Infusomat® 362033Recalls & reported problems

B. Braun Medical Inc. · UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04046964837821), received between Mar 14, 2024 and Jul 2, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (3 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 29, 2025Open, Classified

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

    Z-0640-2026 · Root cause: Process control · FDA record ↗

  2. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2379-2024 · Root cause: Process control · FDA record ↗

  3. Nov 1, 2023Open, Classified

    In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.

    Z-0355-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Air/Gas in Device3
  • Device Alarm System2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Distress1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 2, 2025Device Alarm System; Air/Gas in Device; Device Alarm System; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: DETAILS OF COMPLAINT (REPORTED ISSUE): BACK TO BACK SHIFTS AS PT STARTED TPN @ 1800 AND INFUSED OVERNIGHT, THERE WAS CONSISTENT AIR BUBBLES AND CONSISTENT ALARMS, TROUBLE SHOOTING TECHNIQUES ATTEMPTED, DESPITE PRIMING ON THE PUMP STILL HAD NUMEROUS AIR BUBBLES, AS WELL AS ATTEMPTS TO CHANGE THE TUBING NUMEROUS TIMES AND CHANGING OUT THE PUMP. ALARM EVENTS AIR IN LINE - NOT VISIBLE AND UNRESOLVED. DELAY IN PT RECEIVING ORDERED AMOUNT OF LIPIDS FOR PARENTERAL NUTRITION. IMPACT TO PATIENT: DELAY IN TREATMENT, EMOTIONAL DISTRESS, INTERRUPTIONS TO CLINICAL CARE DUE TO TIME REQUIRED TO RESOLVE ALARMS, RISK FOR INFECTION / INTRODUCTION OF CONTAMINANTS, UNDER DOSING OF ORDERED MEDICATION / FLUIDS. ACTION(S) TAKEN CHANGED OUT PUMP (ISOLATE AND SEND TO BIOENGINEERING DEPARTMENT), FOLLOWED B BRAUN TIPS TO RESOLVE ALARMS (AIR-IN-LINE AND OCCLUSION), REPEATED PRIMING TO CLEAR AIR, TUBING REPLACED (TRIPLE BAG AND SEND TO MATERIAL MANAGEMENT). MEDICATION / FLUID LIPIDS IV RATE APPROX 12ML.

    Patient problems: Distress

    MAUDE report 22382737 ↗
  • MalfunctionJun 24, 2025Device Alarm System; Air/Gas in Device; Device Alarm System; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: INCIDENT DESCRIPTION: LIPID INFUSION CONTINUED TO RING AIR IN LINE". WRITER USED A 10CC SYRINGE CONNECTED IT TO THE SINGLE PORT AND ATTEMPTED TO PRIME 10MLS OUT TO OBTAIN THE AIR BUBBLE. NO AIR BUBBLE WAS VISUALIZED IN THE LINE. THE PUMP CONTINUED TO RING "AIR IN LINE" AFTER THIS. BEDSIDE NURSE CHANGED OUT B. BRAUN PUMP (USED SAME LIPIDS TUBING). PUMP STILL CONTINUED TO RING "AIR IN LINE". BEDSIDE NURSE THEN CHANGED THE LIPIDS TUBING WHICH SEEMED TO BE EFFECTIVE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22318323 ↗
  • MalfunctionMar 14, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: AIR IN LINE ALARM WITH TPN TUBING SET. DETAILED INQUIRY DESCRIPTION: "TPN TUBING IS ALARMING "AIR IN LINE" OVER AND OVER AND OVER.... IN SPITE OF MULTIPLE ATTEMPTS BY SEVERAL NURSES TO CLEAR THE LINE. WE JUST FIXED THE TRAVASOL LINE, THEN THE LIPID LINE ALARMED. AGAIN. FINALLY REPLACED THE TUBING." NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18904248 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →