Biotronik Fortress Introducer Sheath System recallZ-2532-2025 · Contract Medical International GmbH · initiated Jul 30, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Contract Medical International GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Units may have been packaged with an incorrect pouch label, which lists the reference number and unique device identifier of the 4F version of the device, rather than the packaged 6F version.
FDA root cause category: Under Investigation by firm
What the company is doing
On July 30, 2025, the firm notified affected consignees via email. Customers were instructed to verify if the labeling information on received products is consistent. If customers identify incorrect labeling, they should contact Contract Medical International GmbH for replacement.
Recall details
- Recall number
- Z-2532-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 30, 2025
- Posted by the FDA
- Sep 5, 2025
- Recalling firm
- Contract Medical International GmbH
- Quantity
- 2500
- Product code
- DYB
- FDA submissions
- K153197
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, CT, FL, GA, KY, MA, MI, MO, NJ, NY, OR, RI, SC, TN, TX and the countries of Switzerland, Hungary, Italy, France, Denmark, Austria, Australia, Belgium, Poland, Germany, Bulgaria, Kazakhstan, Finland, Ireland, South Korea, Taiwan, Israel, Spain, Vietnam, Malaysia, Czechia, Thailand, Libya, Great Britain, Chile, Slovakia, Slovenia, Croatia, Sweden, Argentina.
Product description
Brand Name: Fortress Introducer Sheath System Product Name: Fortress Introducer Sheath System Model/Catalog Number: 386594 Product Description: Fortress Introducer Sheath System, 6F Straight 45cm:
Code information (models, lots, serial numbers)
Lot Code: Model Number: 386594 Lot Number: 800989 Unique Device Identifier: 04251244503703 Expiration Date: 31 OCT 2027.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Biotronik Fortress Introducer Sheath System 6F; 45 cm; straightContract Medical International GmbH · Introducer, Catheter1 recall2 reports
Product lines
Other devices under the same FDA submission
7 devices cleared under K153197 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Biotronik Fortress Introducer Sheath System 45 cm, 6F CrossoverContract Medical International GmbH · Introducer, Catheter4 reports
- Biotronik Fortress Introducer Sheath System 90 cm, 6F StraightContract Medical International GmbH · Introducer, Catheter2 reports
- Biotronik Fortress Introducer Sheath System 6F; 45 cm; crossoverContract Medical International GmbH · Introducer, Catheter2 reports
- Biotronik Fortress Introducer Sheath System 45 cm, 5F StraightContract Medical International GmbH · Introducer, Catheter1 report
- Biotronik Fortress Introducer Sheath System 100 cm, 6F StraightContract Medical International GmbH · Introducer, Catheter1 report
- Biotronik Fortress Introducer Sheath System 65 cm, 5F StraightContract Medical International GmbH · Introducer, Catheter1 death1 report
- Biotronik Fortress Introducer Sheath System 90 cm, 5F StraightContract Medical International GmbH · Introducer, Catheter1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page