SOMATOM Scope 10046799Recalls & reported problems
Siemens Healthcare GmbH · Class 2
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04056869013053), received between Jul 8, 2020 and Mar 21, 2023 — classified as 1 death, 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K140912); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Use of Device Problem2
- Adverse Event Without Identified Device or Use Problem1
- Defective Component1
- Manufacturing, Packaging or Shipping Problem1
Patient problems
- Death1
- Hypersensitivity/Allergic reaction1
- Laceration(s)1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 21, 2023Defective Component; Manufacturing, Packaging or Shipping Problem; Defective Component; Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE SOMATOM SCOPE POWER CT SYSTEM. DURING SYSTEM CALIBRATION WITHOUT ANY PATIENT INVOLVEMENT, THE GANTRY ROTATION BELT BROKE APART. AS A RESULT, IT PUSHED THE PLEXI-RING OUT OF ITS HOLDER AND THE BELT CAME OUT OF THE GANTRY. THERE IS NO REPORT OF IMPACT TO THE STATE OF HEALTH OF ANY PATIENT OR USER INVOLVED, HOWEVER, IT CANNOT BE EXCLUDED THAT THIS SITUATION COULD HAVE LED TO ANY INJURIES IF A PATIENT WOULD HAVE BEEN ON THE TABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16585371 ↗InjuryApr 23, 2021Use of Device Problem; Use of Device Problem
IT WAS REPORTED TO SIEMENS THAT THE PLEXI-RING OF THE CT GANTRY COVER HAS JUMPED OUT OF ITS TRACK DUE TO A MISSING/BROKEN PLASTIC LOCK. THE FACILITY TECHNICIAN ATTEMPTED TO PUSH THE PLEXI-RING BACK INTO ITS ORIGINAL POSITION BUT DID NOT WAIT UNTIL THE GANTRY STOPPED ROTATING. THE TECHNICIAN'S FINGER GOT CAUGHT IN THE ROTATING PARTS AND A MODERATE INJURY TO THE DIGIT WAS SUSTAINED. IT WAS REPORTED THAT BOTH THE SKIN AND MUSCLE OF THE DIGIT WERE INJURED. THE INJURY WAS TREATED BY SUTURING THE LACERATION. THERE WAS NO PERMANENT DAMAGE TO THE TECHNICIAN'S FINGER. THIS INCIDENT IS CONSIDERED A WORK ACCIDENT AND NO MALFUNCTION OR GENERAL DESIGN ISSUE WERE REPORTED. THE REPORTED EVENT OCCURRED IN (B)(6).
Patient problems: Laceration(s)
MAUDE report 11715155 ↗DeathMar 5, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED. THE PATIENT HAD A STRONG ADVERSE REACTION TO THE INJECTED CONTRAST MEDIA BEFORE THE CT SCAN WAS STARTED.
Patient problems: Death; Hypersensitivity/Allergic reaction
MAUDE report 11423133 ↗MalfunctionJul 8, 2020Use of Device Problem; Use of Device Problem
IT WAS REPORTED TO SIEMENS THAT DURING THE EXAMINATION THE UNRESTRAINED PATIENT KICKED THE PLEXI RING AND BROKE IT. THE PATIENT WAS NOT INJURED DURING THE INCIDENT. IT CANNOT BE COMPLETELY EXCLUDED THAT CRITICAL INJURY DUE THE PATIENT CONTACTING THE ROTATING PARTS INSIDE THE GANTRY COULD RESULT IF THIS EVENT WERE TO REOCCUR. THIS REPORT WAS FILED WITH AN ABUNDANCE OF CAUTION. THE REPORTED EVENT OCCURRED IN (B)(6).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10245204 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K140912. One submission often covers many models, so check the product and lot details in each record.
- Z-2477-2018 · Feb 1, 2018 · TerminatedSOMATOM Scope Power (Model 10967888)
- Z-2476-2018 · Feb 1, 2018 · TerminatedSOMATOM Scope ( Model 10967666)
Other System, x-ray, tomography, computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls6 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Brilliance CT Big BorePhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed7 recalls1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.