COBRA VISION 7356071Recalls & reported problems

Richard Wolf GmbH · SENSOR-URETERORENOSCOPE 9.9FR WL 680MM flexible, working channel 1: 3.6 Fr, working channel 2: 2.4 Fr, double-slanted distal tip Ø 5.2 Fr, TL 980 mm, deflection +270°/-270°, laser shifter for fixing and guiding the laser fiber, locking lever for distal tip, control lever to distal, downward DEF, two function buttons, for connection to ENDOCAM Logic, LED illumination, distal sensor, Fixfocus · Class 2

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04055207054864), received between Sep 11, 2018 and Mar 25, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Material Fragmentation2
  • Mechanical Jam2
  • Blocked Connection1
  • Connection Problem1
  • Display or Visual Feedback Problem1
  • Material Integrity Problem1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient2
  • Insufficient Information1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2025Break; Break

    ACCORDING TO THE INFORMATION RECEIVED, THE DEVICE BROKE WHEN THE PHYSICIAN WAS TRYING TO ADVANCE THE FLEXIBLE WORKING CHANNEL INTO THE PATIENT'S NARROW URETERAL ORIFICE. DUE TO THE NARROWED URETER, THE PRIMARY AND THE BACKUP DEVICE COULD NOT BE USED. ALSO, SINCE DOCTOR COULD NOT GET ANY ACCESS TO THE URINE OUTPUT (UO), THE SCHEDULED PROCEDURE WAS NOT COMPLETED AND NEEDED TO BE RESCHEDULED. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21688532 ↗
  • MalfunctionOct 1, 2022Break; Material Fragmentation; Break; Material Fragmentation

    RWMIC REFERENCE COMPLAINT NO. (B)(4). THE PURPOSE OF THIS SUBMISSION IS TO UPDATE MDR WITH PATIENT INFORMATION AND ADDITIONAL PROCEDURAL INFORMATION. NEW INFORMATION ADDED BY THE USER FACILITY (SOURCE: RICHARD WOLF QUESTIONNAIRE, FORM 5206 (03); RECEIVED 10/03/2022). PURPOSE OF PROCEDURE: TREATMENT WHAT WAS THE PROCEDURE? CYSTO, URETEROSCOPY, LITHOTRIPSY WITH HOLMIUM LASER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15525696 ↗
  • MalfunctionOct 1, 2022Break; Material Fragmentation; Break; Material Fragmentation

    RWMIC REFERENCE COMPLAINT (B)(4). ON SEPTEMBER 9, 2022, IT WAS REPORTED TO RICHARD WOLF MEDICAL INSTRUMENTS CORP. (RWMIC) BY A USER FACILITY REPRESENTATIVE REGARDING PRODUCT ID# 7356071 (B)(4). "THE PLASTIC WAS FREE-FLOATING IN THE KIDNEY. PHYSICIAN DOES BELIEVE THAT HE COLLECT ALL OF THE PIECES AND THE ENTIRE ROOM STAFF; INCLUDING MYSELF PUT OUR EYES ON IT TO AID IN HIS RETRIEVAL. AFTER I GOT THE PIECES I DID ACTUALLY PUT THEM ON THE SPOT AT THE TIP OF THE SCOPE, BEFORE THEY FELL APART THEY DID FIT." ADDITIONAL INFORMATION REPORTED BY THE USER FACILITY REPRESENTATIVE: (EMAIL, DATED SEPTEMBER 20, 2022) WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED ON A PATIENT WHEN THE REPORTING ISSUE OCCURRED? YES. WAS THERE ANY INJURY OR ILLNESS TO THE PATIENT DUE TO THE REPORTED ISSUE? NO. WAS THERE ANY INJURY OR ILLNESS TO ANY OTHER PERSONNEL DUE TO THE REPORTED ISSUE? NO. DID THE ISSUE CAUSE A DELAY IN THE PROCEDURE BEING PERFORMED? YES. DID THE DELAY PUT THE PATIENT AT RISK? YES. WAS THERE A SIMILAR BACK-UP DEVICE AVAILABLE FOR USE? YES. WAS THE SCHEDULED PROCEDURE COMPLETED? YES. NEW INFORMATION OBTAINED FROM THE USER FACILITY REPRESENTATIVE: (EMAIL, DATED SEPTEMBER 22, 2022) H

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15523744 ↗
  • MalfunctionJan 11, 2022Vibration; Vibration

    THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF ALL 2016-2017 COMPLAINT FILES THAT RWMIC IS COMPLETING. RECENTLY THE COMPLAINT SPECIALIST DETERMINED A RE-ASSESSMENT OF THE MDR DECISION FOR THIS COMPLAINT WAS NEEDED AND THAT FOR THIS COMPLAINT, DUE TO A DELAY IN THE PROCEDURE BEING PERFORMED THAT RESULTED IN PATIENT BEING AT RISK, AN MDR SHOULD BE FILED. PREVIOUSLY IN 2017, THE COMPLAINT SPECIALIST AT THAT TIME, DETERMINED AN MDR WAS NOT NEEDED DUE TO "NO INJURY TO PATIENT OR STAFF. BACKUP DEVICE READILY AVAILABLE FOR USE. MINIMAL DELAY, MINIMAL RISK TO PATIENT. PROCEDURE COMPLETED AS SCHEDULED." FACTS OF THE CASE AS REPORTED TO RICHARD WOLF BY THE USER FACILITY: "DISTAL TIP OF SCOPE HIGHLY SENSITIVE TO VIBRATIONS OF THE LASER, EVENT TURNING DOWN THE LASER "WATTAGE" THE IMAGE DID NOT IMPROVE. ADDITIONAL INFORMATION: WILL THE DEVICE BE RETURNED? YES WAS THE DEVICE BEING USED DURING A PROCEDURE WHEN THE ISSUE OCCURRED? YES SPECIFICALLY, WAS THE DEVICE BEING USED ON A PATIENT WHEN THE ISSUE OCCURRED? YES WAS THERE ANY INJURY OR ILLNESS TO PATIENT OR OTHER PERSONNEL DUE TO ISSUE? NO DID THE ISSUE CAUSE A DELAY IN THE PROCEDURE BEING PERFORMED THAT PUT THE PATIENT AT RIS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13232269 ↗
  • MalfunctionFeb 9, 2021Material Integrity Problem; Material Integrity Problem

    THE FOLLOWING WAS REPORTED TO RW (B)(4): THE 7356071 SENSOR-URETEROROSCOPE 9.9CH NL 680MM GOT STUCK IN THE ACCESS SLEEVE WHEN IT WAS FIRST USED. THE PROBLEM WAS THAT THE SLEEVE WAS LOOSE AT THE DISTAL END AND APPARENTLY TOO LONG TO BE ROLLED UP . THIS POSED QUITE A PROBLEM DURING THE PROCEDURE. FURTHER DETAILED INFORMATION WAS NOT PROVIDED BY THE USER, DESPITE THREE REQUESTS FROM RICHARD WOLF.

    Patient problems: Insufficient Information

    MAUDE report 11302910 ↗
  • MalfunctionAug 29, 2019Mechanical Problem; Material Puncture/Hole; Mechanical Problem; Material Puncture/Hole

    THE PURPOSE OF THIS SUBMISSION IS TO REPORT THE RESULTS OF THE DEVICE INVESTIGATION. ACCORDING TO THE INVESTIGATION COMPLETED BY RICHARD WOLF: PRODUCT APPEARANCE: USED. COMPLAINT CONDITION VERIFIED: YES. REPEAT REPAIR: NO. MET SPECIFICATIONS: NO (SEE PHYSICAL FINDINGS). TEST METHOD: OTHER (SEE EQUIPMENT USED). EQUIPMENT USED: CONTROLLER TYPE 5225108, LEAK TESTER. PROBABLE ROOT CAUSE: USER ERROR/MISUSE. COMPLAINT VALID: NO, HANDLING ERROR DURING LASER APPLICATION. ACTION REQUIRED: ROUTINE MAINTENANCE. OTHER: REPAIR EXCHANGE RECOMMENDED. AS WRITTIN IN INSTRUCTION MANUAL GA-D370 CHAPTER 7.6 "DO NO WORK OUTSIDE THE SCOPES FIELD OF VIEW!". PHYSICAL FINDINGS: THE DEFLECTION, THE LASER CHANNEL AND THE DISTAL TUBE OF THE INSTRUMENT HAVE BEEN DAMAGED BY LASER ENERGY. CONCLUSION: THE CAUSE OF THE DAMAGES IS DUE TO A HANDLING ERROR DURING LASER APPLICATION. USER DIDN'T FOLLOW INSTRUCTION MANUAL GA-D370 CHAPTER 7.6. ADDITIONAL INFORMATION FROM THE MANUFACTURER: "THE DAMAGE DESCRIBED HERE IS AS FOLLOWS ARE CAUSED BY INCORRECT HANDLING DURING LASER APPLICATION CAUSED BY THE FACT THAT A LASER FIBER IS NOT PRESENT IN THE ENDOSCOPIC VISUAL RANGE HAS BEEN ACTIVATED, IN THE GA-D375 CHAPTER 7.5 THE US

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 8947691 ↗
  • MalfunctionAug 23, 2019Blocked Connection; Mechanical Jam; Blocked Connection; Mechanical Jam

    ON JULY 24, 2019, THE USER FACILITY REPORTED THE FOLLOWING TO RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC): DURING A PROCEDURE, "200 TRAC TIP" FIBER WAS ADVANCED THROUGH BOTH WORKING CHANNELS AND IT GOT HUNG UP AND WOULD NOT ADVANCE ALL THE WAY. WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED DURING A PROCEDURE WHEN THE ISSUE OCCURRED? YES. SPECIFICALLY, WAS THE DEVICE BEING USED ON A PATIENT WHEN THE ISSUE OCCURRED? YES. WAS THERE ANY INJURY OR ILLNESS TO PATIENT OR OTHER PERSONNEL DUE TO ISSUE? UNKNOWN. DID THE ISSUE CAUSE A DELAY IN THE PROCEDURE BEING PERFORMED THAT PUT THE PATIENT AT RISK? YES. WAS THERE A SIMILAR BACK-UP DEVICE AVAILABLE FOR USE? YES. WAS THE SCHEDULED PROCEDURE COMPLETED? YES. HOW WAS THE PATIENT ANESTHETIZED? GENERAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 8927182 ↗
  • MalfunctionAug 19, 2019Connection Problem; Mechanical Jam; Connection Problem; Mechanical Jam

    THE PURPOSE OF THIS SUBMISSION IS TO REPORT THE RESULTS OF THE DEVICE INVESTIGATION. ACCORDING TO THE INVESTIGATION COMPLETED BY RICHARD WOLF: PRODUCT APPEARANCE: OTHER: NO PROBLEM FOUND. COMPLAINT CONDITION VERIFIED: NO, NO PROBLEM FOUND. REPEAT REPAIR: NO, NO PROBLEM FOUND. MET SPECIFICATIONS: YES. TEST METHOD: VISUAL, FUNCTIONAL. EQUIPMENT USED: LEAK RESTER, CAMERA SETUP. PROBABLE ROOT CAUSE: OTHER, NO PROBLEM FOUND. COMPLAINT VALID: NO PROBLEM FOUND. ACTION REQUIRED: ROUTINE MAINTENANCE. PHYSICAL FINDINGS: NO PROBLEM FOUND. CONCLUSION: SCOPE PASS ALL TESTS. THE USER IS INFORMED IN THE RELATED IFU GA-D 370 / USA / 2015-04 V3.0 / ECO 2015-0069 ABOUT THE CAUTIONS AND FUNCTIONAL CHECKS OF THE DEVICE: CAUTION! TO AVOID DAMAGE TO THE LASER FIBER CHANNEL (17), - DO NOT USE FORCE TO INSERT OR WITHDRAW THE LASER FIBER (X) INTO/FROM THE LASER FIBER CHANNEL (18). - INSERT AND WITHDRAW THE LASER FIBER (X) THROUGH THE LASER FIBER CHANNEL ONLY IF THE INSTRUMENT SHEATH (5) IS STRAIGHT AND THE INSTRUMENT TIP (4) IS STRAIGHT (17). ANGLE (DEFLECT) THE URS ONLY WHEN THE LASER FIBER (X) IS CORRECTLY POSITIONED. CAUTION! TO AVOID DAMAGE TO THE WORKING CHANNEL (1) AND THE BIOPSY VALVE SET (10), - DO

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 8908088 ↗
  • MalfunctionSep 11, 2018No Display/Image; Display or Visual Feedback Problem; Display or Visual Feedback Problem

    RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION, HEREINAFTER REFERRED TO AS RWMIC WAS NOTIFIED OF AN EVENT THAT OCCURRED AT THE HOSPITAL, ON AN UNSPECIFIED DATE, DURING A PATIENT PROCEDURE. THE PICTURE ALLEGEDLY FROZE UP WHILE SCOPE WAS IN THE URETER. A BACK UP DEVICE WAS AVAILABLE AND THE PROCEDURE WAS COMPLETED. THERE WAS A DELAY IN THERAPY THAT PUT THE PATIENT AT RISK BUT THERE WAS NO ADVERSE OUTCOME REPORTED AS A CONSEQUENCE OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7866793 ↗

Other Ureteroscope and accessories, flexible/rigid devices