Flex-X / IMAGE 1 S 11278VSURecalls & reported problems
Karl Storz GmbH & Co. KG · Video Uretero-Renoscope FLEX-XC · Class 2
All models of Flex-X / IMAGE 1 S (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04048551344385), received between Dec 29, 2025 and Sep 2, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 1, 2022Class IIOpen, Classified
Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.
Z-1070-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Apr 1, 2022Class IIOpen, Classified
Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.
Z-1069-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information1
- Mechanical Jam1
- Temperature Problem1
- Unintended Movement1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 2, 2026Mechanical Jam; Unintended Movement
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50571550 ↗MalfunctionDec 29, 2025Temperature Problem; Insufficient Device Problem Information; Temperature Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT THE SCOPE STOPPED WORKING MID CASE. THE SURGEON BELIEVES THIS IS BECAUSE OF THE INCORRECT DEFLECTION MECHANISM. THE CASE HAD TO BE ABANDONED AS THEY DID NOT HAVE A SPARE SCOPE.
Patient problems: Insufficient Information
MAUDE report 23917743 ↗
FDA clearance and classification
The FDA regulates product code FGB (ureteroscope and accessories, flexible/rigid) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the Flex-X / IMAGE 1 S 11278VSU
Has the Flex-X / IMAGE 1 S 11278VSU been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1070-2022 (Class II), began on Apr 1, 2022: Failure to achieve the expected six-log reduction in microorganisms following the disinfection process. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Flex-X / IMAGE 1 S 11278VSU?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Dec 29, 2025 and Sep 2, 2026. In those reports, the most common specific device problems are “Mechanical Jam” (1), “Temperature Problem” (1) and “Unintended Movement” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Flex-X / IMAGE 1 S 11278VSU?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Flex-X / IMAGE 1 S 11278VSU, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand Flex-X / IMAGE 1 S, model 11278VSU). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Ureteroscope and accessories, flexible/rigid devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall1,122 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls393 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall303 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls187 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls170 reports
- ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-10-3DOlympus Medical Systems Corp. · Ureteroscope And Accessories, Flexible/Rigid186 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P7ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall121 reports
- n.a. 091279-01Karl Storz GmbH & Co. KG · Ureteroscope and accessories, flexible/rigid111 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.