIC, APPROXIMATELY 1/4" LONG THAT WAS STUCK INSIDE OF THE TUBING. NO INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8080240 ↗
  • MalfunctionNov 16, 2018Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 1: CUSTOMER STATED WHEN THE NURSES WERE SETTING UP/PRIMING THE IV SET, THEY WAS DIFFICULTY GETTING FLUID TO FLOW INTO THE SET. ANOTHER IV SET WAS RETRIEVED AND THE SAME INCIDENT OCCURRED. THE NURSES THEN NOTICES A PIECE OF CLEAR PLASTIC, APPROXIMATELY 1/4" LONG THAT WAS STUCK INSIDE OF THE TUBING. NO INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8080195 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page