LifeStar® Biliary Stent VIUS14060Recalls & reported problems
Bard Peripheral Vascular, Inc. · LifeStar® Biliary Stent 14 mm x 60 mm (80 cm delivery catheter) · Class 2 · Implant
All models of LifeStar Biliary Stent (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04049519005690), received between Feb 19, 2019 and May 18, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Entrapment of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Entrapment of Device1
- Fracture1
- Material Deformation1
- Material Twisted/Bent1
Patient problems
- Great Vessel Perforation1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 18, 2020Fracture; Material Twisted/Bent; Fracture; Material Twisted/Bent
IT WAS REPORTED THAT APPROXIMATELY THREE YEAR AFTER STENT PLACEMENT IN THE RENAL VEINS, THE STENT ALLEGEDLY KINKED AND PROTRUDED THROUGH THE VEIN WALL. IT WAS FURTHER REPORTED THAT THE STENT HAD TO BE REMOVED ALONG WITH SECTIONS OF THE RENAL VEIN DUE TO THE DAMAGE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Great Vessel Perforation
MAUDE report 10067855 ↗MalfunctionFeb 19, 2019Entrapment of Device; Material Deformation; Entrapment of Device; Material Deformation
IT WAS REPORTED THAT AFTER SUCCESSFULLY IMPLANTING THE BILIARY STENT TO TREAT STENOSIS IN A CURVED RIGHT INNOMINATE VEIN VIA ACCESS THROUGH A RIGHT BRACHIOBASILIC FISTULA AND UPON REMOVAL OF THE POST DILATATING PTA BALLOON, THE STENT ALLEGEDLY WAS PULLED BACK WITH THE BALLOON AND COLLAPSED. IT WAS FURTHER REPORTED THAT A STENT GRAFT WAS USED TO COVER THE COLLAPSED STENT AND BALLOON DILATATION WAS PERFORMED TO APOSE THE STENTS TO THE VESSEL WALL. ANOTHER DEVICE WAS USED TO COVER THE TARGET LESION AND COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8351315 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page