E-LINE 8654204Recalls & reported problems
Richard Wolf GmbH · WORKING ELEMENT ACTIVE MONO 12/30° open, compatible with telescopes Ø 4mm, bayonet-cap, reusable · Class 2
All models of E-LINE (19)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04055207018712), received between Feb 22, 2019 and Feb 22, 2019 — classified as 1 injury.
The most frequently reported specific device problem is “Temperature Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Temperature Problem1
Patient problems
- Full thickness (Third Degree) Burn1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryFeb 22, 2019Temperature Problem; Temperature Problem
ON NOVEMBER 07, 2018, RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) RECEIVED A REPORT FROM RICHARD WOLF (B)(4) INVOLVING THE FOLLOWING DEVICES: 815.052 LOT 523/573 (MDR 9611102-2018-00015), 815.032 LOT 006/086 (MDR 9611102-2018-00016), 8674.205 LOT 1259038 (MDR 9611102-2018-00017), 8680.205 LOT 1255367 (MDR 9611102-2018-00018), 8654.204 LOT 1202777 (THIS MDR). IT WAS REPORTED THAT THE PATIENT SUFFERED THIRD DEGREE BURN ON THE RIGHT FINGER DUE TO HF APPLICATION. THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE INSPECTION SHOWED THAT THE CAUSE OF THE FAULT DESCRIBED BY THE USER CANNOT BE TRACED BACK TO THE WORKING ELEMENTS OR HF CONNECTION CABLE MONO. PROBABLE ROOT CAUSE WAS DETERMINED TO BE USE ERROR. NO DEFECTS WERE FOUND IN THE MEDICAL DEVICES. IN CONCLUSION, THE COMBUSTION WAS NOT CAUSED BY A FAULTY HF DEVICE OR ACCESSORY.
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 8363755 ↗
Other RESECTOSCOPE, WORKING ELEMENT devices
- S-LINE 8680224Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT7 reports
- E-LINE 86553841Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT7 reports
- S-LINE 8680225Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT5 reports
- S-LINE 8680204Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT5 reports
- S-LINE 8680205Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT1 death5 reports
- NA 8654282Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT3 reports
- PRINCESS 8653018Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT2 reports
- SHARK 8674225Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page