SOMATOM go.Sim 11061660Recalls & reported problems
Siemens Healthcare GmbH · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04056869249230), received between Nov 18, 2022 and Nov 18, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Scratched Material” (1 report).
2 FDA recall records name this device's identifier, 1 still open.
4 other recalls cover devices cleared under the same FDA submission (K192061, K211373, K233650, K250822); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Mar 7, 2022Class IITerminated
Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.
Z-0996-2022 · Root cause: Software change control · Recall details · FDA record ↗
- Mar 7, 2022Class IIOpen, Classified
Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional workflows. Results in potential patient issues: Possible rescan , Unexpected X-Ray dose and additional contrast media, Delay in diagnosis, scan abort,and patient rescan
Z-0962-2022 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Scratched Material1
Patient problems
- Abrasion1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 18, 2022Scratched Material; Scratched Material
IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED WHILE OPERATING THE SOMATOM GO.SIM CT SYSTEM. IT WAS REPORTED THAT A CT TECHNOLOGIST CUT HER RIGHT INDEX FINGER ON THE UNDERSIDE OF THE RADIATION TABLETOP DUE TO A SHARP EDGE WHILE REMOVING THE HEADREST. THE TECHNOLOGISTS' FINGER WAS CLEANED, AND BANDAGED, AND NO FURTHER MEDICAL TREATMENT WAS REQUIRED. THE SERVICE ENGINEER CHECKED THE INVOLVED HARDWARE, AND NO VISIBLE IRREGULARITIES WERE OBSERVED ON THE TABLE, TABLETOP, OR HEADREST. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT. FOLLOWING THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED IF A DEVICE MALFUNCTION IS IDENTIFIED.
Patient problems: Abrasion
MAUDE report 15825368 ↗
FDA clearance and classification
Cleared through 510(k) K250822 on Jul 3, 2025 (decision: Substantially Equivalent). Submitted by Siemens Medical Solutions USA, Inc. as “SOMATOM go.Now; SOMATOM go.Up; SOMATOM go.All; SOMATOM go.Top; SOMATOM go.Sim; SOMATOM go.Open Pro; SOMATOM Pro.Pulse”. The FDA review took 107 days. FDA record ↗
Cleared through 510(k) K233650 on Mar 26, 2024 (decision: Substantially Equivalent). Submitted by Siemens Medical Solutions USA, Inc. as “SOMATOM go.Up; SOMATOM go.Now; SOMATOM go.All; SOMATOM go.Top; SOMATOM go.Sim; SOMATOM go.Open Pro; SOMATOM X.cite; SOMATOM X.ceed”. The FDA review took 133 days. FDA record ↗
Cleared through 510(k) K211373 on Aug 27, 2021 (decision: Substantially Equivalent). Submitted by Siemens Medical Solutions USA, Inc. as “SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM X. Platform Scanners - SOMATOM X.cite and SOMATOM X.ceed, Scan&GO Software”. The FDA review took 115 days. FDA record ↗
…and 1 more FDA submissions listed for this device.
The FDA regulates product code JAK (system, x-ray, tomography, computed) under 21 CFR 892.1750 as a Class II device. 21 CFR 892.1750 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K192061, K211373, K233650, K250822. One submission often covers many models, so check the product and lot details in each record.
- Z-1239-2026 · Dec 19, 2025 · Open, ClassifiedSOMATOM X.ceed Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS
- Z-1003-2023 · Dec 22, 2022 · Open, ClassifiedSoftware : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D license. Used on the following CT (Computed tomography systems): SOMATOM go.
- Z-0995-2022 · Mar 7, 2022 · Terminatedsyngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Up Model Number 11061620. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo
- Z-1253-2021 · Mar 3, 2021 · TerminatedSoftware versions syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a in the Computed tomography systems as follows: SOMATOM go.Now (Model 11061610)SOMATOM go.Up (Models 11061620 & 11061628
Common questions about the SOMATOM go.Sim 11061660
Has the SOMATOM go.Sim 11061660 been recalled?
Yes. 2 FDA recall records name this device's identifier: 1 still open and 1 closed (terminated) by the FDA. The most recent, Z-0996-2022 (Class II), began on Mar 7, 2022: Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the SOMATOM go.Sim 11061660?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 18, 2022. That report lists device problem “Scratched Material” (1); patient problem “Abrasion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SOMATOM go.Sim 11061660?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SOMATOM go.Sim 11061660 FDA approved?
It was cleared through 510(k) K192061 on Nov 21, 2019; its record lists 4 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the SOMATOM go.Sim 11061660, and is it still sold?
The FDA's device database lists Siemens Healthcare GmbH as the company responsible for it (brand SOMATOM go.Sim, model 11061660). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System, x-ray, tomography, computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths49 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls23 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls5 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Brilliance CT Big BorePhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed7 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page