Skip to content

LifeStent® Biliary Stent EX080801CDRecalls & reported problems

Bard Peripheral Vascular, Inc. · Lifestent® Biliary Stent, 8mm x 80mm (80cm catheter length) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04049519000411), received between Nov 10, 2016 and Nov 10, 2016 — classified as 1 malfunction.

The most frequently reported specific device problem is “Positioning Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Positioning Failure1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2016Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT DURING STENTING OF A PRE-DILATED LESION IN THE AXILLIARY VEIN VIA ACCESS THROUGH THE LOOP OF A GRAFT, THE BILIARY STENT COULD NOT BE DEPLOYED BY ROTATING THE THUMBWEEL OF THE DEPLOYMENT MECHANISM. REPORTEDLY, THERE WAS NO PARTIAL RELEASE OF THE STENT AND THE DELIVERY SYSTEM COULD BE RETRACTED WITHOUT ISSUES. ANOTHER STENT OF THE SAME BRAND WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6094084 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →