Biotronik Fortress Introducer Sheath System 6F; 45 cm; straightRecalls & reported problems
Contract Medical International GmbH · Introducer Sheath System containing One (1) introducer Sheath with haemostasis valve and One (1) dilator. · Class 2
All models of Biotronik Fortress Introducer Sheath System (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04251244503703), received between Sep 24, 2024 and Nov 18, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Stretched” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 30, 2025Class IIOpen, Classified
Units may have been packaged with an incorrect pouch label, which lists the reference number and unique device identifier of the 4F version of the device, rather than the packaged 6F version.
Z-2532-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Stretched1
- Unsealed Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 18, 2024Stretched; Stretched
THE PROBLEM WAS DESCRIBED AS: "IT WAS A LEFT CFA PUNCTURE FOR A RIGHT LEG ANGIOGRAM. UPON REMOVAL OF THE FORTRESS 6F 45CM SHEATH RESISTANCE WAS ENCOUNTERED. SHEATH WAS ATTEMPTED TO BE WITHDRAWN, BUT IT WAS NOTED THAT THE SHEATH WAS STRETCHING. UPON REALIZATION, A 6F 65CM FORTRESS WAS OPENED, IN ORDER TO GET THE 65CM INTRODUCER. IT WAS MANAGED TO INSERT THE INTRODUCER INTO THE SHEATH AND REMOVED IT. NB: CLINICAL NURSE STATED THAT THE DELIVERY OF THE SHEATH OVER THE ILIAC BIFURCATION WAS UNEVENTFUL. FORTRESS VALVE AND SIDE PORT WERE REPLACED WITH A COOK VALVE AND SIDE PORT, AS THERE HAD BEEN A SEPARATE ISSUE WITH OTHER COMPANIES STENT ACCIDENTLY DEPLOYING IN THE FORTRESS SHEATH. THEREFORE, THIS PART OF THE FORTRESS SHEATH NEEDED TO BE REMOVED AND REPLACED. THIS EVENT OCCURRED WELL BEFORE THE ATTEMPT TO REMOVE THE FORTRESS SHEATH." EVENT OCCURRED: INSIDE PATIENT PATIENT INJURY: NO PATIENTS OUTCOME: DISCHARGED AT HOME. ON 23.10.2024 CUSTOMER PROVIDED ADDITIONAL INFORMATION: "I HAVE SPOKEN WITH THE SCRUB NURSE WHO SAID THAT THE 6F FORTRESS 45CM SHEATH DISTAL END WAS IN THE AORTA WHEN THEY ENCOUNTERED RESISTANCE WHEN PULLING BACK THE SHEATH. THEY WERE REMOVING IT WITHOUT THE DILATOR WHIC
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20703826 ↗MalfunctionSep 24, 2024Unsealed Device Packaging; Unsealed Device Packaging
THE PROBLEM WAS DESCRIBED AS: "THE FORTRESS INTRODUCER SHEATH IS NOT SEALED AT THE LOWER END. THE SHEATH IS NOT SEALED AT THE BOTTOM. THIS MEANS THAT WHEN THE SHEATH IS HANDED OVER AND THE OUTER NON-STERILE PACKAGING IS OPENED TO OPEN THE STERILE PACKAGING, THE SHEATH FALLS OUT BY ITSELF... THE BOTTOM SEAL IS MISSING. NOTHING HAS HAPPENED NOW... THE SHEATHS JUST COULDN'T BE USED BECAUSE THEY WERE NO LONGER STERILELY PACKAGED AND SOME OF THEM HAD FALLEN TO THE FLOOR."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20290017 ↗
Other Introducer, Catheter devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page