Volumat™ Line VL SP40-11Recalls & reported problems
Fresenius Kabi AG · For use with Agilia VP MC pumps / Volumat MC Agilia pump or gravity only. · Class 2
All models of Volumat Line (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 04250273733860), received between Mar 20, 2026 and Mar 20, 2026 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Connection Problem” (14 reports).
2 other recalls cover devices cleared under the same FDA submission (K121613, K220301); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Connection Problem14
- No Apparent Adverse Event14
- Obstruction of Flow2
Patient problems
- No Clinical Signs, Symptoms or Conditions14
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2026Obstruction of Flow; Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: WILL NOT FLOW AND NURSE WAS TRYING TO ATTACH A FLUSH TO SEE IF SHE COULD FIGURE WHERE IT WAS CLAMPED AND WHERE SYRINGE IS ATTACHED IS THE PART THAT WILL "POP OFF" AND START DRIPPING. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24651367 ↗MalfunctionMar 20, 2026Obstruction of Flow; Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: WILL NOT FLOW AND NURSE WAS TRYING TO ATTACH A FLUSH TO SEE IF SHE COULD FIGURE WHERE IT WAS CLAMPED AND WHERE SYRINGE IS ATTACHED IS THE PART THAT WILL "POP OFF" AND START DRIPPING. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24651358 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646759 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646758 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646757 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646756 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646755 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646754 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646753 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646752 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646751 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646750 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646749 ↗MalfunctionMar 20, 2026Connection Problem; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED BY THE CUSTOMER: CUSTOMER REPORTED THAT THEIR PRIMARY TUBING (M6445475) WHERE ANOTHER TUBING LINE CAN BE ATTACHED IS POPPING OUT CAUSING DRIPPING ONTO THE FLOOR OR PATIENT. NO ADVERSE EVENTS REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646748 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121613, K220301. One submission often covers many models, so check the product and lot details in each record.
- Z-2064-2019 · Jun 24, 2019 · TerminatedVOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a Product Usage: The Volumat MC Agilia is a piece of transportable medical equipment intended to
- Z-0362-2019 · Dec 2, 2017 · TerminatedVolumat MC Agilia Volumetric Infusion Pump, Model No. Z021135, Software version 1.7. Product Usage - The Volumat MC Agilia is a piece of transportable medical equipment intended to
Other Pump, Infusion devices
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,514 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion2 recalls45 deaths148,093 reports
- BD® Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion5 recalls30 deaths51,256 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls4 deaths21,956 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths15,547 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion3 recalls6 deaths13,022 reports
- CADD 21-6500-51Icu Medical, Inc. · Pump, infusion2 deaths8,231 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.