TIPCAM1 26605BARecalls & reported problems
Karl Storz GmbH & Co. KG · TIPCAM®1 S 3D LAP, 30° · Class 2
All models of TIPCAM1 (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Karl Storz TIPCAM 1 3D laparoscope is a rigid surgical scope used with a camera head for laparoscopic surgery. Most of the 37 reports describe image problems — flickering or frozen picture, a foggy or distorted view, and generally poor image quality — sometimes interrupting a procedure until the scope or cables were reconnected or replaced. Five injuries and no deaths were reported. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 37
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 04048551347843), received between Oct 4, 2021 and Aug 27, 2026 — classified as 5 injuries, 32 malfunctions.
The most frequently reported specific device problem is “Erratic or Intermittent Display” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Erratic or Intermittent Display12
- Optical Distortion6
- Poor Quality Image5
- Insufficient Device Problem Information4
- Optical Problem4
- No Display/Image3
- Output Problem3
- Device Handling Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions29
- Insufficient Information6
- Bowel Burn2
- Foreign Body In Patient1
Recent adverse event reports
The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 27, 2026Optical Distortion
The following was reported: "On (b)(6) 2026, medical staff of the anesthesiology department encountered the issue of screen bar flickering on the lens while using the 3D high definition electronic laparoscope, which prevented the medical staff from accurately assessing the patient's condition." No negative impact in state of health reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50482791 ↗MalfunctionJul 28, 2026Erratic or Intermittent Display
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50060729 ↗MalfunctionJul 6, 2026Erratic or Intermittent Display
IT WAS REPORTED THAT "3D CAMERA HEAD FLICKERING.". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25720971 ↗MalfunctionJun 22, 2026Optical Distortion
THE FOLLOWING WAS REPORTED: \"FOGGY.\"
Patient problems: Insufficient Information
MAUDE report 25587390 ↗MalfunctionJun 11, 2026Optical Problem
THE FOLLOWING WAS REPORTED: "INTRA-OPERATIVELY, STORZ ENDOSCOPE EXPERIENCED SEVERAL ENDOSCOPE IMAGE FROZEN ISSUES. THE ISSUES WERE RESOLVED AFTER UNPLUGGING THE ENDOSCOPE CABLE AND PLUGGING IT BACK IN. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT INJURY". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25479980 ↗MalfunctionMay 20, 2026Erratic or Intermittent Display
IT WAS REPORTED THAT "DURING A ROBOTIC LAPAROSCOPIC REPAIR OF A SUPRAPUBIC HERNIA PRIOR TO SCOPE INSERTION, THE ENDOSCOPE SHOWED AN INTERMITTENTLY FLASHING IMAGE ON THE OPERATING ROOM TEAM INTERFACE AND THE SPARE MONITOR, BUT NOT ON THE SURGEON CONSOLE. THE DATA CABLE OF THE TOWER WAS DISCONNECTED AND RECONNECTED AND THE EXTENSION CORD WAS ALSO EXCHANGED, BUT NEITHER ACTION RESOLVED THE ISSUE. THE ONLY THING THAT WORKED WAS REPLACING THE SCOPE, AFTER WHICH THE SYSTEM FUNCTIONED PROPERLY. THERE WAS NO PATIENT INJURY.". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25236000 ↗MalfunctionMay 12, 2026Optical Distortion
THE FOLLOWING WAS REPORTED: "THE SCREEN FLASHES AFTER WORKING FOR A WHILE."
Patient problems: Insufficient Information
MAUDE report 25151355 ↗InjuryApr 21, 2026Break
IT WAS REPORTED THAT DURING A ROBOTIC LAPAROSCOPIC HARTMANN'S REVERSAL PROCEDURE, AT THE TIME OF SPLENIC FLEXURE MOBILIZATION, AN IMAGE QUALITY ISSUE SUDDENLY OCCURRED ON THE ENDOSCOPE WHILE THE SURGEON WAS OPERATING. THE IMAGE CHANGED, PROMPTING THE WITHDRAWAL, CLEANING, AND REINSERTION OF THE ENDOSCOPE WAS DONE BUT THE ISSUE PERSISTED. THE STORZ SYSTEM WAS THEN REBOOTED BY TURNING OFF BOTH THE BOTTOM AND TOP UNITS AND TURNED THEM BACK ON, BUT THE ISSUE REMAINED UNRESOLVED. FINALLY, A NEW ENDOSCOPE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. NEW INFORMATION RECEIVED: THERE WAS A PROLONGATION OF PROCEDURE OF 90 MINUTES. DUE TO THAT, A REPORT IS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24945595 ↗MalfunctionApr 20, 2026Optical Distortion
THE FOLLOWING WAS REPORTED: BLURRED IMAGE/ COLOR CAST. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 24932068 ↗MalfunctionApr 20, 2026Optical Distortion
THE FOLLOWING WAS REPORTED: CCD & CMOS / FOGGY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 24932059 ↗MalfunctionApr 11, 2026Optical Distortion
THE FOLLOWING WAS REPORTED: 3D ELECTRONIC ENDOSCOPE FLICKERING AND CABLE BASE FRACTURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24850209 ↗MalfunctionApr 8, 2026Poor Quality Image
IT WAS REPORTED THAT THE IMAGE FROM THE ENDOSCOPE CAMERA WAS FLASHING, AND WITH CERTAIN CONNECTIONS, THE SYSTEM DISPLAYED A MESSAGE INDICATING THAT UNKNOWN EQUIPMENT HAD BEEN CONNECTED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24818896 ↗MalfunctionMar 9, 2026Output Problem
THE FOLLOWING WAS REPORTED: IT WAS REPORTED THAT, DURING USE, THE ENDOSCOPE HAD FAULTY OPTICS AND A DAMAGED CABLE. IT IS BEING EXCHANGED. THERE WAS NO PATIENT INVOLVEMENT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24552791 ↗MalfunctionFeb 17, 2026Fracture; Insufficient Device Problem Information
IT WAS REPORTED THAT THE DEVICE BROKE DURING SURGERY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 24369342 ↗MalfunctionFeb 9, 2026Erratic or Intermittent Display; Insufficient Device Problem Information
IT WAS REPORTED THAT IMAGE WAS FLICKERING, COLOR DISTORTION, AND GARBLED SCREEN; CANNOT BE USED NORMALLY. PROCEDURE WAS COMPLETED SUCCESSFULLY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24299940 ↗MalfunctionFeb 6, 2026Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT DURING THE PARAESOPHAGEAL HERNIA ROBOTIC LAPAROSCOPIC PROCEDURE, THE ENDOSCOPE WAS DIFFICULT TO MANIPULATE LAPAROSCOPICALLY. WHEN PLACING THE 11MM PORT WITH THE OPTICAL OBTURATOR, THE SURGEON COULD NOT SEE THE TISSUES WELL. THERE WAS NO IMPACT TO THE PATIENT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24278561 ↗MalfunctionJan 29, 2026Erratic or Intermittent Display; No Apparent Adverse Event
IT WAS REPORTED THAT THE IMAGE ON THE SCOPE BEGAN FLICKERING DURING PROCEDURE. PROCEDURE WAS COMPLETED SUCCESSFULLY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24211141 ↗MalfunctionJan 29, 2026Poor Quality Image; Scratched Material
IT WAS REPORTED: "SHADOW AND BLURRED IMAGE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24211012 ↗MalfunctionJan 15, 2026Erratic or Intermittent Display; Insufficient Device Problem Information
IT WAS REPORTED THAT "DURING A ROBOTIC LAPAROSCOPIC SURGERY, THE ENDOSCOPE DELIVERED A FLICKERING IMAGE. THERE WAS NO PATIENT INJURY." IT WAS CONFIRMED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY OR PROLONGATION. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24089606 ↗MalfunctionDec 10, 2025Poor Quality Image; Device Damaged Prior to Use; Defective Device; Poor Quality Image; Device Damaged Prior to Use; Defective Device
IT WAS REPORTED THAT THE 3D IMAGE IS TOO DARK AND THAT THE PICTURE CANNOT BE PROVIDED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23770494 ↗MalfunctionNov 19, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT THE IMAGE WAS FLICKERING & FROZEN. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23598259 ↗MalfunctionSep 3, 2025No Display/Image; No Apparent Adverse Event; No Display/Image; No Apparent Adverse Event
IT WAS REPORTED THAT THE TIPCAM HAS NO IMAGE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22956707 ↗MalfunctionAug 22, 2025Erratic or Intermittent Display; Device Handling Problem; Erratic or Intermittent Display; Device Handling Problem
ACCORDING TO THE INFORMATION RECEIVED, DURING A ROBOTIC LAPAROSCOPIC RADICAL NEPHRECTOMY AT THE END OF THE PROCEDURE, THE ENDOSCOPE'S IMAGE WAS NOT VISIBLE AS IT APPEARED STRIPED AND INTERMITTENT. THEN, AN ERROR APPEARED READING "WARNING: ENDOSCOPE NOT SUPPORTED. ENDOSCOPE FUNCTIONS MAY NOT OPERATE CORRECTLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22877643 ↗MalfunctionJun 9, 2025Corroded; Erratic or Intermittent Display; Output Problem; Scratched Material; Corroded; Erratic or Intermittent Display; Output Problem; Scratched Material
IT WAS REPORTED THAT SPORADIC IMAGE DISAPPEARS AFTER MOVEMENT / LIGHT DIMS. CAMERA DOES NOT CONNECT TO LSK TOWER. MESSAGE: HEAD IS INCOMPATIBLE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22167187 ↗MalfunctionMay 16, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT PRE-OPERATIVELY, AS SOON AS THE ENDOSCOPE WAS CONNECTED, THERE WAS A RED WARNING ALARM: "THE ENDOSCOPE IMAGE IS FROZEN". THE ENDOSCOPE WAS CHANGED BUT STILL RECEIVED THE SAME ERROR. THE STORZ WAS REBOOTED BUT STILL THE SAME ERROR. THE SURGEON DECIDED TO CONVERT TO ANOTHER ROBOTIC SYSTEM. THERE WAS NO HARM TO THE PATIENT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFRENCE MDR: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22039482 ↗
FDA clearance and classification
The FDA regulates product code FGB (ureteroscope and accessories, flexible/rigid) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the TIPCAM1 26605BA
Has the TIPCAM1 26605BA been recalled?
No FDA recall record in our data names this exact device identifier (DI 04048551347843), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the TIPCAM1 26605BA?
The FDA's MAUDE database lists 37 adverse event reports for this device, received between Oct 4, 2021 and Aug 27, 2026. In those reports, the most common specific device problems are “Erratic or Intermittent Display” (12), “Optical Distortion” (6) and “Poor Quality Image” (5); the most common patient problems are “Bowel Burn” (2) and “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TIPCAM1 26605BA?
None of the reports are classified as deaths, 5 as injuries and 32 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the TIPCAM1 26605BA, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand TIPCAM1, model 26605BA). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Ureteroscope and accessories, flexible/rigid devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall1,122 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls393 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall303 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls187 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls170 reports
- ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-10-3DOlympus Medical Systems Corp. · Ureteroscope And Accessories, Flexible/Rigid186 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P7ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall121 reports
- n.a. 091279-01Karl Storz GmbH & Co. KG · Ureteroscope and accessories, flexible/rigid111 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page