Skip to content

MULTIX Impact C 11020896Recalls & reported problems

Siemens Shanghai Medical Equipment Ltd. · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04056869262147), received between May 18, 2022 and Mar 18, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Installation-Related Problem1
  • Key or Button Unresponsive/not Working1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Unspecified Tissue Injury1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2026Key or Button Unresponsive/not Working

    SIEMENS HEALTHINEERS WAS NOTIFIED OF A PROBLEM INVOLVING THE MULTIX IMPACT C SYSTEM. DURING THE PRE-USE PREPARATION, THE USER FOUND THE E-STOP AT TABLETOP WAS BROKEN AND LOST ITS FUNCTION. THERE WAS NO INJURY REPORTED. ALTHOUGH THERE ARE OTHER EMERGENCY STOPS THAT CAN BE USED IF NEEDED, IN A WORST-CASE SCENARIO, CRITICAL INJURY COULD OCCUR DUE TO COLLISION IF THE USER IS UNABLE TO STOP THE TABLE USING THE OTHER EMERGENCY STOPS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24629187 ↗
  • MalfunctionOct 31, 2022Installation-Related Problem; Installation-Related Problem

    SIEMENS BECAME AWARE ON AN INCIDENT WITH THE MULTIX IMPACT C UNIT. THE USER'S FINGER WAS CUT BY A SHARP EDGE OR CORNER WHEN PULLING OUT THE TRAY OF THE BUCKY WALL STAND (BWS). NO MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 15698338 ↗
  • InjuryMay 18, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    SIEMENS BECAME AWARE OF AN INCIDENT THAT OCCURRED ON THE MULTIX IMPACT C DEVICE. WHEN THE X-RAY TECHNICIAN WAS PREPARING FOR A SCOLI SERIES EXAMINATION AT THE WALL STAND, HE NEITHER REMOVED THE STRETCH GRIP NOR CHECKED WHETHER THE OVERHEAD HAND GRIP WAS FIRMLY ATTACHED TO THE WALL STAND. AS THE EXAMINATION BEGAN, THE STRETCH GRIP CAUGHT THE TOP OF THE MULTI PURPOSE ORTHO STAND CAUSING THE HANDLE TO BECOME LOOSE AND FALL FORWARD. THE PART FELL BUT MISSED THE PATIENT. THERE ARE NO INJURIES ATTRIBUTED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14427378 ↗

Other System, x-ray, stationary devices

All System, x-ray, stationary devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.