SOMATOM go.Now 11061610Recalls & reported problems
Siemens Healthcare GmbH · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04056869051291), received between Oct 13, 2022 and Dec 12, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Computer Software Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Mar 7, 2022Open, Classified
Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional workflows. Results in potential patient issues: Possible rescan , Unexpected X-Ray dose and additional contrast media, Delay in diagnosis, scan abort,and patient rescan
Z-0962-2022 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Computer Software Problem1
- Electrical /Electronic Property Problem1
- No Apparent Adverse Event1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2022Computer Software Problem; Computer Software Problem
IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT AND MALFUNCTION OCCURRED WHILE OPERATING THE SOMATOM GO.NOW CT SYSTEM. ON (B)(6) 2022, THE CT SCANNER CAUSED A DELAY IN DIAGNOSIS TO AN INTUBATED 50-YEAR-OLD FEMALE PATIENT. DURING POSITIONING OF THE PATIENT ON THE TABLE FOR A CHEST EXAM, THE CT SOFTWARE HUNG UP AND THE PATIENT WAS TAKEN OFF THE TABLE TO RESTART THE CT SYSTEM. AFTER THE PATIENT WAS REMOVED FROM THE EXAM TABLE, THE PATIENT COLLAPSED AND WAS IMMEDIATELY TRANSFERRED TO THE ICU. THE EXAMINATION WAS NOT REPEATED. IT WAS REPORTED TO SIEMENS ON (B)(6) 2022, THAT THE PATIENT HAD PASSED AWAY. CURRENTLY, THERE IS NO CORRELATION BETWEEN THE SYSTEM MALFUNCTION AND THE DEATH OF THE PATIENT. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: Insufficient Information
MAUDE report 15959005 ↗MalfunctionOct 13, 2022Electrical /Electronic Property Problem; No Apparent Adverse Event; Electrical /Electronic Property Problem; No Apparent Adverse Event
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WITH THE SOMATOM GO.NOW CT SYSTEM. A SERVICE ENGINEER VISITED THE CUSTOMER TO CHECK THE CONDITION OF THE SITE AND WAS NOTIFIED THAT A FIRE HAD OCCURRED IN THE GANTRY ROOM ON (B)(6), 2022. ACCORDING TO AVAILABLE INFORMATION, THE ROOT CAUSE WAS A SHORT CIRCUIT ON THE CUSTOMER´S POWER CABLES. THE CUSTOMER REPORTED THAT THE FIRE WAS EXTINGUISHED BY THE FIRE DEPARTMENT. AS A RESULT OF THE FIRE, THE GANTRY COVERS WERE MELTED. THERE IS NO REPORT OF IMPACT TO THE STATE OF HEALTH OF ANY PATIENT OR USER INVOLVED. AT THIS TIME, THERE IS NO INDICATION THAT THE CT DEVICE MAY HAVE CONTRIBUTED TO THIS EVENT. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15594326 ↗
Other System, x-ray, tomography, computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls6 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Brilliance CT Big BorePhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed7 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page