Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04056702006273), received between May 28, 2020 and Sep 5, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
1 other recall cover devices cleared under the same FDA submission (K173489); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Insufficient Device Problem Information1
- Self-Activation or Keying1
Patient problems
- Foreign Body In Patient1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 5, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
WE HAVE BEEN INFORMED OF THE FOLLOWING EVENT: "ACCORDING TO THE REPORTER, THE UNIT HAD DEFICIT CALCULATION ISSUES. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION 1.HOW WAS THE ISSUE NOTICED? DURING SET UP. 2.COULD WE PLEASE GET MORE INFORMATION, IF THERE WAS ISSUE WITH LOW DEFICIT OR HIGHER DEFICIT? LOW DEFICIT. AS THE COMPLAINT IS REGARDING THE DEFICIT, WHICH COULD BE RELATED TO AN ISSUE WITH SCALE. 3.COULD YOU PLEASE PROVIDE SN OF THE SCALE AND SEND BACK ALSO THE SCALE BACK AS WELL? REP SENT IT OFF."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20143351 ↗InjurySep 18, 2020Incorrect, Inadequate or Imprecise Result or Readings
WE HAVE BEEN INFORMED OF THE FOLLOWING EVENT: FROM: (B)(6), (B)(6) 2020. "ACCORDING TO THE REPORTER, DURING HYSTEROSCOPY, THERE WAS A DEFI-CIT READ 1800 AND DROPPED RAPIDLY TO 0 FOR NO APPARENT REASON. A SMALL AMOUNT OF FLUID COMING FROM UNDER BUTTOCKS DRAPE, THAT WAS BEING SUCTIONED TO THE CANISTERS IN THE STAND WAS NOTED. BUT NOT ENOUGH TO CAUSE A LARGE CHANGE IN DEFICIT. THE SYSTEM WERE RE-PRIMED WITH NEW TUBING AFTER THEY NOTICED THE QUICK AND UNWAR-RANTED CHANGE IN DEFICIT. THEY WERE UNABLE TO FINISH THE CASE AND NEEDED TO BRING BACK THE PATIENT TO REMOVE THE REST OF THE TISSUE DUE TO DEFICIT INACCURACY. ANOTHER HYSTEROSCOPY WAS PERFORMED." FROM: (B)(6), (B)(6) 2020. "AT THIS POINT THE COMPLAINT IS CLOSED AS A NO SAMPLE ON OUR SIDE. THE COMMUNICATIONS TASK IS CLOSED AND WE HAVE NOT RECEIVED THE PRODUCT BACK. WE HAVE REQUESTED IT TO COME BACK BUT IT HAS NOT BEEN RECEIVED. IF IT IS RECEIVED WE WILL REOPEN THE COMPLAINT AND PROCESS THE UNIT. IT IS NOT TAGGED AS STILL IN USE AS WE WERE NOT ABLE TO VERIFY THAT, AND WE HAVE NO EVIDENCE THAT IT WAS SENT FROM THE CUSTOMER, AT THIS POINT."
Patient problems: Insufficient Information
MAUDE report 10552465 ↗InjuryMay 28, 2020Fluid/Blood Leak; Self-Activation or Keying
WE HAVE BEEN INFORMED OF THE FOLLOWING EVENT: "ACCORDING TO THE REPORTER, DURING A HYSTEROSCOPY, THE MACHINE WAS WORKING NORMALLY AND THEN SWITCHED TO DIAGNOSTIC MODE IN THE MIDDLE OF THE CASE. SINCE THERE WERE NO ALARMS OF HIGH DEFICIT, THE UNIT KEPT GOING AND THEY NOTICED BY DOING MANUAL COUNT (B)(4) WAS MISSING AND PRESUMED TO BE INSIDE THE BODY. AFTER REBOOTING THE MACHINE MANY TIMES, THE UNIT WOULD WORK AND MOVED TO DIAGNOSTIC MODE AND THEN THEY WOULD GET A SCALES DEFECTIVE SAYING. THE SCALE TEST WAS STILL 0. THE SURGICAL TIME EXTENDED 30 MINUTES AND PROCEDURE WAS ABORTED. THE (B)(4) FLUID WAS NOT REMOVED AND WAS ABSORBED TO THE BLOOD STREAM. THE PATIENT HAD AN EXTENDED 2-NIGHT STAY HOSPITALIZATION AND WAS DOING WELL. "
Patient problems: Foreign Body In Patient
MAUDE report 10095937 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K173489. One submission often covers many models, so check the product and lot details in each record.
- Z-2561-2021 · Jun 18, 2021 · Open, ClassifiedCovidien HysteroLux Fluid Management System Control Unit, REF: 72205000
Other Insufflator, Hysteroscopic devices
- Fluent Fluid Management System FLT-112SHologic, Inc. · INSUFFLATOR, HYSTEROSCOPIC140 reports
- Aquilex® Fluid Control System AQL-100CBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall18 reports
- Aquilex Fluid Control System AQL-100PBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall8 reports
- Aquilex® Fluid Control System AQL-100CSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic15 reports
- ENDOMAT UP210Karl Storz GmbH & Co. KG · Insufflator, hysteroscopic4 deaths15 reports
- Fluent Console FLT-100Hologic, Inc. · INSUFFLATOR, HYSTEROSCOPIC1 death13 reports
- Hysteroscopic Fluid Management System Stand 7210165Smith & Nephew, Inc. · INSUFFLATOR, HYSTEROSCOPIC9 reports
- Hysteroscopic Fluid Management 7210164Smith & Nephew, Inc. · INSUFFLATOR, HYSTEROSCOPIC6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page