ManoScan 3885Recalls & reported problems
Given Imaging, Inc. · AR 3D Probe · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04260167480954), received between May 24, 2019 and Jan 8, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Deflation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Deflation Problem1
- Difficult to Remove1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 8, 2020Deflation Problem; Deflation Problem
ACCORDING TO THE REPORTER, WHEN THEY REMOVED THE CATHETER AFTER THE PROCEDURE, THE BALLOON WAS NOT DEFLATED EVEN THOUGH THE SOFTWARE SHOWED THERE WAS NO PRESSURE ON THE BALLOON AND THE BALLOON WAS DETACHED AND REMAINED WITH THE PATIENT. THEY THOUGHT THAT THE PATIENT'S ANATOMY WAS THE ISSUE WHICH CAUSED AN AIRLOCK WITHIN THE SYRINGE. THE PATIENT SUBSEQUENTLY DID HAVE A FLEXIBLE SIGMOIDOSCOPY TO RETRIEVE THE DEFLATED, DETACHED BALLOON. THE CATHETER WORKED CORRECTLY DURING THE PREVIOUS PROCEDURE. THERE WAS NO USER HARM.
Patient problems: Foreign Body In Patient; Patient Problem/Medical Problem
MAUDE report 9567293 ↗MalfunctionMay 24, 2019Difficult to Remove; Difficult to Remove
ACCORDING TO THE REPORTER, THE SHEATH BALLOON BROKE OF THE CATHETER WHEN DOCTOR REMOVED IT FROM THE PATIENT WITH 30CC OF AIR STILL IN THE BALLOON. THE PHYSICIAN ATTEMPTED TO REMOVE THE BALLOON FROM THE PATIENT WITHOUT SUCCESS AND ADVISED THAT THE BALLOON WILL PASS. THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: Device Embedded In Tissue or Plaque; No Consequences Or Impact To Patient
MAUDE report 8640776 ↗
Other SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL) devices
- ManoScan 3890Given Imaging, Inc. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)90 reports
- Solar GI LGI H20, Trolley system G3-6Medical Measurement Systems B.V. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall3 reports
- Solar GI HRM H20, 24p, Trolley system G3-8Medical Measurement Systems B.V. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall2 reports
- Solar GI HRM H2O, 36p, Trolley system G3-12Medical Measurement Systems B.V. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall1 report
- EndoFLIP EF-100Covidien LP · System, Gastrointestinal Motility (Electrical)1 recall0 reports
- Solar GI HRM H20, 24p, Compact Pole system G3-7Medical Measurement Systems B.V. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall0 reports
- Zvu® ZVU-3Diversatek Healthcare, Inc. · System, Gastrointestinal Motility (Electrical)1 recall0 reports
- Endoflip RFG-EF100Covidien LP · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.