NAEOTOM Alpha 11330003Recalls & reported problems
Siemens Healthcare GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04056869263168), received between Jul 28, 2022 and Jul 23, 2024 — classified as 1 death, 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
2 other recalls cover devices cleared under the same FDA submission (K211591, K220814, K233657, K243523, K251061); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 19, 2025Class IIOpen, Classified
To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
Z-1240-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 22, 2022Class IIOpen, Classified
In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped
Z-1003-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Electrical /Electronic Property Problem2
- Improper or Incorrect Procedure or Method2
- Unexpected Shutdown2
- Adverse Event Without Identified Device or Use Problem1
- Delayed Program or Algorithm Execution1
- Image Display Error/Artifact1
- Output Problem1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Abrasion1
- Cardiac Arrest1
- Crushing Injury1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 23, 2024Image Display Error/Artifact; Output Problem; Image Display Error/Artifact; Output Problem
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE NAEOTOM ALPHA CT SCANNER IN THE UNITED STATES ON (B)(6) 2024. AFTER A SUCCESSFUL CT HEAD SCAN OF A STROKE PATIENT, THE READING RADIOLOGIST IDENTIFIED A LESION INFERIOR TO THE HEMORRHAGE. A SHORT TIME LATER, THE PATIENT¿S CONDITION DECLINED, AND A SUBSEQUENT CT WAS OBTAINED ON A DIFFERENT SCANNER. COMPARISON REVEALED THAT THE ORIGINAL OBSERVATION FROM THE NAEOTOM ALPHA SYSTEM WAS AN ARTIFACT AND NOT A LESION. AS PER OUR UNDERSTANDING, THERE WAS A DIAGNOSIS SCAN FOR STROKE TREATMENT DURING WHICH AN ADDITIONAL LESION WAS SEEN IN RECONSTRUCTED VNC (VIRTUAL NON CONTRAST) IMAGES, PRESUMABLY AS A CHANCE FINDING. THE SECOND SCAN BECAME NECESSARY DUE TO THE PATIENTS¿ HEALTH SITUATION. WHEN COMPARING THE IMAGES, THE CUSTOMER IDENTIFIED THE PREVIOUSLY SEEN LESION AS AN ARTIFACT. BASED ON OUR INVESTIGATION TO DATE, IT IS OUR UNDERSTANDING THAT THERE IS NO CONNECTION BETWEEN THE MENTIONED LESION AND THE REPORTED STROKE, BUT THE EXACT RELEVANCE FOR THIS CLINICAL CASE IS STILL UNCLEAR. THE PATIENT¿S CURRENT HEALTH STATUS IS UNKNOWN AT THIS TIME. SIEMENS IS CONDUCTING A THOROUGH INVESTIGATION OF THE REPORTED EVENTS. AS WE C
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19806732 ↗InjuryAug 10, 2023Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method
IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED WHILE OPERATING THE NAEOTOM ALPHA CT SYSTEM. A 60-69-YEAR-OLD FEMALE PATIENT WAS PLANNED FOR TWO SCANS (HEAD AND THORAX/ABDOMEN). THE PATIENT WAS POSITIONED WITH FEET FIRST AND WAS INSTRUCTED BY THE OPERATOR TO HAVE BOTH ARMS CROSSED OVER THE BODY. AFTER THE HEAD SCAN, WITHOUT THE KNOWLEDGE OF THE OPERATOR, THE PATIENT LOWERED HER ARMS TO THE SIDES OF THE BODY. DURING POSITIONING OF THE SECOND SCAN, THE PATIENT¿S FINGER WAS PINCHED BETWEEN THE TABLETOP AND ITS COVER. AS A RESULT, THE PATIENT SUFFERED AN ABRASION TO THE FINGER. THE PATIENT WAS TREATED WITH MINOR FIRST AID AND NO FURTHER MEDICAL INTERVENTION WAS NECESSARY.
Patient problems: Abrasion
MAUDE report 17497614 ↗MalfunctionJun 5, 2023Delayed Program or Algorithm Execution; Delayed Program or Algorithm Execution
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE NAEOTOM ALPHA CT SYSTEM. ON (B)(6) 2023, A 12-YEAR-OLD FEMALE PATIENT WAS UNDERGOING A CT EXAMINATION. FOLLOWING THE TOPOGRAM, THE SPIRAL SCAN WAS STARTED AND THEN ABORTED. REMOTE TROUBLESHOOTING WITH SIEMENS CUSTOMER SERVICE CORRECTED THE ISSUE TEMPORARILY AND THE SCAN WAS REPEATED SUCCESSFULLY. NO FURTHER NEGATIVE HEALTH CONSEQUENCES WERE REPORTED. SIEMENS HEALTHINEERS CONSIDERS CHILDREN A VULNERABLE PATIENT GROUP, THEREFORE, THIS REPORT IS FILED WITH AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17060173 ↗MalfunctionMay 17, 2023Electrical /Electronic Property Problem; Unexpected Shutdown; Electrical /Electronic Property Problem; Unexpected Shutdown
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE NAEOTOM ALPHA CT SYSTEM. ON (B)(6) 2023, A 6-YEAR-OLD MALE PATIENT UNDERWENT AN INNER EAR EXAMINATION IN WHICH THE SYSTEM HUNG UP DURING THE SCAN. AS A RESULT, A RESCAN OF THE PATIENT WAS NECESSARY. NO NEGATIVE HEALTH CONSEQUENCES WERE REPORTED FOR THE PATIENT ASSOCIATED WITH THE REPORTED EVENT. THEREFORE, THIS REPORT IS SUBMITTED WITH AN ABUNDANCE OF CAUTION AS SIEMENS CONSIDERS CHILDREN AS A VULNERABLE PATIENT GROUP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16950153 ↗MalfunctionMar 20, 2023Electrical /Electronic Property Problem; Unexpected Shutdown; Electrical /Electronic Property Problem; Unexpected Shutdown
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE NEOTOMA ALPHA CT SYSTEM. AFTER A TOPO SCAN OF A 3 YEAR OLD MALE PATIENT, WHEN LOADING THE NEXT SCAN, THE LOADING PROCESS WAS ABORTED WITH AN ERROR. THE USER RESTARTED THE CT SYSTEM AND REPEATED THE TOPOGRAM SCAN WITHOUT SUCCESS. THE PATIENT WAS THEN EXAMINED ON ANOTHER CT SYSTEM. THE PATIENT WAS SEDATED DURING THIS PROCEDURE TO PERFORM THE EXAMINATION. NO NEGATIVE HEALTH CONSEQUENCES WERE REPORTED FOR THE CHILD ASSOCIATED WITH THE REPORTED EVENT. THEREFORE, THIS REPORT HAS BEEN SUBMITTED WITH AN ABUNDANCE OF CAUTION AS SIEMENS CONSIDERS CHILDREN AS A VULNERABLE PATIENT GROUP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16573768 ↗DeathAug 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED WHILE OPERATING THE NAETOM ALPHA CT SYSTEM. SIEMENS WAS NOTIFIED THAT A PATIENT WITH A SUSPECTED PULMONARY EMBOLISM COULD NOT BE SCANNED ON (B)(6) 2022. THE PATIENT PASSED AWAY, ALTHOUGH, THE EXACT TIME OF DEATH IS UNKNOWN AT THE TIME. DURING INITIAL ANALYSIS, THE SIEMENS SERVICE SPECIALIST FOUND THAT NO VALID CUSTOMER SCAN PROTOCOL FOR PULMONARY EMBOLISM WITH CONTRAST AGENT WAS AVAILABLE. A TECHNICAL INVESTIGATION HAS BEEN STARTED TO DETERMINE WHETHER THE CT SCANNER OR THE CONTRAST AGENT INJECTOR (3RD PARTY DEVICE) MAY HAVE CONTRIBUTED TO THE EVENT. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
Patient problems: Cardiac Arrest
MAUDE report 15309899 ↗InjuryJul 28, 2022Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED WHILE OPERATING THE NAEOTOM ALPHA CT SYSTEM. AT THE END OF AN EXAMINATION, DURING HORIZONTAL MOVEMENT OF THE PATIENT TABLE, THE PATIENT´S (90 YEARS, FEMALE) LEFT INDEX FINGER WAS TRAPPED IN THE GAP BETWEEN THE TABLETOP AND TABLE SUPPORT. THE PATIENT WAS SENT TO THE EMERGENCY ROOM OF A HOSPITAL THAT SPECIALIZES IN HAND TREATMENT. NO FURTHER INFORMATION WAS PROVIDED REGARDING THE EXTENT OF THE INJURY NOR HOW HER FINGER WAS MEDICALLY TREATED. ACCORDING TO THE AVAILABLE INFORMATION THE PATIENT RETURNED HOME THE SAME DAY. THIS REPORT HAS BEEN SUBMITTED WITH AN ABUNDANCE OF CAUTION.
Patient problems: Crushing Injury
MAUDE report 15120744 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K211591, K220814, K233657, K243523, K251061. One submission often covers many models, so check the product and lot details in each record.
- Z-1242-2026 · Dec 19, 2025 · Open, ClassifiedNAEOTOM Alpha.Pro Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS
- Z-1241-2026 · Dec 19, 2025 · Open, ClassifiedNAEOTOM Alpha.Prime Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS
Other System, x-ray, tomography, computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls6 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Brilliance CT Big BorePhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed7 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page