n.a. 27002KRecalls & reported problems
Karl Storz GmbH & Co. KG · Ureteroscope, 9.5 Fr., 6°, 34 cm · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04048551231227), received between Jun 24, 2026 and Jun 24, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Poor Quality Image” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Dec 19, 2024Class IIOpen, Classified
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
Z-1063-2025 · Root cause: Labeling Change Control · Recall details · FDA record ↗
- Dec 19, 2024Class IIOpen, Classified
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
Z-1062-2025 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Poor Quality Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 24, 2026Poor Quality Image
IT WAS REPORTED: OPTICAL SYSTEM FOGGY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25613791 ↗
FDA clearance and classification
The FDA regulates product code FGB (ureteroscope and accessories, flexible/rigid) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the n.a. 27002K
Has the n.a. 27002K been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1063-2025 (Class II), began on Dec 19, 2024: Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the n.a. 27002K?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jun 24, 2026. That report lists device problem “Poor Quality Image” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the n.a. 27002K?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the n.a. 27002K, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand n.a., model 27002K). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Ureteroscope and accessories, flexible/rigid devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall1,122 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls393 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall303 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls187 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls170 reports
- ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-10-3DOlympus Medical Systems Corp. · Ureteroscope And Accessories, Flexible/Rigid186 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P7ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall121 reports
- n.a. 091279-01Karl Storz GmbH & Co. KG · Ureteroscope and accessories, flexible/rigid111 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page