Infusomat® 363424Recalls & reported problems

B. Braun Medical Inc. · UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04046964837906), received between Oct 10, 2024 and Oct 10, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (4 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 4 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 29, 2025Open, Classified

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

    Z-0640-2026 · Root cause: Process control · FDA record ↗

  2. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2391-2024 · Root cause: Process control · FDA record ↗

  3. Nov 1, 2023Open, Classified

    In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.

    Z-0355-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Air/Gas in Device4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 10, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED, EVENT 1 : THE STAFF WHO USED THIS PRODUCT CHANGED IT 4 TIMES WITH THE SAME LOT NUMBER. THE USUAL TROUBLESHOOTING WAS FOLLOWED FOR AIR BUBBLES IN THE TUBING, TRIED A DIFFERENT IV PUMP, BUT NO SUCCESS. A DIFFERENT IV TUBING, WITH A DIFFERENT LOT NUMBER RESOLVED THE PROBLEM ON THE FIRST ATTEMPT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20416718 ↗
  • MalfunctionOct 10, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED, EVENT 2: THE STAFF WHO USED THIS PRODUCT CHANGED IT 4 TIMES WITH THE SAME LOT NUMBER. THE USUAL TROUBLESHOOTING WAS FOLLOWED FOR AIR BUBBLES IN THE TUBING, TRIED A DIFFERENT IV PUMP, BUT NO SUCCESS. A DIFFERENT IV TUBING, WITH A DIFFERENT LOT NUMBER RESOLVED THE PROBLEM ON THE FIRST ATTEMPT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20416712 ↗
  • MalfunctionOct 10, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED, EVENT 3: THE STAFF WHO USED THIS PRODUCT CHANGED IT 4 TIMES WITH THE SAME LOT NUMBER. THE USUAL TROUBLESHOOTING WAS FOLLOWED FOR AIR BUBBLES IN THE TUBING, TRIED A DIFFERENT IV PUMP, BUT NO SUCCESS. A DIFFERENT IV TUBING, WITH A DIFFERENT LOT NUMBER RESOLVED THE PROBLEM ON THE FIRST ATTEMPT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20416711 ↗
  • MalfunctionOct 10, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED, EVENT 4, THE STAFF WHO USED THIS PRODUCT CHANGED IT 4 TIMES WITH THE SAME LOT NUMBER. THE USUAL TROUBLESHOOTING WAS FOLLOWED FOR AIR BUBBLES IN THE TUBING, TRIED A DIFFERENT IV PUMP, BUT NO SUCCESS. A DIFFERENT IV TUBING, WITH A DIFFERENT LOT NUMBER RESOLVED THE PROBLEM ON THE FIRST ATTEMPT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20416710 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →