Perseus A500 MK06000Recalls & reported problems
Drägerwerk AG & Co. KGaA · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Dräger Perseus A500 anesthesia machine has 139 adverse event reports, including 3 deaths and 10 injuries. Many describe gas delivery failures, loss of ventilation, frozen screens and pressure drops during anesthesia. There is 1 recall, open since March 2024, because the machine can shut down unexpectedly while running on battery power. Reports rose to 29 over the last year from 18 in the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 139
- Deaths reported
- 3
- Injuries reported
- 9
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 139 adverse event reports that list this exact device (DI 04048675253600), received between Mar 23, 2017 and Sep 22, 2026 — classified as 3 deaths, 9 injuries, 127 malfunctions.
The most frequently reported specific device problem is “Gas Output Problem” (88 reports).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K133886); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Mar 20, 2024Class IOpen, Classified
Unexpected shutdown while operating on battery power.
Z-1512-2024 · Root cause: Labeling design · Recall details · FDA record ↗
Most reported problems
Device problems
- Gas Output Problem88
- Failure to Deliver78
- Decrease in Pressure55
- Intermittent Communication Failure46
- Circuit Failure20
- Inaccurate Delivery14
- Inappropriate or Unexpected Reset8
- Gas/Air Leak7
Patient problems
- No Clinical Signs, Symptoms or Conditions95
- No Consequences Or Impact To Patient32
- Insufficient Information4
- No Known Impact Or Consequence To Patient3
- Cardiac Arrest2
- Respiratory Arrest2
- Awareness During Anesthesia1
- Bradycardia1
Recent adverse event reports
The 25 most recent of 139 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2026Premature Discharge of Battery; Failure to Deliver; Complete Loss of Power
It was reported that the device suddenly switched off while running on battery. No injury reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50853412 ↗MalfunctionMay 27, 2026Gas Output Problem; Decrease in Pressure; Intermittent Communication Failure
IT WAS REPORTED THAT THE INDUCTION OF THE PATIENT IN MANUAL VENTILATION WAS UNEVENTFUL BUT THAT - SOON AFTER SWITCHOVER TO AUTOMATIC VENTILATION, THE ANESTHESIA DEVICE ALARMED, THE SCREEN TURNED BLACK AND VENTILATION STOPPED. AS PER REPORT, THE EMERGENCY GAS SUPPLY HAS WORKED, THE DEVICE WAS REPLACED AND THE EVENT DID NOT RESULT IN HEALTH CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25301791 ↗MalfunctionMay 22, 2026Gas Output Problem; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT THERE WAS A VENTILATOR FAILURE DURING USE. THE DEVICE GENERATED AN ALARM. THERE WAS NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25253136 ↗MalfunctionMay 11, 2026Human-Device Interface Problem; Output Problem
IT WAS REPORTED THAT "THE TOUCHSCREEN MALFUNCTIONED DURING MACHINE OPERATION, MAKING IT IMPOSSIBLE TO ADJUST PARAMETERS." THERE WAS NEITHER AN INDICATION FOR A PATIENT INVOLVEMENT NOR AN INJURY OR HEALTH IMPAIRMENT IN THE REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25134642 ↗MalfunctionMay 11, 2026Gas Output Problem; Decrease in Pressure; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT DURING ANESTHESIA, THE SCREEN FROZE JUST AFTER THE FRESH GAS FLOW WAS INCREASED. THERE WAS NO RESPONSE FROM THE SCREEN. A FEW SECONDS LATER, THE SCREEN TURNED WHITE FOLLOWED BY A RESTART FROM THE DEVICE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25134641 ↗MalfunctionMay 11, 2026Gas Output Problem; Decrease in Pressure; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT DURING ANESTHESIA, A LOUD ALARM SOUND WAS HEARD, LOW AIR AND O2 SUPPLY WAS DISPLAYED. THE SCREEN FROZE AND THERE WAS NO CO2 CURVE, AND THE PATIENT WAS NOT VENTILATED. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25134639 ↗MalfunctionMay 7, 2026Gas Output Problem; Decrease in Pressure; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT THE PERSEUS FAILED DURING USE AND CONTINUED OPERATION AFTER APPROX. 10 SECONDS WITH AN ALARM MESSAGE INDICATING THAT THE FLOWSENSOR WAS NOT CALIBRATED. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25109965 ↗MalfunctionMay 5, 2026Gas Output Problem; Decrease in Pressure; Intermittent Communication Failure
IT WAS REPORTED THAT ON (B)(6) 2026, AT 14:44 PM, DURING OPERATION, THE PERSEUS WAS NOT POSSIBLE TO OPERATE ANYMORE. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25078176 ↗MalfunctionApr 29, 2026Gas Output Problem; Decrease in Pressure; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT THE PERSEUS SCREEN TURNED WHITE AND, SHORTLY AFTERWARDS, AN ERROR MESSAGE APPEARED STATING THAT THE FLOW SENSORS NEEDED TO BE CALIBRATED. THE OPERATION WAS SUCCESSFULLY PERFORMED USING THE PERSEUS. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25027523 ↗MalfunctionApr 24, 2026Gas Output Problem; Decrease in Pressure; Intermittent Communication Failure
IT WAS REPORTED THAT DURING ANESTHESIA, THERE IS SUDDENLY A VERY LOUD ALARM SOUND, AFTER WHICH THE ANESTHESIA MACHINE SHUTS DOWN, THE CONTROL PANEL GOES COMPLETELY BLACK AND THE VENTILATION STOPS. IT IS VERY BRIEF - LESS THAN 2 MINUTES - BEFORE THE ENTIRE SYSTEM RESTARTS AUTOMATICALLY. THE PATIENT HAS NOT LACKED VENTILATION AT ANY TIME. THERE HAVE BEEN NO CHANGES IN THE PATIENT'S MEASURED VALUES. THE PATIENT WAS ENSURED OXYGEN ALL THE TIME. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24979000 ↗MalfunctionApr 22, 2026No Display/Image; Gas Output Problem; Decrease in Pressure; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT THE SCREEN TURNED BLANK, THEN THE DEVICE PERFORMED A WARM START DURING SURGERY. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954038 ↗MalfunctionApr 17, 2026Erratic or Intermittent Display; Sensing Intermittently; Failure to Deliver
IT WAS REPORTED THAT THE VENTILATOR FAILED DURING USE. THERE WAS NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24909217 ↗MalfunctionApr 17, 2026Erratic or Intermittent Display; Sensing Intermittently; Failure to Deliver
IT WAS REPORTED THAT THE VENTILATOR FAILED DURING USE. THERE WAS AN INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24909216 ↗MalfunctionApr 16, 2026Gas Output Problem; Failure to Deliver; Inaccurate Delivery
IT WAS REPORTED: "DURING THE ANAESTHESIA OF AN 18-YEAR-OLD MALE PATIENT IN OPERATION NORTH 2, THE DEPARTMENT OF ANAESTHESIA & SURGERY NORTH, AARHUS UNIVERSITY HOSPITAL, A LOUD WHINING NOISE SUDDENLY EMANATED FROM THE ANAESTHESIA MACHINE, AND THE ENTIRE SCREEN WENT BLACK. THE FAULT LASTED APPROX. 3 SECONDS AND RECURRED TWICE DURING THE ANAESTHESIA. AFTER THE TWO EPISODES, THE SCREEN RETURNED TO NORMAL DISPLAY, ALTHOUGH SOME OF THE FIGURES APPEARED LIGHT GREY. THE MACHINE THEN DISPLAYED A MESSAGE STATING "FLOW SENSOR CALIBRATION REQUIRED". THE DEVICE HAD PREVIOUSLY BEEN TESTED AND USED ON SEVERAL PATIENTS WITHOUT ANY PROBLEMS. DURING THE PERIODS OF MALFUNCTION, THE ANAESTHESIA SYSTEM¿S FUNCTION WAS TEMPORARILY UNAVAILABLE, WHICH COULD POTENTIALLY HAVE RESULTED IN A FAILURE TO DETECT CHANGES IN THE PATIENT¿S CONDITION OR IN THE COMPOSITION OF THE ANAESTHETIC GAS. CONSEQUENCES WERE AVOIDED BECAUSE ACTION WAS TAKEN.".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24902352 ↗MalfunctionApr 14, 2026Gas Output Problem; Decrease in Pressure; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT THE DEVICE SHUT DOWN DURING USE. THE VENTILATOR SCREEN WENT BLUE AND THE VENTILATOR THEN PERFORMED A RESTART AND OPERATED NORMALLY WITH PRE-SET ADJUSTMENTS. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24868790 ↗MalfunctionMar 18, 2026Gas Output Problem; Decrease in Pressure; Reset Problem
IT WAS REPORTED THAT THE DEVICE FAILED DURING USE. AN INTERNAL FG FLOW FAILURE AND A SYSTEM FAILTURE ALARM WERE GENERATED. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24631696 ↗MalfunctionMar 13, 2026Gas Output Problem; Decrease in Pressure; Failure to Deliver; Inaccurate Delivery; Gas/Air Leak
IT WAS REPORTED THAT THE MACHINE VENTILATOR WOULD RUN AND STOP WITH A BREAK IN-BETWEEN. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24588856 ↗InjuryMar 10, 2026Gas Output Problem; Inaccurate Delivery; Misassembled During Installation
IT WAS REPORTED THAT THE VAPORIZER WAS NOT SEATED PROPERLY LOCKED DOWN. THE PATIENT WAS AWAKE DURING PROCEDURE. POST-OP THE PATIENT RECEIVED ADDITIONAL MEDICATION TO ASSIST WITH THE ¿DISCOMFORT/EXPERIENCE¿.
Patient problems: Emotional Changes
MAUDE report 24556179 ↗MalfunctionMar 4, 2026Gas Output Problem; Decrease in Pressure; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT THE PATIENT WAS INTUBATED WITHOUT COMPLICATIONS. MANUAL VENTILATION WAS USED INITIALLY, THEN SWITCHED TO MECHANICAL VENTILATION. SUDDENLY, A BEEPING SOUND WAS HEARD, ACCOMPANIED BY A SIMULTANEOUS FAILURE OF THE VENTILATION AND DISPLAY. THE ALARMING AND EMERGENCY DOSING WERE STILL FUNCTIONAL. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24504532 ↗MalfunctionFeb 23, 2026Gas Output Problem; Decrease in Pressure; Intermittent Communication Failure
IT WAS REPORTED THAT ON (B)(6) AT APPROXIMATELY 2:52 PM, THE PERSEUS RESTARTED UNEXPECTEDLY DURING ANESTHESIA, AFTER APPROXIMATELY 3¿3½ HOURS OF USE. BEFORE RESTARTING, THE PERSEUS PLAYED AN ACOUSTIC ALARM. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24412606 ↗MalfunctionFeb 19, 2026Gas Output Problem; Decrease in Pressure; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT AUTOMATIC VENTILATION WAS INTERRUPTED WHILE THE DEVICE PERFORMED A RESTART. THERE WAS NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24385619 ↗MalfunctionJan 12, 2026Gas Output Problem; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT THE DEVICE SCREEN FROZE AND COULD NOT BE OPERATED DURING A CASE. THE DEVICE EMITTED A LOUD INTERMITTENT HIGH-FREQUENCY BEEPING. THE PATIENT WAS IMMEDIATELY VENTILATED MANUALLY AND A REPLACEMENT DEVICE WAS USED. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24035676 ↗MalfunctionDec 8, 2025Circuit Failure; Gas Output Problem; Failure to Deliver; Circuit Failure; Gas Output Problem; Failure to Deliver
IT WAS REPORTED THAT THE DEVICE STOPPED VENTILATING AND ALARMED FOR ¿TURBOVENT 2 FAILURE.¿ THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23740461 ↗MalfunctionDec 8, 2025Circuit Failure; Gas Output Problem; Failure to Deliver; Circuit Failure; Gas Output Problem; Failure to Deliver
IT WAS REPORTED THAT THE DEVICE STOPPED VENTILATING AND ALARMED FOR ¿TURBOVENT 2 FAILURE.¿ THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23739991 ↗InjuryDec 5, 2025Decrease in Pressure; Inaccurate Delivery; Gas/Air Leak; Decrease in Pressure; Inaccurate Delivery; Gas/Air Leak
IT WAS REPORTED THAT DURING USE ON A PATIENT, THE ANESTHESIA WORKSTATION TRIGGERED A "GAS LEAK" ALARM AND THAT VENTILATION WAS SIGNIFICANTLY IMPAIRED. REPORTEDLY THE OBSERVED MALFUNCTION CAUSED A DELAY IN THE PATIENT'S SURGERY, DISRUPTING THE ORIGINALLY SCHEDULED SURGICAL PROCEDURE BUT NO INJURY OCCURED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23726193 ↗
FDA clearance and classification
Cleared through 510(k) K133886 on Jun 20, 2014 (decision: Substantially Equivalent). Submitted by Draeger Medical GmbH as “Perseus A500”. The FDA review took 182 days. FDA record ↗
The FDA regulates product code BSZ (gas-machine, anesthesia) under 21 CFR 868.5160 as a Class II device. 21 CFR 868.5160 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Recalls under the same FDA submission
These recalls involve devices cleared under K133886. One submission often covers many models, so check the product and lot details in each record.
- Z-2254-2018 · May 24, 2018 · TerminatedPerseus A500 Anesthesia Machine; Cat. no. MK06000 Product V Product Usage: Intended for use in anesthetizing adults, children, and neonates and can be used for automatic and manu
- Z-0360-2016 · Nov 12, 2015 · TerminatedPerseus A500 Anesthesia Workstation, MK06000. Intended for use in anesthetizing adults, children, and neonates and can be used for automatic and manual ventilation, pressure-suppo
Common questions about the Perseus A500 MK06000
Has the Perseus A500 MK06000 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1512-2024 (Class I), began on Mar 20, 2024: Unexpected shutdown while operating on battery power. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Perseus A500 MK06000?
The FDA's MAUDE database lists 139 adverse event reports for this device, received between Mar 23, 2017 and Sep 22, 2026. In those reports, the most common specific device problems are “Gas Output Problem” (88), “Failure to Deliver” (78) and “Decrease in Pressure” (55); the most common patient problems are “Cardiac Arrest” (2), “Respiratory Arrest” (2) and “Awareness During Anesthesia” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Perseus A500 MK06000?
3 reports are classified as deaths, 9 as injuries and 127 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Perseus A500 MK06000 FDA approved?
It was cleared through 510(k) K133886 on Jun 20, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Perseus A500 MK06000 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Perseus A500 MK06000, and is it still sold?
The FDA's device database lists Drägerwerk AG & Co. KGaA as the company responsible for it (brand Perseus A500, model MK06000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Gas-Machine, Anesthesia devices
- Carestation 650 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls3,073 reports
- Carestation 750 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,193 reports
- Aisys CS2Datex-Ohmeda Inc · Gas-machine, anesthesia5 recalls2,029 reports
- Carestation 620 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls1,977 reports
- Avance CS2Datex-Ohmeda Inc · Gas-machine, anesthesia6 recalls1,535 reports
- Flow-c Anesthesia System 6887700Maquet Critical Care AB · Gas-Machine, Anesthesia6 recalls72 reports
- FLOW-i C20 6677200Maquet Critical Care AB · Gas-Machine, Anesthesia4 recalls99 reports
- Flow-e Anesthesia System 6887900Maquet Critical Care AB · Gas-Machine, Anesthesia5 recalls17 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page