n.a, 091279-06Recalls & reported problems
Karl Storz GmbH & Co. KG · Video Uretero-Renoscope FLEX-XC1 · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Karl Storz FLEX-XC1 video uretero-renoscope is a flexible or rigid scope used to look inside the urinary tract and kidneys. The 27 reports mostly describe the scope failing to produce a usable image — blank screens, flickering or blacked-out video, lines across the picture, or a pink-tinted image — and sometimes the tip bending mechanism breaking or jamming mid-procedure, forcing doctors to pull the scope out without a view and switch to another one. No deaths and no injuries were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 27
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 04048551428030), received between Mar 19, 2025 and Jun 9, 2026 — classified as 27 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- No Display/Image13
- Optical Problem9
- Erratic or Intermittent Display5
- Display or Visual Feedback Problem2
- Mechanical Jam2
- Break1
- Dent in Material1
- Difficult to Insert1
Patient problems
- No Clinical Signs, Symptoms or Conditions25
- Insufficient Information2
Recent adverse event reports
The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2026Image Display Error/Artifact
ACCORDING TO THE INFORMATION RECEIVED, ITEM DID NOT WORK WHEN PLUGGED INTO THE TC028 ADAPTER. VIDEO WOULD NOT APPEAR ON THE SCREEN FOR LONGER THAN A SECOND. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25447471 ↗MalfunctionMay 22, 2026Optical Problem
IT HAS BEEN REPORTED THE UROLOGIST HAD TO WITHDRAW THE SCOPE WITHOUT ANY IMAGE (BLIND REMOVAL). THEY HAD TO OPEN A DIFFERENT SCOPE (NOT KARL STORZ) TO FINISH THE PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25259511 ↗MalfunctionApr 28, 2026Break; Positioning Problem
IT HAS BEEN REPORTED THE DEFLECTION BROKE DURING PROCEDURE. DEFLECTION POPPED. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25021204 ↗MalfunctionApr 2, 2026No Display/Image
IT WAS REPORTED THAT IMAGE DID NOT COME UP ON THE SCREEN, CHANGED THE ADAPTER AND STILL WOULD NOT COME UP ON THE SCREEN, CHANGED OUT THE SCOPE AND WORKED FINE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24772104 ↗MalfunctionMar 10, 2026Erratic or Intermittent Display
IT WAS REPORTED THAT LINES FLICKERING ACROSS THE SCREEN, AND EVENTUALLY TOOK OVER THE WHOLE SCREEN, THEN THE IMAGE DROPPED. ALL COMPLAINTS ON THIS ISSUE HAS BEEN WITH CLEARPETRA USING HIGH-PRESSURE IRRIGATION. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24561357 ↗MalfunctionMar 6, 2026Poor Quality Image
IT WAS REPORTED THAT THE SCOPE HAD A POOR VIDEO OUTPUT, WITH A PINK HUE ACROSS THE IMAGE. THE CASE WAS DELAYED BY 5 MINUTES. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24531374 ↗MalfunctionFeb 27, 2026Erratic or Intermittent Display
IT WAS REPORTED THAT THERE WAS INTERMITTENT IMAGING WOULD BLACK OUT AND THERE WERE LINES IN THE IMAGE. SE VISUALIZED THE ISSUE. SE HAD THEM TRY THE EBOX ADAPTER AND IT WAS THE SCOPE. THE SURGEON SWITCHED TO A REUSABLE SCOPE TO FINISH THE PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24474861 ↗MalfunctionJan 12, 2026No Display/Image; Optical Problem
IT WAS REPORTED THAT WHEN SCOPE PLUGGED INTO X LINK THERE WAS NO IMAGE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24039342 ↗MalfunctionDec 5, 2025Mechanical Jam; Material Split, Cut or Torn; Mechanical Jam; Material Split, Cut or Torn
ON (B)(6) 2025 ADDITIONAL INFORMATION WAS RECEIVED THAT THE CONFIRMATION WAS RECEIVED THAT THERE WAS NO ABORTION OF THE SURGERY. THE DEVICE WAS CHANGED, AND THE SURGERY WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23729549 ↗MalfunctionNov 7, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT HAS BEEN REPORTED: AS SOON AS THEY PLUGGED THE SCOPE INTO CAMERA BOX, IT HAD A SOLID WHITE SCREEN, NO IMAGE, COULD NOT TELL WHERE THEY WERE AT. SWAPPED ADAPTOR. TRIED RESETTING CAMERA BOX NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23497147 ↗MalfunctionOct 23, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT THERE WAS A NO VIDEO IMAGE. NO PATIENT INVOLVEMENT REPORTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A LESS THAN 5 MIN DELAY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23367448 ↗MalfunctionAug 29, 2025No Display/Image; Optical Problem; No Display/Image; Optical Problem
IT HAS BEEN REPORTED: THE ISSUE BECAME APPARENT DURING THE PROCEDURE WHEN THE IMAGE WAS EXPERIENCING SOME FORM OF INTERFERENCE OR DISCONNECTION. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22928166 ↗MalfunctionAug 29, 2025Optical Problem; Optical Problem
THE FOLLOWING WAS REPORTED: FAULTY IMAGE AND WORKING CHANNEL ISSUES DURING PROCEDURE. THREE SCOPES WERE OPENED AND DIDN'T WORK. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 2027009-2025-02513 AND 2027009-2025-02514.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22927689 ↗MalfunctionAug 29, 2025Optical Problem; Optical Problem
THE FOLLOWING WAS REPORTED: FAULTY IMAGE AND WORKING CHANNEL ISSUES DURING PROCEDURE. THREE SCOPES WERE OPENED AND DIDN'T WORK. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22927688 ↗MalfunctionAug 29, 2025Optical Problem; Optical Problem
THE FOLLOWING WAS REPORTED: FAULTY IMAGE AND WORKING CHANNEL ISSUES DURING PROCEDURE. THREE SCOPES WERE OPENED AND DIDN'T WORK. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 2027009-2025-02513 2027009-2025-02515.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22927687 ↗MalfunctionAug 5, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT AFTER PLUGGED IN THE SCOPE WILL TURN ON BUT IMAGE DID NOT SHOW ON THE SCREEN. THIS HAPPENED DURING A CYSTO/URETHRAL DILATION/LEFT URS/LASER. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22722334 ↗MalfunctionAug 5, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT AFTER PLUGGED IN THE SCOPE WILL TURN ON BUT IMAGE DID NOT SHOW ON THE SCREEN. THIS HAPPENED DURING A CYSTO/URETHRAL DILATION/LEFT URS/LASER. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MEDWATCH 9610617-2025-1373.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22722319 ↗MalfunctionJun 21, 2025Dent in Material; Mechanical Jam; Dent in Material; Mechanical Jam
IT WAS REPORTED THAT THE LASER FIBER NOT GOING INSIDE THE SCOPE DURING SURGERY. THE SURGICAL PROLONGATION WAS BETWEEN 15 AND 60 MINUTES. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22303759 ↗MalfunctionJun 3, 2025Difficult to Insert; Difficult to Insert
ACCORDING TO THE INFORMATION RECEIVED, THE LASER FIBER NOT GOING INSIDE THE SCOPE DURING SURGERY. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22137380 ↗MalfunctionJun 2, 2025Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image
ACCORDING TO THE INFORMATION RECEIVED, FLICKERED TO A BLACK SCREEN, SCOPE WE PLUGGED IN AND IMMEDIATELY HAD 1/2 LIGHT OUTPUT. IMAGE WAS VERY DARK. SWAPPED OUT TO A 2ND SCOPE BEFORE WE STARTED THE CASE. 2ND SCOPE FAILED HALFWAY THROUGH THE CASE AND THEY PUSHED FLUORO THROUGH WORKING CHANNEL WITH SYRINGE. IMAGE STARTED DIGITIZING GREEN AND PURPLE LINES, FLICKERING, AND WITHIN 30 SECONDS WENT BLACK. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22126491 ↗MalfunctionJun 2, 2025Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image
ACCORDING TO THE INFORMATION RECEIVED, FLICKERED TO A BLACK SCREEN, SCOPE WE PLUGGED IN AND IMMEDIATELY HAD 1/2 LIGHT OUTPUT. IMAGE WAS VERY DARK. SWAPPED OUT TO A 2ND SCOPE BEFORE WE STARTED THE CASE. 2ND SCOPE FAILED HALFWAY THROUGH THE CASE AND THEY PUSHED FLUORO THROUGH WORKING CHANNEL WITH SYRINGE....IMAGE STARTED DIGITIZING GREEN AND PURPLE LINES, FLICKERING, AND WITHIN 30 SECONDS WENT BLACK. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22126477 ↗MalfunctionJun 2, 2025Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image
ACCORDING TO THE INFORMATION RECEIVED, FLICKERED TO A BLACK SCREEN, SCOPE WE PLUGGED IN AND IMMEDIATELY HAD 1/2 LIGHT OUTPUT. IMAGE WAS VERY DARK. SWAPPED OUT TO A 2ND SCOPE BEFORE WE STARTED THE CASE. 2ND SCOPE FAILED HALFWAY THROUGH THE CASE AND THEY PUSHED FLUORO THROUGH WORKING CHANNEL WITH SYRINGE. IMAGE STARTED DIGITIZING GREEN AND PURPLE LINES, FLICKERING, AND WITHIN 30 SECONDS WENT BLACK. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22126475 ↗MalfunctionMay 28, 2025No Display/Image; Optical Problem; No Display/Image; Optical Problem
IT WAS REPORTED THAT THE IMAGE INTERFERENCE WAS PRESENT, ONLY LINES ON THE SCREEN. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22097295 ↗MalfunctionMay 27, 2025Optical Obstruction; Optical Obstruction
IT WAS REPORTED THAT DURING A URETEROSCOPY LASER LITHTRIPSY PROCEDURE THE SCOPE OBSTRUCTED IN THE WORKING CHANNEL. NO PATIENT INJURY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22089627 ↗MalfunctionMay 12, 2025No Display/Image; Display or Visual Feedback Problem; No Display/Image; Display or Visual Feedback Problem
IT WAS REPORTED THAT THERE IS A GREEN HORIZONTAL LINE FLICKERING OCCURRING RIGHT WHEN THE SCOPE IS FIRST PLUGGED IN. SOMETIMES IT GETS SO BAD WHERE THE GREEN HORIZONTAL LINES BECOME A FLASH DISRUPTION OF THE IMAGE. THIS HAS HAPPENED TO MULTIPLE SCOPES AND SOME HAVE WORSE FLICKERING OR FLASHING OF THE IMAGE. AT TIMES IT ENDS UP WHERE THE SCOPE COMPLETELY DIES AND IT GOES BACK TO THE BLUE DASHBOARD SCREEN. THEY TRIED DIFFERENT CAMERA BOXES AND DIFFERENT TC028 ADAPTERS. THE SYMPTOMS PERSISTED AND IT'S OCCURRING TO ANOTHER FACILITY WITH SIMILAR LOT CODE. THEY WERE ABLE TO COMPLETE THE PROCEDURE AFTER OPENING A NEW SCOPE WHICH WAS A REUSABLE ONE INSTEAD OF A DISPOSABLE SCOPE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REF: COMPLAINT ID: (B)(4). CROSS REF: COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22005847 ↗
Other Ureteroscope and accessories, flexible/rigid devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall1,125 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls397 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall305 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls190 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls170 reports
- ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-10-3DOlympus Medical Systems Corp. · Ureteroscope And Accessories, Flexible/Rigid186 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P7ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall121 reports
- n.a. 091279-01Karl Storz GmbH & Co. KG · Ureteroscope and accessories, flexible/rigid118 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.