Karl Storz GmbH & Co. KG · HOPKINS® Telescope 12°, 4 mm, 30 cm · Class 2
All models of HOPKINS (49)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04048551093610), received between Feb 13, 2026 and Aug 25, 2026 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Optical Distortion” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Optical Distortion5
- Chemical Problem2
- Optical Problem2
- Poor Quality Image2
- Device Contaminated at the User Facility1
- Failure to Clean Adequately1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information3
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 25, 2026Optical Problem
New information was received August 10, 2026: The affected material number is 26105FA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50439101 ↗MalfunctionJul 13, 2026Poor Quality Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25801811 ↗MalfunctionJul 9, 2026Optical Distortion
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25769183 ↗InjuryJun 2, 2026Optical Distortion
THE FOLLOWING WAS REPORTED: THE PATIENT ENTERED THE OPERATING ROOM AT 08:40 ON (B)(6) 2026, AND ROUTINE PREOPERATIVE PREPARATION WAS PERFORMED. THE LITHOTOMY POSITION WAS ASSUMED AT 09:00, FOLLOWED BY ROUTINE DISINFECTION AND DRAPING. TRANSURETHRAL VAPORIZATION RESECTION OF THE PROSTATE WAS FORMALLY INITIATED AT 09:23. THE PRIMARY SURGICAL OBJECTIVE WAS TO RESECT HYPERPLASTIC PROSTATIC TISSUE TO RELIEVE THE PATIENT'S URINARY OBSTRUCTION AND RELATED SYMPTOMS AND IMPROVE PROGNOSIS. DURING THE PROCEDURE, WHILE USING THE HIGH-FREQUENCY ELECTROSURGICAL SYSTEM TO RESECT HYPERPLASTIC PROSTATIC TISSUE, THE RESECTOSCOPE DISPLAYED A BLURRED IMAGE WITH GHOSTING SHADOWS, SEVERELY OBSTRUCTING THE SURGICAL FIELD OF VIEW, WHICH LED TO A SURGICAL ERROR CAUSING PROSTATIC BLEEDING IN THE PATIENT, WITH AN ESTIMATED BLOOD LOSS OF APPROXIMATELY 200 ML. THE SURGEON IMMEDIATELY USED THE RESECTOSCOPE LOOP FOR ELECTROCOAGULATION HEMOSTASIS, WHICH WAS COMPLETED AT 10:55. THE PATIENT'S VITAL SIGNS WERE MONITORED TO ENSURE STABILITY. FOLLOWING THE MALFUNCTION, A BACKUP RESECTOSCOPE WAS IMMEDIATELY REPLACED. THE SURGERY CONCLUDED UNEVENTFULLY AT 11:30, AND THE PATIENT WAS SAFELY RETURNED TO THE WARD. DUE TO TH
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25366159 ↗MalfunctionApr 2, 2026Optical Distortion
THE FOLLOWING WAS REPORTED: "OPTICAL SYSTEM FOGGY". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 24769437 ↗MalfunctionMar 26, 2026Optical Distortion; Device Contaminated at the User Facility
THE FOLLOWING WAS REPORTED: "OPTICAL SYSTEM: BLURRED IMAGE. FOGGY." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 24708184 ↗MalfunctionMar 26, 2026Chemical Problem; Optical Distortion
THE FOLLOWING WAS REPORTED: "OPTICAL SYSTEM: FOGGY." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 24707781 ↗MalfunctionMar 4, 2026Poor Quality Image; Chemical Problem; Failure to Clean Adequately
IT WAS REPORTED THAT "OPTICAL SYSTEM: BLURRED IMAGE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24510352 ↗MalfunctionFeb 13, 2026Optical Problem
IT WAS REPORTED THAT DURING THE USE OF THE HYSTEROSCOPIC RESECTOSCOPE, THE MEDICAL STAFF FOUND BLURRY AND ABNORMAL IMAGES ON THE LENS. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24337828 ↗
Other Cystoscope and accessories, flexible/rigid devices
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHOlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls1 death1,380 reports
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHROlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls377 reports
- VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2Olympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid1 recall418 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall353 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall1 death332 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid2 recalls75 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid114 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall54 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.