E·Luminexx® Biliary Stent ZBM12040Recalls & reported problems
Bard Peripheral Vascular, Inc. · E·Luminexx® Biliary Stent 12 mm x 40 mm (80 cm delivery catheter) · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04049519010366), received between Feb 7, 2017 and Mar 9, 2026 — classified as 1 death, 4 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Detachment of Device or Device Component1
- Device Contamination with Chemical or Other Material1
- Difficult or Delayed Positioning1
- Failure to Advance1
- Fracture1
- Malposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- Death1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 9, 2026Difficult or Delayed Positioning; Malposition of Device
ON (B)(6) 2026, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING AN E-LUMINEXX VASCULAR STENT. IT WAS REPORTED THAT DURING THE PROCEDURE, THE STENT WAS DIFFICULT TO DEPLOY. IT WAS FURTHER REPORTED THAT THE STENT ALLEGEDLY BECAME JAMMED IN THE DELIVERY SYSTEM AND SUBSEQUENTLY JUMPED OUT, MISSING THE INTENDED TARGET LOCATION. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24543274 ↗MalfunctionJan 5, 2021Fracture; Migration or Expulsion of Device; Fracture; Migration or Expulsion of Device
IT WAS REPORTED THAT POST STENT PLACEMENT, THE TANTALUM MARKER WAS ALLEGEDLY FRACTURED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11116231 ↗MalfunctionNov 30, 2017Failure to Advance; Device Contamination with Chemical or Other Material; Failure to Advance; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT DURING DEPLOYMENT PROCEDURE A BILIARY STENT ALLEGEDLY COULD NOT BE ADVANCED BECAUSE A PIECE OF MELTED GLUE OR PLASTIC WAS ALLEGEDLY FOUND ATTACHED TO THE OUTER SURFACE. REPORTEDLY, THE DEVICE WAS REMOVED WITHOUT ANY ISSUES AND ANOTHER STENT SYSTEM WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7072854 ↗DeathOct 23, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE STENT PLACEMENT PROCEDURE IN THE INFRA-RENAL ABDOMINAL ARTERY AN AORTIC LEAK DEVELOPED, AND DESPITE RESUSCITATION ATTEMPTS, THE PATIENT EXPIRED AFTER BEING TRANSFERRED TO ANOTHER FACILITY FOR VASCULAR INTERVENTION.
Patient problems: Death
MAUDE report 6970631 ↗MalfunctionFeb 7, 2017Break; Positioning Failure; Detachment of Device or Device Component; Break; Positioning Failure; Detachment of Device or Device Component
IT WAS REPORTED THAT THE WHITE END PIECE OF THE DELIVERY SYSTEM ALLEGEDLY BROKE BEFORE THE STENT WAS DEPLOYED IN THE LEFT BRACHIOCEPHALIC VEIN VIA ACCESS THROUGH THE LEFT ARM. REPORTEDLY, THE VESSEL HAD BEEN PRE-DILATED AND THE USER HAD DIFFICULTIES IN ADVANCING THE DELIVERY SYSTEM TO THE LESION SITE. THE DELIVERY SYSTEM WAS REMOVED WITHOUT ISSUE AND ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6307042 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page