Skip to content

SOMATOM go.Up 11061620Recalls & reported problems

Siemens Healthcare GmbH · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
1
Injuries reported
2
Recalls naming it
3

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04056869051314), received between May 21, 2021 and Dec 5, 2024 — classified as 1 death, 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

3 FDA recall records name this device's identifier, 2 still open.

6 other recalls cover devices cleared under the same FDA submission (K163296, K211373, K233650, K250822); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 22, 2022Class IIOpen, Classified

    In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped

    Z-1003-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Mar 7, 2022Class IIOpen, Classified

    Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional workflows. Results in potential patient issues: Possible rescan , Unexpected X-Ray dose and additional contrast media, Delay in diagnosis, scan abort,and patient rescan

    Z-0962-2022 · Root cause: Software design · Recall details · FDA record ↗

  3. Mar 7, 2022Class IITerminated

    Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.

    Z-0995-2022 · Root cause: Software change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1
  • Inadequate or Insufficient Training1
  • Inaudible or Unclear Audible Prompt/Feedback1
  • Radiation Output Problem1
  • Use of Device Problem1

Patient problems

  • Hypersensitivity/Allergic reaction1
  • Insufficient Information1
  • Laceration(s)1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 5, 2024Radiation Output Problem; Radiation Output Problem

    IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE SOMATOM GO. UP CT SYSTEM. THE USER REPORTED THAT A SCAN OF AN 84-YEAR-OLD FEMALE POTENTIAL STROKE PATIENT WAS NOT POSSIBLE AFTER CREATING THE TOPO. THE PREPARATION SYMBOL (GREEN CIRCLE) WAS SPINNING; BUT NEVER CHANGED. ACCORDING TO THE USER, THIS RESULTED IN A 30-MINUTE DELAY IN DIAGNOSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20852121 ↗
  • InjuryAug 10, 2023Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method

    IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED WHILE OPERATING THE SOMATOM GO.UP CT SYSTEM. DURING EXAMINATION, WHILE THE TABLE WAS BEING DRIVEN, THE PATIENT¿S FINGER WAS PINCHED BETWEEN THE TABLETOP SUPPORT AND THE TABLETOP. AS A RESULT, THE PATIENT SUFFERED A LACERATION TO THE FINGER WHICH REQUIRED FOUR STITCHES. WE ARE UNAWARE OF ANY FURTHER IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED.

    Patient problems: Laceration(s)

    MAUDE report 17497611 ↗
  • InjuryMay 26, 2021Inaudible or Unclear Audible Prompt/Feedback; Inaudible or Unclear Audible Prompt/Feedback

    IT WAS REPORTED TO SIEMENS THAT AFTER INJECTION OF A CONTRAST MEDIUM DURING A ROUTINE CT SCAN, THE PATIENT STARTED TO CHOKE ON THE PATIENT TABLE OF THE SOMATOM SYSTEM. OVER TIME, THE PATIENT'S CONDITION DETERIORATED FURTHER. THE PATIENT REQUIRED CARDIOPULMONARY RESUSCITATION TO REVIVE HIM. THE SITUATION APPEARED TO BE A SERIOUS ALLERGIC REACTION TO THE CONTRAST MEDIUM. THE PATIENT WAS STABILIZED WITHOUT ANY PERMANENT HEALTH CONSEQUENCES. THE USER ALLEGED THAT THE INTERCOM SYSTEM USED TO COMMUNICATE WITH PATIENTS DURING EXAMINATION, WAS EMITTING NOISY FEEDBACK DURING THE EXAMINATION, AND DISTURBED COMMUNICATION BETWEEN THE USER AND THE PATIENT. THE USER REPORTED THE COMMUNICATION INTERFERENCE MAY HAVE CONTRIBUTED TO THE USER MISSING THE FIRST SIGNS OF THE PATIENT'S ALLERGIC REACTION AND DISTRESS. THE REPORTED EVENT OCCURRED IN (B)(6).

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 11886039 ↗
  • DeathMay 21, 2021Inadequate or Insufficient Training; Inadequate or Insufficient Training

    IT WAS REPORTED TO SIEMENS THAT A PLANNED CT EXAMINATION OF AN EMERGENCY PATIENT FROM THE FACILITY ICU WAS NOT POSSIBLE BECAUSE THE CT COULD NOT BE STARTED DUE TO THE CALCULATED TUBE COOLING WAIT TIME. THE PATIENT HEALTH STATUS WAS CRITICAL. THE USER REPORTED THAT THE CT SCAN WAS PLANNED ONE WEEK EARLIER BUT WAS POSTPONED DUE THE PATIENT'S CONDITION AND LACK OF TRANSPORTATION CAPABILITY. A SIGNIFICANT DELAY IN DIAGNOSIS (1.5 HOURS) WAS ASSOCIATED WITH THE TUBE COOLING WAIT TIME WARNING. AFTER THE USER CONTACTED TECHNICAL SUPPORT, THE SCAN WAS ABLE TO BE COMPLETED. THE USER DID NOT ALLEGE A PRODUCT MALFUNCTION. UNFORTUNATELY, THE PATIENT PASSED AWAY IN THE ICU AT AN UNKNOWN TIME AFTER THE SCAN ON (B)(6) 2021. THE REPORTED EVENT OCCURRED IN (B)(6).

    Patient problems: Insufficient Information

    MAUDE report 11863495 ↗

FDA clearance and classification

Cleared through 510(k) K250822 on Jul 3, 2025 (decision: Substantially Equivalent). Submitted by Siemens Medical Solutions USA, Inc. as “SOMATOM go.Now; SOMATOM go.Up; SOMATOM go.All; SOMATOM go.Top; SOMATOM go.Sim; SOMATOM go.Open Pro; SOMATOM Pro.Pulse”. The FDA review took 107 days. FDA record ↗

Cleared through 510(k) K233650 on Mar 26, 2024 (decision: Substantially Equivalent). Submitted by Siemens Medical Solutions USA, Inc. as “SOMATOM go.Up; SOMATOM go.Now; SOMATOM go.All; SOMATOM go.Top; SOMATOM go.Sim; SOMATOM go.Open Pro; SOMATOM X.cite; SOMATOM X.ceed”. The FDA review took 133 days. FDA record ↗

Cleared through 510(k) K211373 on Aug 27, 2021 (decision: Substantially Equivalent). Submitted by Siemens Medical Solutions USA, Inc. as “SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM X. Platform Scanners - SOMATOM X.cite and SOMATOM X.ceed, Scan&GO Software”. The FDA review took 115 days. FDA record ↗

…and 1 more FDA submissions listed for this device.

The FDA regulates product code JAK (system, x-ray, tomography, computed) under 21 CFR 892.1750 as a Class II device. 21 CFR 892.1750 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K163296, K211373, K233650, K250822. One submission often covers many models, so check the product and lot details in each record.

  • Z-1239-2026 · Dec 19, 2025 · Open, Classified
    SOMATOM X.ceed Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS
  • Z-1253-2021 · Mar 3, 2021 · Terminated
    Software versions syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a in the Computed tomography systems as follows: SOMATOM go.Now (Model 11061610)SOMATOM go.Up (Models 11061620 & 11061628
  • Z-1939-2018 · Apr 20, 2018 · Terminated
    SOMATOM go.Now, Material Number 11061618 There is a potential for a software issue that may cause the need for necessary patient rescans.
  • Z-1940-2018 · Apr 20, 2018 · Terminated
    SOMATOM Go.Up, Material Number 11061628 There is a potential for a software issue that may cause the need for necessary patient rescans.
  • Z-0714-2018 · Oct 2, 2017 · Terminated
    SOMATOM Go.Now computed tomography (CT) x-ray system table, Model Number: 11061618 Product Usage: Intended to generate and process cross-sectional images of patients by computer r
  • Z-0713-2018 · Oct 2, 2017 · Terminated
    SOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number: 11061628 Product Usage: Intended to generate and process cross-sectional images of patients by computer re

Common questions about the SOMATOM go.Up 11061620

Has the SOMATOM go.Up 11061620 been recalled?

Yes. 3 FDA recall records name this device's identifier: 2 still open and 1 closed (terminated) by the FDA. The most recent, Z-1003-2023 (Class II), began on Dec 22, 2022: In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the SOMATOM go.Up 11061620?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between May 21, 2021 and Dec 5, 2024. In those reports, the most common specific device problems are “Improper or Incorrect Procedure or Method” (1), “Inadequate or Insufficient Training” (1) and “Inaudible or Unclear Audible Prompt/Feedback” (1); the most common patient problems are “Hypersensitivity/Allergic reaction” (1) and “Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SOMATOM go.Up 11061620?

1 report is classified as a death, 2 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SOMATOM go.Up 11061620 FDA approved?

It was cleared through 510(k) K163296 on Mar 21, 2017; its record lists 4 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the SOMATOM go.Up 11061620, and is it still sold?

The FDA's device database lists Siemens Healthcare GmbH as the company responsible for it (brand SOMATOM go.Up, model 11061620). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page