LifeStent® Vascular & Biliary Stent System EX050403CSRecalls & reported problems
Bard Peripheral Vascular, Inc. · BARD® LIFESTENT® Vascular and Biliary Stent System 5 mm x 40 mm (130 cm delivery catheter) · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04049519007861), received between Feb 26, 2016 and Nov 6, 2020 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Device Contamination with Chemical or Other Material1
- Difficult to Remove1
- Failure to Advance1
- Material Deformation1
- Positioning Problem1
- Sticking1
Patient problems
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 6, 2020Failure to Advance; Positioning Problem; Failure to Advance; Positioning Problem
IT WAS REPORTED THAT DURING TREATMENT IN THE LEFT ILIAC ARTERY, THE STENT ALLEGEDLY FAILED TO ADVANCE. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10799327 ↗MalfunctionSep 4, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT AN ALLEGED HAIR WAS IDENTIFIED IN THE PACKAGING. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 7842998 ↗InjuryFeb 26, 2016Break; Difficult to Remove; Material Deformation; Break; Difficult to Remove; Sticking; Material Deformation
IT WAS REPORTED THAT A STENT WAS SUCCESSFULLY DEPLOYED TO TREAT A FOCAL 90% STENOSIS IN THE DISTAL SFA/PROXIMAL POPLITEAL. DURING THE ATTEMPT TO REMOVE THE DELIVERY SYSTEM THE TIP OF THE DELIVERY SYSTEM GET CAUGHT ON THE STENOSIS. IN ORDER TO REMOVE THE DELIVERY SYSTEM THE USER PULLED HARDER BUT THE ARTERY MOVED AND THE USER WAS CONCERNED ABOUT TEARING THE VESSEL. THEREFORE, A SECOND WIRE WAS ADVANCED AND AN ANGIOPLASTY BALLOON WAS INFLATED AT THE STENT WAIST TO CREATE SPACE. FURTHERMORE, THE HAND PIECE OF THE DELIVERY SYSTEM WAS REMOVED TO DISENGAGE THE DELIVERY SYSTEM. ULTIMATELY, A LONG 7 FR SHEATH WAS ADVANCED INTO THE CEPHALAD PORTION OF THE STENT TO PROVIDE COUNTER PRESSURE AS THE DELIVERY SYSTEM WAS PULLED OUT. THE PROCEDURE WAS COMPLETED WITH THE USE OF AN ADDITIONAL ANGIOPLASTY BALLOON AND A COVERED STENT TO TREAT THE REMAINING STENOSIS.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 5462989 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page