Multix Fusion Max 11020660Recalls & reported problems
Siemens Shanghai Medical Equipment Ltd. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04056869050713), received between Feb 24, 2021 and Mar 13, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem1
- Difficult or Delayed Positioning1
- Use of Device Problem1
Patient problems
- Muscle/Tendon Damage1
- Unspecified Tissue Injury1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 13, 2024Difficult or Delayed Positioning; Activation, Positioning or Separation Problem; Difficult or Delayed Positioning; Activation, Positioning or Separation Problem
SIEMENS HEALTHINEERS WAS INFORMED OF A PRODUCT MALFUNCTION AND ADVERSE EVENT ASSOCIATED WITH THE MULTIX FUSION MAX SYSTEM. THE TECHNICIAN WAS MOVING THE X-RAY TUBE FROM THE WALL BUCKY TO THE TABLE TO PERFORM A TABLETOP X-RAY WHEN THE TUBE STAND UNEXPECTEDLY ENGAGED INTO DETENT POSITION. THE TECHNICIAN WAS AT THE HAND CONTROL WHEN THE X-RAY TUBE MOVEMENT LOCKED. THE TECHNICIAN¿S ARM WRENCHED UP AND BACK, AND THE MOMENTUM CARRIED HIM FORWARD. THE TECHNICIAN'S SHOULDER WAS INJURED DURING THIS EVENT, AND HE WENT TO THE EMERGENCY ROOM WHERE HE RECEIVED AN ULTRASOUND EXAMINATION. IT WAS DETERMINED THAT THE TECHNICIAN SUFFERED A FULL THICKNESS TEAR OF THE SUPRASPINATUS TENDON. ADDITIONAL INFORMATION WAS RECEIVED ON (B)(6) 2024, STATING THE TECHNICIAN IS STILL ABLE TO PERFORM HIS WORK DUTIES AND HAS NOT NEEDED TIME OFF WORK DUE TO HIS INJURY. THE TECHNICIAN STARTED PHYSIOTHERAPY AT THE BEGINNING OF (B)(6) 2024 (EXACT DATE UNKNOWN) AND IS UNDER THE CARE OF A SPORTS MEDICINE SPECIALIST. THERE WAS NO PATIENT INVOLVEMENT IN THIS INCIDENT.
Patient problems: Muscle/Tendon Damage
MAUDE report 18893716 ↗InjuryFeb 24, 2021Use of Device Problem; Use of Device Problem
WHILE PERFORMING SERVICE ON THE MULTIX FUSION MAX UNIT, SIEMENS LOCAL SERVICE ENGINEER NOTICED THAT ONE OF THE FOUR STEEL CORNERS OF THE DETECTOR POPPED OFF FROM THE DETECTOR. THE STEEL CORNER WAS JAMMED INSIDE THE DETECTOR'S BUCKY TRAY. THE USER WAS COVERING THE PEELING DETECTOR COVER WITH TAPE. SIEMENS WAS INFORMED OF A MINOR INJURY CAUSED BY THE PEELING DETECTOR COVER. THE OPERATOR SLIGHTLY CUT HAND ON THE PEELING EDGE; HOWEVER, NO MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: Unspecified Tissue Injury
MAUDE report 11372664 ↗
Other System, x-ray, stationary devices
- FDR Visionary Suite 566-16130-23Shimadzu Corporation · System, X-Ray, Stationary4 recalls0 reports
- FDR Visionary Suite 566-16130-31Shimadzu Corporation · System, X-Ray, Stationary2 recalls0 reports
- Proteus XRaGE Hualun Medical Systems Co., Ltd · System, x-ray, stationary2 recalls0 reports
- FDR Visionary Suite 566-16130-33Shimadzu Corporation · System, X-Ray, Stationary2 recalls0 reports
- Radiography 7300 C 712037Philips Medical Systems DMC GmbH · System, X-Ray, Stationary1 recall1 death8 reports
- DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Ltd. · System, x-ray, stationary1 recall8 reports
- DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Ltd. · System, x-ray, stationary1 recall5 reports
- DRX-Compass X-ray SystemCarestream Health, Inc. · System, x-ray, stationary1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page