HOPKINS 26105BARecalls & reported problems
Karl Storz GmbH & Co. KG · HOPKINS® Telescope 30°, 4 mm, 30 cm · Class 2
All models of HOPKINS (49)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04048551093597), received between Feb 6, 2020 and Jul 15, 2026 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Insufficient Device Problem Information3
- Optical Distortion3
- Poor Quality Image3
- Crack1
- Optical Problem1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information4
- Full thickness (Third Degree) Burn1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2026Optical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25830820 ↗MalfunctionJun 23, 2026Poor Quality Image
IT WAS REPORTED THAT "THE HYSTEROSCOPE IMAGE WAS BLURRY AFTER BEING CONNECTED TO THE SYSTEM AND COULD NOT MEET SURGICAL IMAGING REQUIREMENTS.". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25600850 ↗MalfunctionMar 19, 2026Optical Distortion
THE FOLLOWING WAS REPORTED: FOGGY IMAGE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24644465 ↗MalfunctionFeb 5, 2025Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information
IT WAS REPORTED THE ROD LENS OF FOUR HYSTEROSCOPES HAS ALREADY BROKEN 4 TIMES. UNFORTUNATELY, THERE IS NO LONGER SUFFICIENT VISIBILITY TO OPERATE. NO DEATH OR UNANTICIPATED SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 9610617-2025-00131 9610617-2025-00133 9610617-2025-00134.
Patient problems: Insufficient Information
MAUDE report 21318177 ↗MalfunctionFeb 5, 2025Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information
IT WAS REPORTED THE ROD LENS OF FOUR HYSTEROSCOPES HAS ALREADY BROKEN 4 TIMES. UNFORTUNATELY, THERE IS NO LONGER SUFFICIENT VISIBILITY TO OPERATE. NO DEATH OR UNANTICIPATED SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: (B)(4).
Patient problems: Insufficient Information
MAUDE report 21318174 ↗MalfunctionFeb 5, 2025Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information
IT WAS REPORTED THE ROD LENS OF FOUR HYSTEROSCOPES HAS ALREADY BROKEN 4 TIMES. UNFORTUNATELY, THERE IS NO LONGER SUFFICIENT VISIBILITY TO OPERATE. NO DEATH OR UNANTICIPATED SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 9610617-2025-00131. 9610617-2025-00132. 9610617-2025-00134.
Patient problems: Insufficient Information
MAUDE report 21318172 ↗MalfunctionFeb 5, 2025Break; Break
IT WAS REPORTED THE ROD LENS OF FOUR HYSTEROSCOPES HAS ALREADY BROKEN 4 TIMES. UNFORTUNATELY, THERE IS NO LONGER SUFFICIENT VISIBILITY TO OPERATE. NO DEATH OR UNANTICIPATED SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 9610617-2025-00132, 9610617-2025-00133 AND 9610617-2025-00134.
Patient problems: Insufficient Information
MAUDE report 21318169 ↗MalfunctionSep 6, 2023Poor Quality Image; Optical Distortion; Poor Quality Image; Optical Distortion
IT WAS REPORTED THAT THE PATIENT HAD CONSENTED TO HYSTEROSCOPY AND TCRF. THE SCOPE HAD BEEN CHECKED BEFORE THE PROCEDURE AND THE VISION WAS CLEAR EVEN DURING HYSTEROSCOPY. AFTER CONNECTING TO BIPOLAR RESECTOSCOPE THE SURGEON REPORTED THAT THE VISION BECAME FOGGY AND BLURRY WHILE IN THE CAVITY. WE CHANGED TO ANOTHER SCOPE WHICH WAS CLEAR AND THE SURGEON WAS HAPPY WITH IT. WHEN HE INSERTED IT IN THE PATIENT'S UTERINE CAVITY THE SURGEON REPORTED THE SECOND SCOPE WAS ALSO FOGGY AND BLURRY. AS BOTH SCOPES WERE NOT CLEAR THE SURGEON DECIDED IT WAS NOT SAFE TO PROCEED AND ABANDONED THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17691480 ↗MalfunctionSep 6, 2023Poor Quality Image; Optical Distortion; Poor Quality Image; Optical Distortion
IT WAS REPORTED THAT THE PATIENT HAD CONSENTED TO HYSTEROSCOPY AND TCRF. THE SCOPE HAD BEEN CHECKED BEFORE THE PROCEDURE AND THE VISION WAS CLEAR EVEN DURING HYSTEROSCOPY. AFTER CONNECTING TO BIPOLAR RESECTOSCOPE THE SURGEON REPORTED THAT THE VISION BECAME FOGGY AND BLURRY WHILE IN THE CAVITY. WE CHANGED TO ANOTHER SCOPE WHICH WAS CLEAR AND THE SURGEON WAS HAPPY WITH IT. WHEN HE INSERTED IT IN THE PATIENT'S UTERINE CAVITY THE SURGEON REPORTED THE SECOND SCOPE WAS ALSO FOGGY AND BLURRY. AS BOTH SCOPES WERE NOT CLEAR THE SURGEON DECIDED IT WAS NOT SAFE TO PROCEED AND ABANDONED THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17691479 ↗MalfunctionSep 5, 2023Crack; Crack
IT WAS REPORTED THAT THE DURING A HOLEP RESECTION THE REUSABLE BIPOLAR LOOP SEPARATED IN THE PATIENT. ACCORDING TO THE INFORMATION RECEIVED, NO HARM TO PATIENT, USER OR THIRD OCCURRED. THE CASE COULD BE COMPLETED SUCCESSFULLY WITH NO DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17686918 ↗InjuryFeb 6, 2020Use of Device Problem; Use of Device Problem
AS PER A VIGILANCE REPORT RECEIVED FROM THE FACTORY IN (B)(4), A PATIENT RECEIVED A BURN ON THE INNER ASPECT OF THE LEFT THIGH THOUGHT TO HAVE OCCURRED BY THE END OF THE HYSTEROSCOPE COMING INTO CONTACT WITH THE PATIENT'S LEG WHEN THE POWER WAS ON. THE LESION WAS ROUGHLY THE SAME SIZE AS THE HYSTEROSCOPE TIP (4 - 5MM). ACCESSING THE ENDOMETRIAL CAVITY WAS DIFFICULT AND THE OPERATORS WERE SWITCHING BETWEEN USING THE HYSTEROSCOPE TO FACILITATE DILATATION, AND HEGAR DILATORS. WHEN THE HYSTEROSCOPE WAS NOT BEING ACTIVELY USED, IT WAS PLACED ON THE PATIENT'S ABDOMEN. THE CONSULTANT WOULD USUALLY ASK A MEMBER OF THE NURSING TEAM IN THEATRE TO TURN THE POWER OFF WHEN THE HYSTEROSCOPE IS NOT BEING ACTIVELY USED, BUT IN THIS CASE, POSSIBLY BECAUSE THERE WAS CONTINUAL SWITCHING BETWEEN MEANS OF DILATING THE CERVIX, THIS DOES NOT SEEM TO HAVE BEEN THE CASE. ONCE THE INJURY WAS IDENTIFIED, THE MEDICAL AND NURSING TEAMS INSPECTED THE DRAPES BUT FOUND NO EVIDENCE OF CHARRING OR BURNING. PATIENT ASSESSED AT THE BURNS CLINIC THE FOLLOWING DAY. THE BURN WAS FOUND TO BE FULL THICKNESS, AND THE PATIENT WENT TO SURGERY THERE FOR EXCISION AND PRIMARY CLOSURE OF THE LESION. SHE WAS DISCHARGED THE FOLLOW
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 9679558 ↗
Other Cystoscope and accessories, flexible/rigid devices
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHOlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls1 death1,380 reports
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHROlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls377 reports
- VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2Olympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid1 recall418 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall353 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall1 death332 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid2 recalls75 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid114 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall54 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page