VOLUMAT MC AGILIA US Z021135Recalls & reported problems
Fresenius Kabi AG · The VOLUMAT MC AGILIA US is a transportable equipment intended to be used in healthcare facilities environment by healthcare trained professionals, according to hospital protocols on adults, paediatrics and neonates human patients to administer - intermittent or continuous delivery of parenteral fluids (solutions, colloids, parenteral nutrition) and medications (including but not limited to diluted drugs, chemotherapy) through clinically accepted IV routes of administration. - transfusion of blood and blood derivatives products. Note that for transfusion procedures, use of a special administration set is required. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Fresenius Volumat MC Agilia infusion pump has 106 reports, including 2 deaths and 1 injury. Many reports describe the pump showing error codes, failing during pre-use checks, or having charging problems, and a few describe air in the device or the pump delivering too much or too little fluid. Most reports note no clinical signs in patients. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 102
- Deaths reported
- 2
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 102 adverse event reports that list this exact device (DI 04086000851794), received between Jun 21, 2024 and Sep 28, 2026 — classified as 2 deaths, 1 injury, 99 malfunctions.
The most frequently reported specific device problem is “Charging Problem” (8 reports).
2 other recalls cover devices cleared under the same FDA submission (K121613); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Apparent Adverse Event99
- Charging Problem8
- Excess Flow or Over-Infusion8
- Infusion or Flow Problem6
- Air/Gas in Device4
- Calibration Problem2
- Device Alarm System2
- Display or Visual Feedback Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions97
- Reaction to Medicinal Component of Device3
- Localized Skin Lesion2
Recent adverse event reports
The 25 most recent of 102 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026No Apparent Adverse Event
The following has been reported:Unit does not power oncustomer sent in the attached request form reporting unit not power on.Reporting as a conservative measure. No Adverse Event effects were reported.The device was repaired by the customer; Fresenius Kabi is communicating the investigation and repair results here. Fresenius Kabi will not be receiving the device or the parts back for investigationAs this device was manufactured greater than 3 years ago (Manufacturing date: May 2015), preventative maintenance (PM) is expected to have been performed as per the device's approved maintenance schedule. Therefore, the device would have undergone two quality verifications of each essential performance parameter (at release and first PM), indicating the likelihood of a manufacturing-related issue is remote. Review of the Device History Record (DHR) has been deemed not required in accordance with internal procedure.Device history log was reviewed. Several Technical errors (37 and 38) related to the reported event were found.The reported device was received in Lake Zurich and its investigation was performed by our Local Technical Team.According to provided information, external inspection co
Patient problems: Localized Skin Lesion
MAUDE report 50919129 ↗MalfunctionSep 28, 2026No Apparent Adverse Event
Patient problems: Localized Skin Lesion
MAUDE report 50919128 ↗MalfunctionAug 26, 2026No Apparent Adverse Event
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50460992 ↗MalfunctionAug 26, 2026No Apparent Adverse Event
The following has been reported:Pump failed mid infusion with the error code "error 01(0001)".The customer ended up switching out the old pump with a new pump for their patient and are returned the defective pump.Reporting as a conservative measure. No Adverse Event effects were reported.Additional information is needed to complete the investigation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50460988 ↗MalfunctionAug 3, 2026No Apparent Adverse Event
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50117623 ↗MalfunctionAug 3, 2026No Apparent Adverse Event
The following has been reported by Customer:The Air Sensor failed during pre testing / inspection.Reporting as a conservative measure. Adverse effects were not reported.Device History Record was not reviewed as the device has already gone through its first preventive maintenance.The Device Log was reviewed, no error or issue linked to the reported event was found.The device was received in Lake Zurich for Preventive Maintenance during which our Local Technical Team found the failure of the air sensor.According to provided information, nothing was noticed during both external and internal check.The test 14 (Air sensor) failed.The air sensor has been replaced to solved the issues. The problem of defective air sensors is a known topic. An investigation is underway through an internal CAPA in order to find all possible root causes and correct them. When the sample will be received, it will be made available to the technical team to increase the test samples if necessary. Therefore this complaint is considered as VALID and the root cause is linked to a defective air sensor.To our knowledge this complaint is not being related to patient safetyThis complaint is considered as: VALID.The
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50117610 ↗MalfunctionJun 30, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: CUSTOMER REPORTS THAT THIS IS A NEWLY UNBOXED UNIT THAT THEY TESTED AND IT THROWS THEM ERROR CODE 49 WHILE PERFORMING THE PM. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. DEVICE HISTORY RECORD WAS NOT REVIEWED AS THE DEVICE HAS ALREADY GONE THROUGH ITS FIRST PREVENTIVE MAINTENANCE. DEVICE LOG HISTORY WAS NOT PROVIDED THEREFORE NO REVIEW COULD BE PERFORMED. THE REPORTED DEVICE WAS NOT RETURNED TO LAKE ZURICH NOR BRÉZINS FOR INVESTIGATION. ACCORDING TO LOCAL PROTOCOL, IF AFTER 20 DAYS THE DEVICE STILL HASN'T BEEN RETURNED, THE SERVICE ORDER IS CLOSED TO NO DEVICE. SO, WE DID NOT RECEIVE ANYTHING THAT WOULD ALLOW US TO IDENTIFY THE ACTUAL ROOT CAUSE OF THIS EVENT / TO GO FURTHER WITH THIS INVESTIGATION. THEREFORE THE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE IDENTIFIED. HOWEVER, IF FURTHER INFORMATION ARE PROVIDED LATER ON WE MAY RE-OPEN THE COMPLAINT AND PURSUE ITS. TO OUR KNOWLEDGE THIS COMPLAINT IS NOT BEING RELATED TO PATIENT SAFETY. THIS COMPLAINT IS CONSIDERED AS: NOT INVESTIGATED. THE TREND IS: NORMAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25667983 ↗MalfunctionJun 30, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: CUSTOMER REPORTS THAT THIS IS A NEWLY UNBOXED UNIT THAT THEY TESTED AND IT THROWS THEM ERROR CODE 49 WHILE PERFORMING THE PM. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION. DEVICE HISTORY RECORD WAS NOT REVIEWED AS THE DEVICE HAS ALREADY GONE THROUGH ITS FIRST PREVENTIVE MAINTENANCE. DEVICE LOG HISTORY WAS NOT PROVIDED THEREFORE NO REVIEW COULD BE PERFORMED. THE REPORTED DEVICE WAS NOT RETURNED TO LAKE ZURICH NOR BRÉZINS FOR INVESTIGATION. ACCORDING TO LOCAL PROTOCOL, IF AFTER 20 DAYS THE DEVICE STILL HASN'T BEEN RETURNED, THE SERVICE ORDER IS CLOSED TO NO DEVICE. SO, WE DID NOT RECEIVE ANYTHING THAT WOULD ALLOW US TO IDENTIFY THE ACTUAL ROOT CAUSE OF THIS EVENT / TO GO FURTHER WITH THIS INVESTIGATION. THEREFORE THE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE IDENTIFIED. HOWEVER, IF FURTHER INFORMATION ARE PROVIDED LATER ON WE MAY RE-OPEN THE COMPLAINT AND PURSUE ITS. TO OUR KNOWLEDGE THIS COMPLAINT IS NOT BEING RELATED TO PATIENT SAFETY THIS COMPLAINT IS CONSIDERED AS: NOT INVESTIGATED THE TREND IS: NORMAL
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25667981 ↗MalfunctionJun 30, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: CUSTOMER REPORTS THAT THIS IS A NEWLY UNBOXED UNIT THAT THEY TESTED AND IT THROWS THEM ERROR CODE 49 WHILE PERFORMING THE PM. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. DEVICE HISTORY RECORD WAS NOT REVIEWED AS THE DEVICE HAS ALREADY GONE THROUGH ITS FIRST PREVENTIVE MAINTENANCE. DEVICE LOG HISTORY WAS NOT PROVIDED THEREFORE NO REVIEW COULD BE PERFORMED. THE REPORTED DEVICE WAS NOT RETURNED TO LAKE ZURICH NOR BRÉZINS FOR INVESTIGATION. ACCORDING TO LOCAL PROTOCOL, IF AFTER 20 DAYS THE DEVICE STILL HASN'T BEEN RETURNED, THE SERVICE ORDER IS CLOSED TO NO DEVICE. SO, WE DID NOT RECEIVE ANYTHING THAT WOULD ALLOW US TO IDENTIFY THE ACTUAL ROOT CAUSE OF THIS EVENT / TO GO FURTHER WITH THIS INVESTIGATION. THEREFORE THE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE IDENTIFIED. HOWEVER, IF FURTHER INFORMATION ARE PROVIDED LATER ON WE MAY RE-OPEN THE COMPLAINT AND PURSUE ITS. TO OUR KNOWLEDGE THIS COMPLAINT IS NOT BEING RELATED TO PATIENT SAFETY. THIS COMPLAINT IS CONSIDERED AS: NOT INVESTIGATED. THE TREND IS: NORMAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25667978 ↗MalfunctionJun 30, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: CUSTOMER REPORTS THAT THIS IS A NEWLY UNBOXED UNIT THAT THEY TESTED AND IT THROWS THEM ERROR CODE 49 WHILE PERFORMING THE PM. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. DEVICE HISTORY RECORD WAS NOT REVIEWED AS THE DEVICE HAS ALREADY GONE THROUGH ITS FIRST PREVENTIVE MAINTENANCE. DEVICE LOG HISTORY WAS NOT PROVIDED THEREFORE NO REVIEW COULD BE PERFORMED. THE REPORTED DEVICE WAS NOT RETURNED TO LAKE ZURICH NOR BRÉZINS FOR INVESTIGATION. ACCORDING TO LOCAL PROTOCOL, IF AFTER 20 DAYS THE DEVICE STILL HASN'T BEEN RETURNED, THE SERVICE ORDER IS CLOSED TO NO DEVICE. SO, WE DID NOT RECEIVE ANYTHING THAT WOULD ALLOW US TO IDENTIFY THE ACTUAL ROOT CAUSE OF THIS EVENT / TO GO FURTHER WITH THIS INVESTIGATION. THEREFORE, THE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE IDENTIFIED. HOWEVER, IF FURTHER INFORMATION ARE PROVIDED LATER ON, WE MAY RE-OPEN THE COMPLAINT AND PURSUE ITS. TO OUR KNOWLEDGE, THIS COMPLAINT IS NOT BEING RELATED TO PATIENT SAFETY. THIS COMPLAINT IS CONSIDERED AS: NOT INVESTIGATED. THE TREND IS: NORMAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25667977 ↗MalfunctionJun 5, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: ERROR 37 - CAN'T OPERATE PUMP CUSTOMER SENT IN THE REQUEST FORM REPORTING: ERROR 37 - CAN'T OPERATE PUMP. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. DEVICE HISTORY RECORD WAS NOT REVIEWED AS THE DEVICE HAS ALREADY GONE THROUGH ITS FIRST PREVENTIVE MAINTENANCE. DEVICE HISTORY LOG WAS REVIEWED. SEVERAL TECHNICAL ERRORS 37 RELATED TO THE REPORTED EVENT WERE FOUND. THE REPORTED DEVICE WAS RECEIVED IN LAKE ZURICH AND ITS INVESTIGATION WAS PERFORMED BY OUR LOCAL TECHNICAL TEAM. ACCORDING TO PROVIDED INFORMATION, THE REPORTED EVENT WAS NOT REPRODUCED BUT EXTERNAL INSPECTION FOUND DELAMINATED START/ OK BUTTON THAT COULD FIRMED THE REPORTED EVENT. DURING TESTS, OCS TEST FAILED (MISSING RIBBON CABLE Z179413). INTERNAL INSPECTION FOUND THAT RIBBON CABLE FROM CPU BOARD TO POWER SUPPLY BOARD IS MISSING LIKELY DUE TO A MISHANDLING OF THE DEVICE. THIS COMPLAINT REGARDING THE ERROR 37 IS CONSIDERED AS VALID, LINKED TO AN AGEING OF THE KEYPAD. TO OUR KNOWLEDGE THIS COMPLAINT IS NOT BEING RELATED TO PATIENT SAFETY. THIS COMPLAINT IS CONSIDERED AS: VALID. THE TREND IS: NORMAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25408798 ↗MalfunctionJun 5, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: ERROR 37 - CAN'T OPERATE PUMP CUSTOMER SENT IN THE REQUEST FORM REPORTING: ERROR 37 - CAN'T OPERATE PUMP. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. DEVICE HISTORY RECORD WAS NOT REVIEWED AS THE DEVICE HAS ALREADY GONE THROUGH ITS FIRST PREVENTIVE MAINTENANCE. DEVICE HISTORY LOG WAS REVIEWED. SEVERAL TECHNICAL ERRORS 37 RELATED TO THE REPORTED EVENT WERE FOUND. THE REPORTED DEVICE WAS RECEIVED IN LAKE ZURICH AND ITS INVESTIGATION WAS PERFORMED BY OUR LOCAL TECHNICAL TEAM. ACCORDING TO PROVIDED INFORMATION, THE REPORTED EVENT WAS NOT REPRODUCED BUT EXTERNAL INSPECTION FOUND DELAMINATED START/ OK BUTTON THAT COULD FIRMED THE REPORTED EVENT. DURING TESTS, OCS TEST FAILED (MISSING RIBBON CABLE Z179413). INTERNAL INSPECTION FOUND THAT RIBBON CABLE FROM CPU BOARD TO POWER SUPPLY BOARD IS MISSING LIKELY DUE TO A MISHANDLING OF THE DEVICE. THIS COMPLAINT REGARDING THE ERROR 37 IS CONSIDERED AS VALID, LINKED TO AN AGEING OF THE KEYPAD. TO OUR KNOWLEDGE THIS COMPLAINT IS NOT BEING RELATED TO PATIENT SAFETY THIS COMPLAINT IS CONSIDERED AS: VALID THE TREND IS: NORMAL
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25408794 ↗MalfunctionJun 3, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: DAMAGE TO FRONT LOCKING MECHANISM CUSTOMER SENT IN THE REQUEST FORM REPORTING: DAMAGE TO FRONT LOCKING MECHANISM FOR TUBING. ACTION TAKEN: RECEIVED PUMP (B)(6). CONFIRMED CUSTOMER REPORTED ISSUE. THE DOOR LOCKING MECHANISM WAS FOUND TO BE DAMAGED, AND THE DOOR ASSEMBLY WAS REPLACED. ADDITIONALLY, THE LCD DISPLAY BACKLIGHT WAS INOPERATIVE; AS A RESULT, THE DISPLAY WAS REPLACED. THE AIR SENSOR FAILED CALIBRATION AND WAS ALSO REPLACED. A FULL CALIBRATION WAS PERFORMED ON THE DEVICE. EQUIPMENT CLEANED, POST SERVICE TESTED AND RELEASED FOR USE. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION. DEVICE HISTORY RECORD WAS NOT REVIEWED AS THE DEVICE HAS ALREADY GONE THROUGH ITS FIRST PREVENTIVE MAINTENANCE. DEVICE LOG WAS REVIEWED, NO ERROR OR ISSUE LINKED TO THE REPORTED EVENT WAS FOUND. THE REPORTED DEVICE WAS RECEIVED IN LAKE ZURICH AND ITS INVESTIGATION WAS PERFORMED BY OUR LOCAL TECHNICAL TEAM. DURING THIS INVESTIGATION, THE SCREEN WAS FOUND VERY DIM, BACK LIGHT WAS NOT FUNCTIONING. THIS COMPLAINT IS CONSIDERED AS VALID, DIM SCREEN IS DUE TO AGEING, DEVICE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25375229 ↗MalfunctionJun 3, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: DAMAGE TO FRONT LOCKING MECHANISM CUSTOMER SENT IN THE REQUEST FORM REPORTING: DAMAGE TO FRONT LOCKING MECHANISM FOR TUBING. ACTION TAKEN: RECEIVED PUMP (B)(6). CONFIRMED CUSTOMER REPORTED ISSUE. THE DOOR LOCKING MECHANISM WAS FOUND TO BE DAMAGED, AND THE DOOR ASSEMBLY WAS REPLACED. ADDITIONALLY, THE LCD DISPLAY BACKLIGHT WAS INOPERATIVE; AS A RESULT, THE DISPLAY WAS REPLACED. THE AIR SENSOR FAILED CALIBRATION AND WAS ALSO REPLACED. A FULL CALIBRATION WAS PERFORMED ON THE DEVICE. EQUIPMENT CLEANED, POST SERVICE TESTED AND RELEASED FOR USE. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION. DEVICE HISTORY RECORD WAS NOT REVIEWED AS THE DEVICE HAS ALREADY GONE THROUGH ITS FIRST PREVENTIVE MAINTENANCE. DEVICE LOG WAS REVIEWED, NO ERROR OR ISSUE LINKED TO THE REPORTED EVENT WAS FOUND. THE REPORTED DEVICE WAS RECEIVED IN LAKE ZURICH AND ITS INVESTIGATION WAS PERFORMED BY OUR LOCAL TECHNICAL TEAM. DURING THIS INVESTIGATION, THE SCREEN WAS FOUND VERY DIM, BACK LIGHT WAS NOT FUNCTIONING. THIS COMPLAINT IS CONSIDERED AS VALID, DIM SCREEN IS DUE TO AGEING, DEVICE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25375227 ↗MalfunctionApr 28, 2026No Apparent Adverse Event; Air/Gas in Device
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: CUSTOMER SENT IN THE REQUEST FORMS REPORTING: AIR IN LINE ERROR. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25019761 ↗MalfunctionApr 28, 2026No Apparent Adverse Event; Air/Gas in Device
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: CUSTOMER SENT IN THE REQUEST FORMS REPORTING: AIR IN LINE ERROR. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25019737 ↗MalfunctionApr 28, 2026No Apparent Adverse Event; Air/Gas in Device
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: CUSTOMER SENT IN THE REQUEST FORMS REPORTING: AIR IN LINE ERROR. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25019713 ↗MalfunctionApr 28, 2026No Apparent Adverse Event; Air/Gas in Device
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: CUSTOMER SENT IN THE REQUEST FORMS REPORTING: AIR IN LINE ERROR. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25019646 ↗MalfunctionApr 13, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: AIR IN LINE ERROR: CUSTOMER REPORTING AIR IN LINE ERROR-REPORTED TRYING TO CLEAR SEVERAL TIMES AND ERROR WOULD NOT CLEAR. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. DEVICE HISTORY RECORD WAS REVIEWED AND NOTHING WAS FOUND RELATED TO THE REPORTED EVENT. THE DEVICE LOG WAS REVIEWED, NO ERROR OR ISSUE LINKED TO THE REPORTED EVENT WAS FOUND. THE REPORTED DEVICE WAS RECEIVED IN LAKE ZURICH AND ITS INVESTIGATION WAS PERFORMED BY OUR LOCAL TECHNICAL TEAM. ACCORDING TO PROVIDED INFORMATION, NOTHING WAS NOTICED DURING BOTH EXTERNAL AND INTERNAL CHECK. DURING TESTS, DEVICE REPORTS AIR BUBBLE WHEN TUBE WITH WATER IS INSTALLED. CALIBRATION OF THE AIR SENSOR RESOLVED THE ISSUE. THIS COMPLAINT WILL BE CONSIDERED AS A VALID, THE ROOT CAUSE IS LIKELY DUE TO DRIFT OF VALUE, SO AIR CALIBRATION NEEDED. PLEASE BE INFORMED THAT AN INTERNAL CAPA WAS OPENED FOR DEFECTIVE AIR SENSORS, IN PARTICULAR A DRIFT OF THE SENSORS. TO OUR KNOWLEDGE THIS COMPLAINT IS NOT BEING RELATED TO PATIENT SAFETY. THIS COMPLAINT IS CONSIDERED AS: VALID. THE TREND IS: NORMAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24865173 ↗MalfunctionApr 13, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: INTERNAL COMPONENT LOOSE AND RATTLING CUSTOMER SENT IN THE FORM REPORTING THE INTERNAL COMPONENT LOOSE AND RATTLING (LOOSE SCREW). REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24865154 ↗MalfunctionApr 13, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: AIR IN LINE ERROR CUSTOMER REPORTING AIR IN LINE ERROR-REPORTED TRYING TO CLEAR SEVERAL TIMES AND ERROR WOULD NOT CLEAR. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. DEVICE HISTORY RECORD WAS REVIEWED AND NOTHING WAS FOUND RELATED TO THE REPORTED EVENT. THE DEVICE LOG WAS REVIEWED, NO ERROR OR ISSUE LINKED TO THE REPORTED EVENT WAS FOUND. THE REPORTED DEVICE WAS RECEIVED IN LAKE ZURICH AND ITS INVESTIGATION WAS PERFORMED BY OUR LOCAL TECHNICAL TEAM. ACCORDING TO PROVIDED INFORMATION, NOTHING WAS NOTICED DURING BOTH EXTERNAL AND INTERNAL CHECK. DURING TESTS, DEVICE REPORTS AIR BUBBLE WHEN TUBE WITH WATER IS INSTALLED. CALIBRATION OF THE AIR SENSOR RESOLVED THE ISSUE. THIS COMPLAINT WILL BE CONSIDERED AS A VALID, THE ROOT CAUSE IS LIKELY DUE TO DRIFT OF VALUE, SO AIR CALIBRATION NEEDED. PLEASE BE INFORMED THAT AN INTERNAL CAPA WAS OPENED FOR DEFECTIVE AIR SENSORS, IN PARTICULAR A DRIFT OF THE SENSORS. TO OUR KNOWLEDGE THIS COMPLAINT IS NOT BEING RELATED TO PATIENT SAFETY THIS COMPLAINT IS CONSIDERED AS: VALID THE TREND IS: NORMAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24864211 ↗MalfunctionApr 13, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: INTERNAL COMPONENT LOOSE AND RATTLING CUSTOMER SENT IN THE FORM REPORTING THE INTERNAL COMPONENT LOOSE AND RATTLING (LOOSE SCREW). REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24863972 ↗MalfunctionFeb 5, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: AIR ERROR: CUSTOMER EMAILED IN THE REQUEST FORM REPORTING: CANNOT OPERATE PUMP. CONTINUES TO GIVE AIR ERROR. EVEN THOUGH WE HAVE VISUALIZED NO AIR IS PRESENT IN THE LINE. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24272241 ↗MalfunctionFeb 5, 2026No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: AIR ERROR: CUSTOMER EMAILED IN THE REQUEST FORM REPORTING: CANNOT OPERATE PUMP. CONTINUES TO GIVE AIR ERROR. EVEN THOUGH WE HAVE VISUALIZED NO AIR IS PRESENT IN THE LINE. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24272206 ↗MalfunctionJan 30, 2026Excess Flow or Over-Infusion; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY CUSTOMER: PUMP INFUSED MEDICATION AT FAST RATE. CUSTOMER EMAILED IN THE REQUEST FORM REPORTING: PUMP INFUSED MEDICATION AT RATE FASTER THAN IT WAS PROGRAMMED FOR. REPORTING AS A CONSERVATIVE MEASURE. ADVERSE EFFECTS WERE NOT REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24220856 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121613. One submission often covers many models, so check the product and lot details in each record.
- Z-2064-2019 · Jun 24, 2019 · TerminatedVOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a Product Usage: The Volumat MC Agilia is a piece of transportable medical equipment intended to
- Z-0362-2019 · Dec 2, 2017 · TerminatedVolumat MC Agilia Volumetric Infusion Pump, Model No. Z021135, Software version 1.7. Product Usage - The Volumat MC Agilia is a piece of transportable medical equipment intended to
Other Pump, Infusion devices
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,514 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion2 recalls45 deaths148,093 reports
- BD® Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion5 recalls30 deaths51,256 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls4 deaths21,956 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths15,547 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion3 recalls6 deaths13,022 reports
- CADD 21-6500-51Icu Medical, Inc. · Pump, infusion2 deaths8,231 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page