Volumat™ Line VL ST 10-0Recalls & reported problems
Fresenius Kabi AG · Class 2
All models of Volumat Line (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04250273714517), received between Jan 4, 2017 and Jan 5, 2017 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (K121613); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak3
- Chemical Spillage1
- Leak/Splash1
Patient problems
- Chemical Exposure2
- No Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 5, 2017Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THE TRANSFUSION SET SPIKE FELL OUT OF THE IV BAG CAUSING A LEAK. THE EVENT DID NOT OCCUR DURING INFUSION AND THE IV BAG WAS JUST HANGING UP ON THE POLE. THE CUSTOMER HAD TO USE A CHEMO SPILL KIT TO CLEAN THE SPILL. THE DRUG AVASTIN GOT ON THE NURSE BUT WAS NOT HARMED. THE DRUG SPILL DID NOT GET ON THE PATIENT. THE CUSTOMER DID NOT PROVIDE ANY SAMPLE, PICTURES OR INFORMATION REGARDING THE BATCH NUMBER. THEREFORE, A DETAILED ANALYSIS COULD NOT BE PERFORMED AND A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Information
MAUDE report 6226866 ↗MalfunctionJan 5, 2017Fluid/Blood Leak; Leak/Splash; Chemical Spillage; Fluid/Blood Leak; Leak/Splash; Chemical Spillage
THE CUSTOMER REPORTED THE SET WAS BROKEN BELOW THE DRIP CHAMBER. THE INFUSION TREATMENT HAD NOT BEEN STARTED AS THE CUSTOMER WAS PUTTING THE TREATMENT ON THE POLE. THE DRUG CAME INTO CONTACT WITH THE NURSE, BUT SHE WAS NOT HARMED. THE DRUG DID NOT SPILL ON THE PATIENT. THE LEAK WAS CLEANED UP USING A CHEMO SPILL KIT. THERE WAS NO SAMPLE RETURNED, BUT PICTURES WERE AVAILABLE FOR REVIEW. THE BATCH NUMBER IS UNKNOWN. THE INSPECTION OF THE PROVIDED PHOTOGRAPHS SHOWED A VISIBLE HOLE IN THE SILICONE TUBE (IT IS A PART OF THE PUMP SEGMENT) NEXT TO THE TRANSPARENT ADAPTER. THE PVC TUBES AND DRIP CHAMBER WERE FOUND TO BE SEPARATED. A VISIBLE MARK OF THE DRIP CHAMBER CONNECTOR (THIS IS THE POINT WHERE THE TUBING IS PUSHED) WAS ALSO PRESENT. WITHOUT THE BATCH NUMBER, THE PRODUCTION DOCUMENTS COULD NOT BE INVESTIGATED AND NO RETAIN SAMPLES COULD BE TESTED. THE CAUSE OF THE LEAK WAS LIKELY DUE TO INSUFFICIENT GLUE APPLIED ON THE TUBE DURING THE MANUFACTURING PROCESS. A CAPA WAS INITIATED IN THE PRODUCTION FACILITY TO ADDRESS THIS MANUFACTURING ISSUE.
Patient problems: Chemical Exposure; No Known Impact Or Consequence To Patient
MAUDE report 6226596 ↗MalfunctionJan 4, 2017Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THE SET WAS LEAKING DUE TO THE SPIKE DISCONNECTING FROM THE BAG DURING SETUP WHILE AT PATIENT BEDSIDE. THE SET WAS LEAKING THE DRUG GEMZAR AND THE NURSE AND PATIENT WERE EXPOSED BY DIRECT CONTACT TO THE DRUG. THE NURSE TAPED THE IV BAG TO THE SET WITH PHARMACY TAPE AND IT IS STILL NOT HOLDING THEM TOGETHER. DUE TO THE EXPOSURE TO THE DRUG THE NURSE AND PATIENT HAD TO WASH UP WITH SOAP AND CHANGE THEIR CLOTHES. AN INVESTIGATION COULD NOT BE PERFORMED SINCE THERE WAS NO SAMPLE, NO PICTURES AND NO BATCH NUMBER AVAILABLE. THEREFORE, A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: Chemical Exposure
MAUDE report 6223852 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121613. One submission often covers many models, so check the product and lot details in each record.
- Z-2064-2019 · Jun 24, 2019 · TerminatedVOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a Product Usage: The Volumat MC Agilia is a piece of transportable medical equipment intended to
- Z-0362-2019 · Dec 2, 2017 · TerminatedVolumat MC Agilia Volumetric Infusion Pump, Model No. Z021135, Software version 1.7. Product Usage - The Volumat MC Agilia is a piece of transportable medical equipment intended to
Other Pump, Infusion devices
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,514 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion2 recalls45 deaths148,093 reports
- BD® Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion5 recalls30 deaths51,256 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls4 deaths21,956 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths15,547 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion3 recalls6 deaths13,022 reports
- CADD 21-6500-51Icu Medical, Inc. · Pump, infusion2 deaths8,231 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page